Page 3
Material in this manual is for information purposes only. The contents and the product it describes are subject to change without notice. Thermo Scientific makes no representations or warranties with respect to this manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to the use of this manual.
Page 4
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Thermo Scientific has contracted with one or more recycling/disposal companies in each EU Member State European Country, and this product should be disposed of or recycled through them.
Page 5
We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products. Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the telephone without a service...
Altitude: < 2,000 meters (6,650 feet) Warning Whenever working with liquid nitrogen storage equipment in a closed environment, the use of personal O detection devices is strongly recommended. Refer to "Handling Liquid Nitrogen" in the appendix of this manual. ▲ Thermo Scientific 7000740r7 Cryo Series...
2. Insert the power cord into the power outlet module. Tighten screw on the power cord retainer. 3. Position power cord/module Figure 1-2. Power Cord Assembly assembly into outlet on unit. Secure with screws on module. 7000740r7 Cryo Series Thermo Scientific...
Turning the power switch on resets the circuit breaker. If the circuit breaker trips again within a short time period, the unit should be checked by a qualified electrician. 7000740r7 Cryo Series Thermo Scientific...
It is imperative that the positioning of the sleeve not block either the fill or liquid level tube orifices located at the bottom rear of the tank. Should blockage occur, it will cause filling and liquid level indicator problems. Thermo Scientific 7000740r7 Cryo Series...
The contacts are activated after a thirty minute delay when the unit goes into low level alarm. The contacts are marked COM = Common, N.O. = Normally Open, N.C. = Normally Closed Figure 1-5. Contacts Thermo Scientific 7000740r7 Cryo Series...
The actual liq- uid level will be approximately one inch lower than the frost line on the ruler. Caution Some shrinkage of the ruler may occur depending on the level of liquid nitrogen in the tank. ▲ Thermo Scientific 7000740r7 Cryo Series...
Page 15
. The Low Level alarm LED will go out. 7. Press the Manual Fill button. The solenoid LED lights. When the unit is full, release the button, the fill stops, and the LED goes out. 7000740r7 Cryo Series Thermo Scientific...
Page 16
10. The High Level alarm LED, remote alarm contacts, and audible alarm will activate in about 2 minutes. 11. Replace the probes and adjust if necessary. Refer to the instructions in Section 2 of this manual. Figure 1-6. Alarms and LED Test Thermo Scientific 7000740r7 Cryo Series...
When Manual Fill button is pressed, the solenoid opens and unit begins to fill. The Fill indicator lights. Power indicator lights when power is applied to the unit and the switch is turned on. Thermo Scientific 7000740r7 Cryo Series...
LED (cycle on and off ), high and low level LEDs (off ), all valves will close and remote alarm contacts will activate. To clear this alarm (after the alarm condition is corrected), the unit power switch must be turned off, and back on. 7000740r7 Cryo Series Thermo Scientific...
3. The double tie-wrap on the sensor wires are the 3-inch low level and the 5-inch high level settings. Caution Allow probe to warm up for 30 minutes before removing from tube. Pulling probe completely out of the tube when cold could damage probe. ▲ Thermo Scientific 7000740r7 Cryo Series...
• audible alarm will cycle on and off • all valves will close • remote alarm contacts will activate. To clear this alarm (after the alarm condition is corrected), the unit power switch must be turned off, then turned on. 7000740r7 Cryo Series Thermo Scientific...
If a higher level is desired, the Manual Fill button must be held down until that level is reached. The yellow LED solenoid indicator lights whenever the fill solenoid valve activates. Thermo Scientific 7000740r7 Cryo Series...
3. Clean and reassemble in reverse order of disassembly, paying careful attention to exploded view provided for identification and placement of parts. After maintenance, operate the valve a few times to ensure proper opening and closing. Thermo Scientific 7000740r7 Cryo Series...
The interior of the CryoPlus unit is made of high quality stainless steel and should not be cleaned with any cleanser containing chlorine. Exterior Cabinet Wash the outside of the cabinet with a mild detergent. 7000740r7 Cryo Series Thermo Scientific...
Allow the tank to warm at room temperature. Storage Tank Caution Do not attempt to speed LN evaporation, or frost removal, by applying direct heat. Avoid using a heater or blowing compressed air into storage tank. ▲ Thermo Scientific 7000740r7 Cryo Series...
Repair or replace bypass solenoid valve. Open sensor alarm (audible alarm and alarms LED flash, low Defective sensor or sensor wiring harness. Repair or replace cryo liquid level probe assembly and high level LEDs are off). Thermo Scientific 7000740r7 Cryo Series...
▲ Control Board Fuses Two 63mA Slo-Blow fuses (P/N 230162) are located on the Cryo control circuit board which protects power transformer T1. Refer to Figure 4-2. Figure 4-2. Fuse Locations 7000740r7 Cryo Series Thermo Scientific...
** Add 5.0" (12.7 cm) for utilities and lid opening *** 93.5" (237.5 cm) total unit height of CryoPlus 4 with lid open **** Rack = 13-2C/100 ***** Use rack #4000112 Continuing research and improvements may result in specification changes at any time. Thermo Scientific 7000740r7 Cryo Series...
990027 Magnetic lid gasket 990028 Magnetic lid gasket 990029 Magnetic lid gasket 990031 Magnetic lid gasket 180143 LN2 dip measuring ruler 4000400 4 foot transfer hose 4000401 6 foot transfer hose 195884 Lid closing strap Thermo Scientific 7000740r7 Cryo Series...
Page 36
Cryo 300 - Model 744/745 Cryo Plus 4 - Model 7406/7407 Cryo 400 - Model 746/747 4 place frame #4000371 for 250ml canister Sliding canister #4000335 for 4R9953 bag Platform divider for frames #4000371 and 25ml canisters #4000335 Thermo Scientific 7000740r7 Cryo Series...
Page 38
Tank - 24” 4000080 Tank - 31” 4000050 Tank - 31” 4000051 Tank - 31” 4000047 Tank - 31” 4000048 Tank - 31” 4000045 Tank - 39.5” 4000053 Tank - 39.5” 4000054 Tank - 39.5” 4000052 Thermo Scientific 7000740r7 Cryo Series...
1. It is extremely cold. At atmospheric pressure, liquid nitrogen boils at - 320° F (-196° C). 2. Very small amounts of liquid vaporize into large amounts of gas. One liter of liquid nitrogen becomes 24.6 cu. ft. (0.7 m3) of gas. Thermo Scientific 7000740r7 Cryo Series...
Page 47
Be sure to follow instructions supplied with the withdrawal device. When liquid cylinders or other large storage containers are used for filling, follow the instructions supplied with those units and their accessories. 7000740r7 Cryo Series Thermo Scientific...
Page 48
Disposal of liquid nitrogen should be done outdoors in a safe place. Pour the liquid slowly on gravel or bare earth where it can evaporate without causing damage. Do not pour the liquid on pavement. Thermo Scientific 7000740r7 Cryo Series...
108° F. Under no circumstance should the water be over 112° F, nor should the frozen part be rubbed either before or after rewarming. The patient should neither smoke nor drink alcohol. 7000740r7 Cryo Series Thermo Scientific...
Page 50
Thermo Scientific 7000740r7 Cryo Series 10-1...
Page 52
Thermo Scientific Controlled Environment Equipment 401 Millcreek Road Marietta, Ohio 45750 United States www.thermo.com...