FREEZER COMPONENTS
1. Outer door: To open the door, grip the handle. On closing, lock the door latch completely.
2. Handle: Always grip this handle to open and close the outer door.
3. Door latch: Always lock the latch when the outer door is closed.
4. Magnetic door gasket: This provides a tight door seal and prevents cold air leak. Keep clean.
5. Access port (rear and bottom):
This is used for leading a cable and sensor of a measuring
equipment, or nozzle of back-up system to chamber.
6. Inner door: The operation of the inner door should be quick to minimize the temperature rise in
chamber. Lock the door latch completely when the door is closed. The door is removable for cleaning
or defrosting. See page 27 "Routine maintenance".
o
7. Lock: Turn clockwise to 180
with a key and the outer door is securely locked.
8. Control panel (on the upper front of the door): Used for temperature setting and indication of
operating status is displayed on the panel. See page 10 for details.
9. Caster: 4 casters are provided to facilitate moving of the cabinet. For the installation, adjust the
leveling foot so that the front 2 casters cannot contact with the floor.
10. Leveling foot: The height of the freezer can be adjusted by this screw type foot. Keep the unit in
level at the installation.
11. Fixture (on back side): 2 fixtures are provided as spacers between the cabinet and wall and also
serve as hooks to fix the unit. See page 12 "Installation".
12. Air intake vent (grille): Do not block this vent to keep the proper cooling performance.
13. Power switch: This is for turning ON/OFF the power to the unit. ON – "I" OFF – "○"
14. Space for temperature recorder: An automatic temperature recorder (optional component) can be
attached here. See page 13 "Temperature recorder".
15. Condenser filter (behind the grille):
This filter prevents the dust from accumulating on the
condenser. The dusty filter may cause failure of refrigerating device. Clean the filter once a month.
See page 26 "Routine maintenance" for the cleaning.
16. Battery switch: This is a switch for a battery for power failure alarm. Normally, turn on this switch.
Be sure to turn off this switch if the freezer is not in operating for the long period (over one month).
17. Remote alarm terminal: This is used to notice an alarm condition of the unit to remote location.
Refer to page 22 "Remote alarm terminal".
18. Space for optional component:
19. Door switch: This switch detects the open/close status of outer door.
9

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mdf-u72vcMdf-u52vMdf-u52vc

Table of Contents