Trouble shooting / Table of malfunctions ....16 Table of malfunctions ..........17 How to open the unit ..........24 Getinge Ultrasonic 30, 40, 70, 80 ......24 Getinge Ultrasonic 60, 100 – 300 ......25 How to close the unit / How to mount the bottom plate ................25...
Page 3
Content Getinge Ultrasonic 120 - 300 ........39 9.3.1 How to remove the PCB interference filter ...39 9.3.2 How to mount the PCB interference filter .....40 How to replace the potentiometer for the cleaning period / for temperature ......... 42 How to replace the front panel ....... 43 How to replace the PCB support ......
Please observe the important safety warnings (section 2), and the information on initial operation (section 17). In case of any further queries Getinge will be glad to assist. For our technical support and contact address please see the last page.
New units The limitation period of warranty claims is 2 years (from the date of purchase). If no proof of purchase can be produced, Getinge can find out the date of manufacturing by means of the serial number (nameplate). Repair Getinge grant a limitation period of 2 years on all exchange parts and on any repair works carried out.
This level should not be exceeded even with immersed cleaning items. Plastic handles (on Getinge Ultrasonic 30 and larger) for the safe transportation of the unit, even when heated. Turning knob for draining the transducer tank (on Getinge...
100-120 | 220-240 *: Impulse sound; 30 – 300: Double half-wave sound. The form of signal has been adapted to the geometry of the transducer tank. Depending on the form of signal the peak of the ultrasonic power is 4 or 8-fold.
Product description Operating and display functions For your information: Specific displays of the unit types Getinge Ultrasonic 30 - 300 are listed separately. Action To do Result Display Switch on the unit press on/off key unit is ready for on/off LED is on...
LED display Various malfunctions are indicated by a flashing LED display: malfunction Ultrasound Getinge Ultrasonic 30 - 300: LED display cleaning period flashes fast („running light“) / LED display in the ultrasound key malfunction flashes fast: There is a fault on the PCB control, on the connection between ultrasound >>...
Always pull the mains plug before you open the unit. For checks that require the unit to be connected use a mains interrupting transformer. Getinge Ultrasonic 30, 40, 70, 80 Fastening of bottom The bottom plate with the unit feet is fastened to the unit plate housing by clips.
How to open the unit Getinge Ultrasonic 60, 100 – 300 Fastening of bottom The bottom plate is screwed to the unit through the unit feet. In plate addition, the bottom plate is screwed to the cooling body of the PCB control with 2 screws.
How to replace the PCB interference filter How to replace the PCB interference filter Getinge Ultrasonic 30 - 40 The PCB interference filter is screwed to the back of the unit. The mains input socket is soldered to the PCB interference filter and must be replaced with the PCB.
Page 34
How to replace the PCB interference filter Illustration 9.9. Electric connections for 1 heating element (e.g. S 30 Voltage supply for PCB control (phase – black). Voltage supply for PCB control (zero – white). Connection for heating control (grey). Connection for heating control (white).
How to replace the PCB interference filter Getinge Ultrasonic 60 - 100 The PCB interference filter is screwed to the back of the unit. The mains input socket is soldered to the PCB interference filter and must be replaced with the PCB.
How to replace the PCB interference filter Getinge Ultrasonic 120 - 300 The PCB interference filter is screwed to the back of the unit. The mains input socket is soldered to the PCB interference filter and must be replaced with the PCB.
On many units the heating element can be replaced without removal of the transducer tank tank. Exceptions: Getinge Ultrasonic 60, 100, 120, 180; 300. On these units the transducer tank must be removed first as described in section 15.
(This does not apply to units with a single transducer system.) Therefore, all transducer disks should be replaced even if only individual piezoceramics are faulty. Getinge deliver preselected sets of piezoceramics. Cause of breakdown Piezoceramics can be damaged for different reasons (mostly tears): e.g.
Putting out of operation and disposal The components can be given to electronic and metal recycling stations or returned to the manufacturer for the same purpose. Contact address Getinge Infecction Control AB Box 69 30505 Getinge Phone +46 (0) 10 335 00 00 e-mail: info@getinge.com...