Mediana YM6000 Service Manual

Mediana YM6000 Service Manual

Patient monitor
Table of Contents

Advertisement

Quick Links

SERVICE MANUAL
YM6000 Patient Monitor
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
EU representative 
TECNOMED 2000 S.L. 
Valencia, 25 ‐ 28012 Madrid Spain 
 
Manufacturer 
Mediana Co., Ltd. 
Wonju Medical Industry Park, 1650‐1 Donghwa‐ri, 
Munmak‐eup, Wonju‐si, Gangwon‐do, Korea 
Tel: (82) 2 542 3375    (82) 33 742 5400   
Fax: (82) 2 542 7447    (82) 33 742‐5483 
 
YM6000 Service Manual 
Revised Date: 0607 
Part Number‐Revision: A7066‐1 
Printed in Korea 
Copyright © 2006‐2007 Mediana Co., Ltd. All rights reserved. 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the YM6000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Mediana YM6000

  • Page 1 SERVICE MANUAL YM6000 Patient Monitor                                 EU representative  TECNOMED 2000 S.L.  Valencia, 25 ‐ 28012 Madrid Spain    Manufacturer  Mediana Co., Ltd.  Wonju Medical Industry Park, 1650‐1 Donghwa‐ri,  Munmak‐eup, Wonju‐si, Gangwon‐do, Korea  Tel: (82) 2 542 3375    (82) 33 742 5400    Fax: (82) 2 542 7447    (82) 33 742‐5483    YM6000 Service Manual  Revised Date: 0607  Part Number‐Revision: A7066‐1 ...
  • Page 2   This document contains proprietary information that is protected by copyright. All Rights Reserved.  Reproduction,  adaptation, or  translation  without  prior  written  permission  is  prohibited,  except  as  allowed under the copyright laws.    Warranty     The information contained in this document is subject to change without notice.  Mediana makes no warranty of any kind with regard to this material, including, but not limited to,  the implied warranties or merchantability and fitness for a particular purpose.  Mediana shall not be liable for errors contained herein or for incidental or consequential damages in  connection with the furnishing, performance, or use of this material.    Revision History     The  documentation  part  number  and  revision  number  indicate  its  current  edition.  The  revision ...
  • Page 3: Table Of Contents

    Contents   Introduction ................................1 Warnings ................................. 1 Cautions ................................2 Manual Overview ............................3 Related Documents............................3 Intended Use for the YM6000 Monitor ....................... 3 Identifying the YM6000 monitor Configurations ..................3 Monitor Controls............................4 Routine Maintenance............................9 Cleaning ................................9 Periodic Safety and Functional Checks....................... 9 Batteries ................................. 10 Environmental Protection........................... 10 Performance Verification............................ 11 General ................................11 Required Equipment ...........................
  • Page 4 Figure 30. Process Unit Block Diagram ..........................84  Figure 31. User‐Control Unit Block Diagram.........................84  Figure 32. Sound Unit Block Diagram............................85  Figure 33. Communication Unit Block Diagram ........................85  Figure 34. GUI Unit Block Diagram............................85  Figure 35. Thermal Printer Unit Block Diagram ........................85  Figure 36. NIBP Unit Block Diagram.............................86  Figure 37. EtCO  Unit Block Diagram ...........................86  Figure 38. ECG Unit Block Diagram ............................86  Figure 39. Respiration Unit Block Diagram..........................87  Figure 40. SpO  Unit Block Diagram............................87  Figure 41. Temperature Unit Block Diagram ..........................87  Figure 42. IBP Unit Block Diagram ............................88  Figure 43. Analog Control Unit Block Diagram ........................88  Figure 44. Oxyhemoglobin Dissociation Curve........................92  Tables  Table 1. Required Equipment..............................11  Table 2. Parameter Alarm Limit Factory Defaults........................18  Table 3. Earth Leakage Current Values ............................31  Table 4. Enclosure Leakage Current............................32  Table 5. Patient Leakage Current Values ..........................33  Table 6. Patient Leakage Current Values ‐ Mains Voltage on Applied Part................33  Table 7. Test Lead Combinations..............................34  Table 8. Allowable Leakage Current............................34  Table 9. Service Menu................................37  YM6000 Service Manual...
  • Page 5 Table 10. Audible Alarm Characteristics ..........................38  Table 11. System Test ................................39  Table 12. NIBP Test .................................40  Table 13. Factory Default Settings ............................40  Table 14. Required Equipments for Firmware download ......................45  Table 15. Completion codes ..............................46  Table 16. Firmware Downloading Error Codes ........................54  Table 17. Alarm Messages and Check Items ..........................54  Table 18. Part Descriptions – Battery and Handle ........................64  Table 19. Part Descriptions – Monitor ............................65  Table 20. Part Descriptions – Side connector holder and Main board ..................66  Table 21. Part Descriptions – CPU module, SpO  module and IBP board ................67  Table 22. Part Descriptions –Bracket and Key board.......................68  Table 23. Part Descriptions – LCD and Bracket........................69  Table 24. Part Descriptions – Knob, LCD window, Encoder assembly and Connectors............70  Table 25. Part Descriptions – EtCO  module ..........................72  Table 26. Part Descriptions – Speaker .............................73  Table 27. Part Descriptions –SMPS case, SMPS and NIBP module ..................74  Table 28. Part Descriptions – Printer module and AC inlet....................75  Table 29. Parts List ..................................79  YM6000 Service Manual...
  • Page 6 Contents           This page is intentionally left blank.    YM6000 Service Manual...
  • Page 7: Introduction

