Tank cover Outlet nozzle Flow control valve Inlet nozzle Elbow Cooling coil Circulation tank Control panel Maintenance panel Drain with stopper Castor (with Stopper) Fig.1 Recirculating Cooler Electrical Installation THIS EQUIPMENT MUST BE EARTHED SRC14 Before connection please ensure that the line Instructions for use supply corresponds to that shown on the rating plate.
General Description The SRC14 accurately controls the temperature of a cooling liquid between -20 to +30ºC. The cooled liquid is then circulated to an external device via the inlet and outlet nozzle. The SRC14 has a capacity of 14 litres. Safety Advice Before Use •...
After operation: Switch the circulation pump off, then switch off the power switch and the mains power. Place an appropriate collection vessel for the cooling medium underneath the drain hose and the drain stopper. Once all fluid has been drained replace the drain stopper. Maintenance &...
General Maintenance Bibby Scientific Ltd. Periodically wipe clean the inside of the circulation Stone, Staffordshire ST15 0SA tank and locate the fuse box at the rear of the unit United Kingdom (Fig. 6). Use mild detergent if necessary. Regularly Tel: +44 (0) 1785 812121 check connections for leakage.
Page 6
Instructions for use Note: Alarm is cleared by turning off the power switch Problem Cause Solution Cooling liquid is not Cooling liquid filter is blocked. Clean the filter – See the Maintenance and circulated. Servicing section for details. Air is trapped in the pump. Briefly remove the drain stopper.
Page 7
This product meets the applicable and so we cannot guarantee that interference EC harmonized standards for radio will not occur in practice. Where there is a frequency interference and may be possibility that injury, damage or loss might occur expected not to interfere with, or be affected if equipment malfunctions due to radio frequency by, other equipment with similar qualifications.
INSPECTION REPORT MODEL SRC14 ELECTRICAL SAFETY 1. Earth continuity 2. Insulation 3. Flash test FUNCTIONAL 1. Indicators 2. Speed/temperature 3. Visual acceptance QUALITY CONTROL INSPECTOR Bibby Scientific France SAS Bibby Scientific Italia Srl ZI du Rocher Vert - BP 79 Via Alcide de Gasperi 56 77793 Nemours Cedex 20077 Riozzo di Cerro al Lambro...
Need help?
Do you have a question about the Stuart SRC14 and is the answer not in the manual?
Questions and answers