Messages about batteries
DESCRIPTION: "E.1" appears, or the unit does not turn on at all.
PROBLEM: the batteries are dead.
SOLUTION: remove the batteries immediately and replace them when neces-
sary (see par. #9).
Messages about thermometer's stabilization
DESCRIPTION: the projection/display shows "E.8".
PROBLEM: the thermometer was moved before the light started flashing, or
the area is subject to strong electromagnetic fields.
SOLUTION: wait until the lights flash before moving the thermometer; make
certain that there are no mobile or cordless telephones in the vicinity.
Messages about room temperature working range
11. MEANING OF DISPLAY MESSAGES
DESCRIPTION: during normal operation, the symbol
"bAt" appears.
PROBLEM: the batteries are running low but it is still
possible to take several readings.
SOLUTION: obtain new batteries for changing them
when the signal "E.1" appears.
DESCRIPTION: the display shows shows "C:AL" and/or
a count-down (in minutes and seconds).
PROBLEM: the thermometer has not stabilized. The sym-
bols prompt you to choose between:
SOLUTION:
- wait until the AQCS countdown has run its course wit-
hout touching the thermometer, or
- perform an MQCS (par. #8).
DESCRIPTION: while in stand-by mode, the room tem-
perature displayed is accompanied by the symbol MQCS
or AQCS.
MEANING: the thermometer has undergone quick auto-
matic or manual calibration.
DESCRIPTION: when you press the "face"
the projection/display shows «Hi.4» and the value alter-
nately.
PROBLEM: the room temperature is between 40,1 and
45,0°C (104.1 and 113.1ºF).
SOLUTION: the temperature reading can be taken but
accuracy is not guaranteed.