Ⅶ.error Message - FLOWTECH KF700H Series Operation Instruments

Electromagnetic flowmeter
Table of Contents

Advertisement

Ⅶ、Error information
Error
Flow measurement value
Upper limit
over than the upper limit
value alarm
Flow measurement value
Lower limit
lower than the lower limit
value alarm
Excitation
Empty tube stage is show
Empty tube
zero or random data.
Zero point
Instant value exceeds
Over range
instrument declare value
Contents
Excitation circuit is not
working correctly
Zero point value too
High on zero trim
CONVERT
Limit alarm set value is lower than the flow
measurement, modify the upper limit alarm
settings
Limit alarm set value is over than the flow
measurement, modify the lower limit alarm
Settings
A) check cables terminal and electrical
excitation of the terminal connection is good
or not
B )check the sensor excitation circuit don't
existence of open or short circuit
C excitation coil temperature is too high
D excitation frequency set too high
A)flow meter sensor is not full of medium
B) electrode surface was completely covered
by insulating layer
C) signal lines to connect the signal is incorrect
or open loop
D )measuring low conductivity medium
E )empty and full trim is not correct, or tube
region % is high sensitivity settings
A) on the zero trim time, the flowmeter sensor
medium in a state of non-full pipes
B )on the zero trim time, the sensor tube in a
non-static state media
C) flowmeter grounding is incorrect or
unreliable and technical requirements of
re-grounding
Over the instruments max allow the value, pls
re-select the more
Large diameter of the flowmeter
TRANSIMITTER
CONVERT PARTS
Reason
Http: // www.flowmeter.cc
40