3 Installation
3.7
Output of current with HART communication
Connection for
Source (passiv input card)
Sink (active input card)
a
The two DIP-switches are located at the front of the device. Settings made to the device with DIP-switches must
occur in a voltage-free state.
3.8
HART® communication
HART communicators (HHT) can be connected as shown in the basic circuit diagram. Test sockets (di-
ameter 2.3 mm) are integrated in the screw connection terminal blocks for this purpose.
3.9
Output Voltage
Connection for
Source – passive input card
a
The two DIP-switches are located at the front of the device. Settings made to the device with DIP-switches must
occur in a voltage-free state.
3.10
Voltage supply
The voltage supply is fed in via the connection terminals 1.1 (+) und 1.2 (-) (DC/AC 24 to 230 V).
3.11
Dimensions
14
Electrical
Connection
circuit im-
of the input card
pedance
at terminal
250
3.1 (+) and 3.2 (-)
250
3.1 (+) and 3.2 (-)
250
3.2 (+) and 3.3 (-)
250
3.2 (+) and 3.3 (-)
Connection of the input card at the terminal
3.1 (+) and 3.2 (-)
112 5 .
99
DIP-switch position
of the HART
S1
communicator
3.1 and 3.2
I
3.2 and 3.3
I
3.2 and 3.3
I
-
I
DIP-switch position
S1
II
17 5 .
a
S2
II
I
II
II
a
S2
II