    Performance  Tests  and  Safety  Tests  listed  in  Performance  Verification  section  of  this  service  manual  have  been  performed.  Failure to perform all tests could result in erroneous monitor readings.      WARNING:  High  voltage  is  generated  by  the  LCD  backlight  driver.  Exercise    caution when operating monitor with covers open.      YM6000 Service Manual...
  • Page 8: Cautions

    Operating  error  may  be  caused  if  the  cover  is  not  closed correctly.    CAUTION: If internal battery cable has been disconnected, pay particular attention  to polarity of the cable before reattaching. If battery cable polarity is reversed, it is  likely that circuit damage will occur.    CAUTION:  For  continued  protection  against  risk  of  fire,  replace  only  with  same  type and rating of fuse.    CAUTION: Ferrite Cores are used for electromagnetic compatibility. Please do not  remove Ferrite Cores while disassembling or reassembling, otherwise the monitor  can be affected by electromagnetic interference and measure inaccurate data to be  displayed or stored.        YM6000 Service Manual...
  • Page 9: Manual Overview

    Hospital‐type  facilities  include  physician  office‐based  facilities,  sleep  labs, skilled nursing facilities, surgicenters, and sub‐acutecenters.    Identifying the YM6000 monitor Configurations    The  following  table  identifies  YM6000  monitor  configurations  and  how  they  are  indicated. The model‐option number and serial number are located on the back of the  monitor.  All  information  in  this  manual,  including  the  illustrations,  is  based  on  a  monitor  configured  with  the  Capnography  (EtCO   and  InCO ), ...
  • Page 10: Monitor Controls

    6  AC indicator  7  Power button  8  Print button  9  Alarm silence/suspension button  10 Trend button  11 NIBP start/stop button  12 Return button  13 Trim knob  14 ECG connector  15 SpO  connector  16 NIBP connector  17 Temperature channel 1  18 Temperature channel 2  19 CO  connector (option)  20 IBP 1 connector (option)  21 IBP 2 connector (option)      Figure 1. Front Panel Controls and Connectors  YM6000 Service Manual...
  • Page 11: Figure 2. Rear Panel Components And Symbols

          Introduction 5   Rear Panel Components              1  Air ventilation  Battery cover  2  Equipotential terminal  Thermal Printer (option) 3  Ac power connector  Handle    4  Speaker              Figure 2. Rear Panel Components and Symbols    YM6000 Service Manual...
  • Page 12: Figure 3. Right Panel Components And Symbols

    6 Introduction Right Panel Components            1    LAN connector  2    RS‐232 interface connector   Figure 3. Right Panel Components and Symbols        YM6000 Service Manual...
  • Page 13: Figure 4. Left Panel Components And Symbols

          Introduction 7 Left Panel Components                        1  Thermal printer (option)   2 Battery cover     Figure 4. Left Panel Components and Symbols    YM6000 Service Manual...
  • Page 14 8 Introduction           This page is intentionally left blank.    YM6000 Service Manual...
  • Page 15: Routine Maintenance

    3. If  the  monitor  has  been  visibly  damaged  or  subjected  to  mechanical  shock  (for  example,  if dropped),  perform  the  performance  tests  as  described  in  Performance  Verification section. If the unit fails these performance tets, refer to Troublehooting  section.  4. Perform  the  electrical  safety  tests  detailed  in  Performance  Verification  section.  If  the unit fails these electrical safety tests, do not attempt to repair. Contact Mediana  Technical Service Department.  5. Inspect the fuses for proper value and rating.    qty 2, 6.3 A, 250 volts for AC mains    YM6000 Service Manual...
  • Page 16: Batteries

