Status screens
The status screens show the most important device data. You can switch between the
individual status screens by pressing "next".
Ambient
P
air pressure [hPa]
RH
internal relative humidity [%]
T
internal temperature [°C]
Alt
calculated height above sea level [m]
Note: Altitude is calculated for average pressure. It may be off by several 100 me-
ters if ambient air pressure is very high or low due to a high- or low-pressure area
passing through.
corr
current ambient pressure and temperature compensation [%]
High voltage
HV
charger corona voltage [V]
Idiff
charger diffusion current [nA]
DV
deposition voltage [V]
Electrometers
EM1
signal of electrometer 1 [mV]
EM2
signal of electrometer 2 [mV]
A1
electrometer 1 amplitude [mV]
A2
electrometer 2 amplitude [mV]
Miscellany
Batt
battery voltage [V]
Error
displays the current error status (0 = no error)
Flow
approximate flow through device [lpm]
ThermoController
Ts_pt sensor set temperature
Ts_pr evaporation tube set temperature
T_pt
sensor measured temperature
T_Pr
evaporation tube measured temperature
PWM_pt heating power sensor
PWM_pr heating power, evaporation tube
Charger
Vbat
battery voltage
Vin
Input voltage
Ichrg
charging current
Fault
error
Config screens
Time and Date: the internal clock of the HEPaC has an accuracy of about 1s per day. It
does not change to daylight saving time and back automatically. Therefore, you may
need to set the clock from time to time.