Field Calibration Instructions Overview Brasch Generation 2 sensor boards are calibrated at the factory and have an expected calibration duration of approximately two years. Environmental conditions may shorten or prolong this period of time, but once the sensor parameters drift outside of acceptable tolerances, the sensor board must be either be recalibrated or replaced.
Page 2
Procedure: 1. Turn power OFF to the sensor board 2. Move the calibration jumpers (JP4) so that both pins are covered by the jumpers 3. Turn power ON to the sensor board 4. Connect the negative voltmeter lead to COM (TP3) 5.
Page 3
Troubleshooting: • If any of the following errors are present, calibration data was corrupted and this procedure needs to be repeated. ◦ Standalone Detectors: 9997 ◦ BMS Transmitters: 0x0008 ◦ Control Panel Remotes: Sensor # Calibration Invalid • If achieving stable voltages is not possible, it may be necessary to replace the sensor and/or the sensor board.