    Operation section of the operator’s manual.    Note: Storing the monitor for a long period without charging the battery may degrade  the  battery  capacity.  The  battery  may  require  a  full  charge/discharge  cycle  to  restore  normal  capacity.  Mediana  recommends  that  the  monitor’s  sealed,  Ni‐ MH  batteries  be  replaced  at  2  year  intervals.  Refer  to  Disassembly  Guide  section.    CAUTION:  If  the  monitor  is  to  be  stored  for  a  period  of  2  months  or  longer,  it  is ...
  • Page 17: Performance Verification

    Sample elbow or tee   gas flow meter  BCI catalogue No. 8133 recommended  Sample line  Sample line  Water trap  Water trap   scrubber     scrubber    (or 0% CO  medical‐grade gas source) (or 0% CO  medical‐grade gas source)  Safety analyzer  METRON QA‐90 or equivalent  Data interface cable  9‐pin RS‐232 cable  PC  Common PC    LAN cable  Common used LAN cable  Stopwatch  Manual or electronic  Note:  Contact  Mediana  Technical  Service  Department  for  pricing  and  ordering  information. YM6000 Service Manual...
  • Page 18: System Tests

    2. The  level  of  the  alarm  tone  will  appear  on  the  screen  as  the  alarm  tone  sounds.  Then, the level goes up gradually. When the tone reaches the maximum level 8, it  returns to the minimum level 1.  3. After  finishing  the  test,  press  the  trim  knob  to  select  Cancel.  The  menu  box  will  disappear.    Pass/Fail Results    When the alarm tone changes to eight steps, the alarm tone is in a normal state.    YM6000 Service Manual...
  • Page 19: Figure 6. Tone Audible Test

    When the tone returns to the minimum, the pitch will change automatically.  There are three pitches – High, Med and Low.  3. Press Cancel to finish the test.  4. Rotate the trim knob to select Key Beep.  The key beep sounds intermittently as the level goes up gradually. When the tone  reaches the maximum level 7, it returns to the minimum 1.  5. After  finishing  the  test,  press  the  trim  knob  to  select  Cancel.  The  menu  box  will  disappear.      Pass/Fail Results      When the HR/PR tone changes to seven steps and the pitch changes to three steps, the  HR/PR tone is normally set.  When the key beep changes to seven steps, the key beep is normally set.  When the completion sound is heard, it is normally set.      YM6000 Service Manual...
  • Page 20: Figure 7. Backup Ram Clear

    14 Performance Verification Backup RAM Clear    When Backup RAM Clear is set to Yes, all settings of the monitor including the service  settings return to the factory defaults from the next cycle.      Figure 7. Backup RAM Clear    Note:  Set  the  monitor  to  appropriate  AC  line  frequency  (50Hz  or  60Hz)  again  after  Back Up RAM clearance.  YM6000 Service Manual...
  • Page 21: Performance Tests

    8. Set NIBP Automatic mode interval of the monitor to 15 minutes.    9. The  monitor  must  operate  for  1  hour  with  one  fully  charged  battery.  The  monitor  must  operate  for  at  least  15  minutes  before  the  monitor  powers  down  due  to  low  battery condition.    10. Verify that low priority audible alarm occurs and a warning message “low battery”  is displayed about 15 minutes before battery fully discharges.  YM6000 Service Manual...
  • Page 22 2. Observe  the  monitor’s  front  panel.  With  monitor  off,  press  Power  button.  The  monitor must perform the following sequence.  The monitor progresses the check sum for the flash memory and displays  the bar while the check sum is proceeding  When  the  check  sum  completed,  the  bar  is  filled,  and  then  Mediana  logo  appears for  a  few  seconds,  with  version  numbers  of the  software and  the  copyright in lower left corner of display.  The monitor emits the high‐pitched beep. ...
  • Page 23   HR/PR Tone Volume Control  1. Press Power button to turn on the monitor.  2. Connect SpO  simulator to sensor input cable and connect cable to monitor.  3. Set SpO  simulator as follows: SpO  of 75% and pulse rate of 60bpm.  4. Verify SpO  and pulse rate values are correctly displayed.  5. Press Alarm silence/suspension button on  front  panel  of  the  monitor  to  temporarily  silence audible alarm.  6. Verify HR/PR source on HR/PR numerical area is set to “SpO ”.  7. Select Setup icon on the screen to display Setup menu.  8. Rotate the trim knob to highlight HR/PR tone volume on Setup menu and press the  trim knob to adjust HR/PR tone volume.  9. Set HR/PR tone volume level 1 to level 7 and return to the monitoring screen. Verify  beeping pulse rate tone increases.  10. Set HR/PR tone volume level 7 to level 1 and return to the monitoring screen. Verify  beeping pulse rate tone decreases.  YM6000 Service Manual...
  • Page 24: Table 2. Parameter Alarm Limit Factory Defaults

     High Alarm Limits  100 %  100 %  95 %  % SpO  Low Alarm Limits  90 %  90 %  80 %  IBP1, 2 SYS High Alarm Limits  160 mmHg  120 mmHg  90 mmHg  IBP1, 2 SYS Low Alarm Limits  90 mmHg  70 mmHg  40 mmHg  IBP1, 2 DIA High Alarm Limits  90 mmHg  70 mmHg  60 mmHg  IBP1, 2 DIA Low Alarm Limits  50 mmHg  40 mmHg  20 mmHg  IBP1, 2 MEAN High Alarm Limits 110 mmHg  90 mmHg  70 mmHg  IBP1, 2 MEAN Low Alarm Limits  60 mmHg  50 mmHg  30 mmHg  RESP High Alarm Limits  30 BPM  30 BPM  100 BPM  YM6000 Service Manual...
  • Page 25 2. Press Power button to turn on the monitor.  3. Connect all necessary simulators to the monitor.  4. Select Setup icon on the screen to display Setup menu.  5. Test #1: 20 sec printing  Set Wave record time to 20 sec via Setup menu.  Press Print button when all the parameter signals display normally.  Verify  that  the  parameter  values  and  waveforms  are  printed  out  for  20  seconds.  6. Test #2: Continuous printing  Set Wave record time to Continuous via Setup menu.  Press Print button when all the parameter signals display normally.  Verify  the  parameter  values  and  waveforms  are  printed  out  continuously.  Verify printing stops with pressing Print button again.  YM6000 Service Manual...
  • Page 26   Network Test  Perform the following procedure to test the Network. The Network connector is located  on the monitor’s right panel, identified with the Network symbol.    CAUTION: Do not change any other settings of the test programs while performing  the Network test.    Note:  Network  Test  is  only  available  when  YM6000 monitor  is  equipped  with TCP/IP  module for Central Monitoring System (CMS). Otherwise this test does not need  to perform.    Note: Contact Mediana Technical Service Department for the software required.    1. Connect the network line to the monitor, then turn on the monitor.  2. Run  “YM6000_Comm_Test.exe”  on  PC  connected  the  network  line  using  the  same  gateway as the monitor. ...
  • Page 27 5  and  1  must  be  between  ‐5  to  ‐12  VDC.  This  verifies  the  RS‐232  Nurse Call function.  7. With the instrument in an alarm condition, use a digital multmeter (DMM) to verify  that there is no continuity (1 mega ohms or greater) between pins 8 and 9 and that  there is continuity (60 ohms or less) between pins 7 and 8.  8. Press the SRC‐MAX simulator %SpO  button to change the %SpO  to 90.  9. Use a DMM to verify that there is continuity between pins 8 and 9 and that there is  no continuity between pins 7 and 8. This verifies the solid state Nurse Call function.    Note:  The  pin  layouts and  signal  descriptions  are included in  the  operator’s manual.  For  the  detailed  information  regarding  the  Nurse  Call,  refer  to  Nurse  Call  Interface section of the operator’s manual.  YM6000 Service Manual...
  • Page 28 HR/PR  numerical  area,  and  that  low  priority audible alarm sounds.  Reconnect the LL lead to ECG simulator. Verify that “ECG: Check ECG leads  &  electrodes”  message  no  longer  appears  and  that  the  audible  alarm  is  stopped.  Repeat this test for the LA and the RA leads.  10. Connect all the leads to the monitor.  Select ECG waveform menu and set Lead select to Lead I.  Verify that the lead selection is displayed on the ECG waveform area.  Repeat step 10‐a for all the ECG Lead selections.  11. Set ECG Lead selection to Lead II.  12. Change ECG waveform size to all the selectable sizes and verify that an appropriate  size is displayed on the ECG waveform area.  YM6000 Service Manual...
  • Page 29 These  tests  verify  the  functionality  of  the  monitor’s  pneumatic  system.  The  monitor  must be placed in Service menu. For a detailed explanation to access the service menu,  refer to Service Menu and Factory Default Settings section.    1. Rotate the trim knob to select NIBP Test in the service menu, and then press the trim  knob.    Note: Before accessing the NIBP Test mode, ensure that current patient mode is proper  for the pneumatic system to be test. You can set Patient mode: Adult, Pediatric or  Neonatal via the setup menu.    Note:  In  the  NIBP  Test  mode,  no  function  button  will  have  no  effect  except  the  trim  knob. All the tests will start to be performed by pressing or rotating the trim knob.  If you would like to stop the test during test progressing, press the trim knob.    Note: Inflation Speed Test and Deflation Speed Test are intended for factory use only.    YM6000 Service Manual...
  • Page 30: Figure 8. Pressure Sensor Accuracy Test

    8. After  finishing  the  test,  press  the  trim  knob  to  select  Cancel.  The  menu  box  will  disappear.  If Cancel is selected  during  the  test  progressing,  the  test  will stop  and  the menu box will disappear.    Pass/Fail Results    The difference between the simulator’s and the monitor’s readings are as follows:    Sphygmomanometer  Monitor’s Readings  0 mmHg  ±0 mmHg    50 mmHg  50±6 mmHg  100 mmHg  100±6 mmHg  200 mmHg  200±6 mmHg    YM6000 Service Manual...
  • Page 31: Figure 9. Air Leakage Test

    4. Rotate the trim knob to select Air Leakage Test, and then press the trim knob.  5. The result will be displayed on the monitor’s screen in four minutes.  6. After  finishing  the  test,  press  the  trim  knob  to  select  Cancel.  The  menu  box  will  disappear.  If Cancel is selected  during  the  test  progressing,  the  test  will stop  and  the menu box will disappear.    Pass/Fail Results    It passes if the leakage value is less than 12mmHg/3minutes.    YM6000 Service Manual...
  • Page 32 The  monitor  will  decrease  to  60  and  stabilize  at  60bpm.  The  test  pass  criteria is 57 to 63 bpm.  The monitor will display:  ‐ 90 % SpO   ‐ 60 bpm  ‐ no alarm  ‐ low level modulation  8. Test #3: Modulation Level  Press  %MODULATION  selection  button  on  the  SpO   simulator.  The %MODULATION LED on the SpO  simulator will light.  The  monitor’s  waveform  area  will  spike  and  stabilizes  at  a  higher  modulation level.  The monitor will display:  ‐ 90 % SpO   YM6000 Service Manual...
  • Page 33 2. Connect  the  temperature  probe  (supplied  with  the  temperature  simulator)  to  an  appropriate jack on temperature simulator.  3. Connect the temperature probe to the temperature connector on the monitor’s front  panel.  4. Set Temperature simulator as follows:  Temperature: 37°C (98.0°F)  Probe type: YSI‐400 series Temperature Probes (Probe accuracy: ±0.1°C)  5. After the normal power‐up sequence, verify that the temperature reads 37°C ±0.1°C  (98.6°F ±0.2°F if Fahrenheit is selected for the temperature units).  6. Turn off the monitor.    Note:  The  accuracy  of  temperature  measurements  is  ±0.1°C  (±0.2°F)  in  the  range  of  25°C to 45°C and ±0.2°C in the range of 15° C to less than 25° C. In the procedure  above, add the tolerance of the simulator and the probe to the acceptable range of  readings.    YM6000 Service Manual...
  • Page 34 9. Press  the  5,  wavef  button  on  the  IBP  simulator,  then  press  enter  button  on  the  IBP  simulator.  10. The monitor will:  ‐ display IBP waveform on the screen.  ‐ display systolic, mean, diastolic measurement values on the IBP numerical area.    Note:  Initial  values  of  IBP  simulator  are  systolic  120mmHg,  diastolic  80mmHg  at  channel 1, systolic 120mmHg, diastolic 0mmHg at channel 2.      YM6000 Service Manual...
  • Page 35 After approximately one minute, observe the CO  reading. The CO  reading  should be between 0.0%~0.3% with well‐ventilated room air.  Proceed  with  the  zero  gas  phase  of  the  calibration  as  defined  by  the  host  system.  Disconnect  the  CO   scrubber  (or  gas  source)  from  the  water  trap  after  the  zero gas calibration.    YM6000 Service Manual...
  • Page 36 Make  sure  the  bleed  line  is  directed  away  from  the monitor. Allow the reference gas to flow for at least one minute.  h. Sample the Span Reference Gas.  After  the  Two‐Point  User  Calibration  steps  have  been  followed,  disconnect  the monitor from the gas supply and test setup.  A verification may be performed using the same gas delivery set up. Verify  the observed CO  gas reading is within 3 mmHg or 10%, whichever is greater,  of  the  CO   value  of  the  test  gas  supplied.  (If  not,  refer  the  user  to  the  appropriate troubleshooting information.)    YM6000 Service Manual...
  • Page 37: Safety Tests

    This  test  is  in  compliance  with  IEC60601‐1  enclosure  leakage  current.  This  test  is  for  ungrounded  enclosure  current,  measured  between  enclosure  parts  and  earth.  The  applied voltage for IEC60601‐1 the applied voltage is 264 Volts AC at 50 to 60 Hz.    1. Connect the monitor AC plug to the electrical safety analyzer as recommended by the  analyzer operating instructions.  2. Place a 200cm  foil in contact with the instrument case making sure the foil is not in  contact with any metal parts of the enclosure that may be grounded.  3. Measure the leakage current between the foil and earth.    Note: The analyzer leakage current indication must not exceed the values listed in Table  4.  YM6000 Service Manual...
  • Page 38: Table 4. Enclosure Leakage Current

    Note:  This  test  requires a test  cable for  each  patient  connector.  For  example,  the  ECG  test  cable  consists  of  the  ECG  cable  connector,  with  all  conductors  shorted  together, connected to a test lead from the electrical safety analyzer. Test cables  for  SpO   and  temperature  can  be  configured  in  a  similar  manner,  by  wrapping  each sensor end individually with aluminum foil filled with conductive gel (only  enough gel to ensure conductivity). Attach a wire to the foil that is connected to a  test lead from the electrical safety analyzer.    YM6000 Service Manual...
  • Page 39: Table 5. Patient Leakage Current Values

    2. Connect  monitor’s  AC  mains  power  cord  to  analyzer  as  recommended  by  analyzer  operating instructions.  3. Connect  ECG  test  cable  between  ECG  connector  on  the  monitor  and  appropriate  input connector on analyzer.  4. Turn on the monitor.  5. Perform the test as recommended by analyzer operating instructions.  6. Repeat the test for SpO  and temperature patient connections, using appropriate test  cables.    Note: Patient leakage current is measured with normal and reverse mains polarity. For  each condition, the measured leakage current must not exceed that indicated in  Table 6.    Table 6. Patient Leakage Current Values ‐ Mains Voltage on Applied Part  Test Condition  Allowable Leakage Current (microamps)  Normal polarity (SFC)  50  Reverse polarity (SFCRM)  50  YM6000 Service Manual...
  • Page 40: Table 7. Test Lead Combinations

    ECG #1 (LA)  ECG #2 (LL)  ECG #1 (LA)  ECG # 3 (RA)  ECG #2 (LL)  ECG #3 (RA)  ECG #1 (LA)  Temperature 1/2  ECG #2 (LL)  Temperature 1/2  ECG #3 (RA)  Temperature 1/2  ECG #1 (LA)    ECG #2 (LL)    ECG #3 (RA)    Temperature 1/2      Table 8. Allowable Leakage Current  Test Condition  Allowable Leakage Current (microamps)  Normal Condition (NC)  10  SFC Open Supply (OS)  50  SFC Open Earth (SFC OE)  50  Normal Condition RM (NCRM)  10  SFC Open Supply RM (SFC OSRM)  50  SRC Open Earth RM (SFC OERM)  50    YM6000 Service Manual...
  • Page 41: Service Menu And Factory Default Settings

    Once  the  Service  menu  is  entered,  physiological  monitoring  is  terminated.  The  screen  layouts  do  not  display  any  information  associated  with  normal  monitoring  operation. Use the following procedure to configure the Service Menu for the monitor  (also see Using the Monitor section, of the operator’s manual):      Figure 10. The access of Service Menu via Set‐up menu    1. Set the monitor to normal monitoring mode.  2. Rotate the trim knob to highlight the Setup icon located on the bottom of the screen,  and then press the trim knob. Setup menu displays.  3. Rotate the trim knob to highlight Service Menu in Setup menu, and then press the  trim knob to access the Service Menu.  4. Three digits are displayed in the Level 2 Menu as shown in Figure 4.  YM6000 Service Manual...
  • Page 42: Figure 11. The Access Of Service Menu Via Set-Up Menu

    8. The  Service  Menu  will  now  be  present.  The  available  Service  Menu  items  are  explained in Figure 6 and Table 9. