  • Page 1 RIDA qLine Orbital Shaker ® User Manual https:/ /clinical.r-biopharm.com...
  • Page 2 RIDA qLine® Orbital Shaker | User Manual...
  • Page 3 User Manual RIDA qLine® Orbital Shaker Item no. ZG2601: Version 16.00 (2022-03-03) Copyright 2022 by R-Biopharm AG © All rights reserved. No part of this publication may be duplicated, communicated, rewritten, or translated into another language or computer language, regardless of the form or means, without explicit prior written permission.
  • Page 4: Table Of Contents

    RIDA qLine® Orbital Shaker | User Manual Contents General information Uninstalling Unpacking the device and disposal Safety information Uninstalling the device Disposal of reagents Operation and and materials description of the RIDA Device disposal qLine® Orbital Shaker Description Packaging and storing the device Intended use Dimensions and installation of...
  • Page 5: General Information

    Failure to observe the instructions in this manual will void the manufacturer’s warranty and can pose a risk to users. Make sure that all users are familiar with the device. When in doubt, contact R-Biopharm AG if you need additional equipment or training before using the device. 1 General information 1.1 Unpacking the device...
  • Page 6 RIDA qLine® Orbital Shaker | User Manual 2. This user manual needs to be available at all times, especially to the people who would like to use this device. Therefore, keep this user manual in the proximity of the device. Strictly observe this user manual.
  • Page 7 Note: Failure to follow the safety and operating instructions in the user manual and improper handling of the device can impact the safety that the device affords and will void the warranty. Important: Only trained users may operate the device. Important: Do not use this device in an unsafe environment and especially not in an explosion- hazardous environment.
  • Page 8: Operation And Description Of The Rida Qline® Orbital Shaker

    RIDA qLine® Orbital Shaker | User Manual 2 Operation and description of the RIDA qLine® Orbital Shaker 2.1 Description The RIDA qLine® Orbital Shaker is a shaker that has orbital shaking motion. The power switch on the front side is used to switch on and off the device. The speed is set to 300 rpm by default.
  • Page 9: Dimensions And Installation Of The Rida Qline® Orbital Shaker

    2.3 Dimensions and installation of the RIDA qLine® Orbital Shaker The following image shows the installation dimensions, such as length × width × height. 90 mm 140 mm 170 mm Important: Make sure that the device is resting on all 4 feet. The device must be placed on a horizontal surface.
  • Page 10: Ventilation Requirements During Operation

    RIDA qLine® Orbital Shaker | User Manual 2.3.1 Ventilation requirements during operation Ensure adequate and constant ventilation of the RIDA qLine® Orbital Shaker during operation. The shaker’s ambient temperature should not exceed 40 °C. As a user, you are responsible for providing adequate ventilation to ensure that the device does not overheat. Users are responsible for the ventilation.
  • Page 11: Required Consumables, Not Included With The Device

    2.4.3 Required consumables, not included with the device Product Item no. RIDA qLine® QC-Kit ZG2701 (for qLine soft) RIDA qLine® Allergy A6442, A6242, A6342, A6142, A6142EC2, A6142VIET, A6142PSMI, A6442UA1, A6242H, A6242PA, A6342UY, A6142UZ, A6442TZA, A6342MENA, A6442UZ, A6442BY, A6342KE, A6342BY, A6142HVEN, A6142UY, A6442UA2, A6242GR, A6342PH, A6242PH, A6142BY, A6242BY, A6242UY, A6142ZW2, A6142EAWU, A6242EAWU, A6342EAWU, A6442EAWU, AW2001, AW2002, AW2003,...
  • Page 12: Device Malfunctions

    RIDA qLine® Orbital Shaker | User Manual 2.6 Device malfunctions Func- Potential Possible Possible Possible check Possible remedies tion error causes effects of individual functions Device does 1) Not Effects External visual Once user errors not shake. connected to on the inspection have been ruled power.
  • Page 13: Maintenance And Service

    3.1 Service and repairs If you have any other questions about installation, operation, or maintenance, contact R-Biopharm AG or your distributor. Do not make any electronic or mechanical repairs to the device. Inside the device, there are no parts that users can service or repair. User repairs can cause damage to the electronics.
  • Page 14: Uninstalling And Disposal

    RIDA qLine® Orbital Shaker | User Manual 4 Uninstalling and disposal Due to its properties and possible applications, the device needs to be handled properly. Follow the instructions below for handling the device: 4.1 Uninstalling the device When uninstalling/removing the device and its components, follow the steps below: 1.
  • Page 15: Packaging And Storing The Device

    center that accepts e-waste. The packaging material also needs to be disposed of in an environmentally friendly way (separation of materials). Devices sold in EU countries must be labeled appropriately. The WEEE Directive stipulates that customers and end users in EU countries may not dispose of electrical and electronic devices or electrical and electronic accessories in the household waste.
  • Page 16: Storing The Device

    RIDA qLine® Orbital Shaker | User Manual 5.2 Storing the device Follow the instructions below for proper storage of the shaker. 1. Package the shaker in a plastic bag. 2. Ensure the following ambient conditions are met: Max. ambient temperature: 5 °C - 40 °C Max.
  • Page 17: Technical Data

    Important: If the device fails, do not attempt to repair it. The device does not contain any parts that can be serviced or repaired by the user. User repairs will void the warranty. Important: Do not open the device. Repairs are reserved to trained service technicians. 7 Technical data Technical data: Standard conditions: 20 °C , 1013 mbar...
  • Page 18: Troubleshooting/Information On Reporting To Authorities

    For users in the European Union: Report all serious adverse events associated with the product to R-Biopharm AG and the appropriate national authorities. 9 Conformity to IEC 61326-2-6 and RoHS3 The CE mark on the RIDA qLine® Orbital Shaker confirms that the RIDA qLine® Orbital Shaker is compliant with Regulation (EU) 2017/746 (In Vitro Diagnostic Medical Devices).
  • Page 19: Symbols Used On The Device

    10 Symbols used on the device Symbol Description RIDA qLine® Orbital Shaker Product designation For in vitro diagnostic use Item number Serial number Important WEEE Waste Electrical and Electronic Equipment Follow user manual Manufacturer Date of manufacture Permissible storage temperature Permissible storage humidity 11 Version history Version number Chapter and description...
  • Page 20 RIDA qLine® Orbital Shaker | User Manual R-Biopharm • An der neuen Bergstrasse 17, 64297 Darmstadt, Germany Email: orders@r-biopharm.de • clinical.r-biopharm.com...