Installation; Direct Adhesive Mount - IMI SENSORS HT640B01 Installation And Operating Manual

High temperature 4-20 ma velocity transmitter
Table of Contents

Advertisement

Installation

Installation
When choosing a mounting method, consider closely the advantages and disadvantages of each technique.
When choosing a mounting method, consider closely the advantages and disadvantages of each technique.
Typical mounting types are stud, direct adhesive, adhesive mounting base and magnetic mounting base. Since
Typical mounting types are stud, direct adhesive, adhesive mounting base and magnetic mounting base. Since
the frequency response is limited to 1 kHz on the 640/641, any of the four methods can be used without seriously
the frequency response is limited to 1 kHz on the 640/641, any of the four methods can be used without seriously
affecting the data values. The mounting method might affect some of the 645/646 series because of the
affecting the data values. The mounting method might affect some of the 645/646 series because of the
increased frequency response.
increased frequency response.
Direct Adhesive Mount Procedure
Direct Adhesive Mount Procedure
For restrictions of space or for convenience, most sensors (with the exception of integral stud models) can be
adhesive-mounted directly to the machine surface.
STEP 1 Prepare a smooth, flat mounting surface. A minimum surface finish of 63 µin [0,00016 mm] generally
works best.
STEP 2 Place a small portion of adhesive on the underside of the sensor. Firmly press down on the top of the
assembly to displace any adhesive. Be aware that excessive amounts of adhesive can make sensor removal
difficult.
Figure 5 – direct adhesive mounting
Firmly Press Down
Adhesive
Typical
Accelerometer
PAGE 9

Advertisement

Table of Contents
loading

Table of Contents