Page 1
5.0 PPM Nitrogen Dioxide Test Gas (for NO sensors) • Gas Flow Regulator • Tubing • Calibration Hood Note: All of the above required parts (except a voltmeter) are included with Brasch CO and Test Gas Kits. FG01 Rev 1.2 – January 10, 2022...
Page 2
Procedure: 1. Turn power OFF to the sensor board 2. Move the calibration jumpers (JP4) so that both pins are covered by the jumpers 3. Turn power ON to the sensor board 4. Connect the negative voltmeter lead to COM (TP3) 5.
Page 3
Troubleshooting: • If any of the following errors are present, calibration data was corrupted and this procedure needs to be repeated. ◦ Standalone Detectors: 9997 ◦ BMS Transmitters: 0x0008 ◦ Control Panel Remotes: Sensor # Calibration Invalid • If achieving stable voltages is not possible, it may be necessary to replace the sensor and/or the sensor board.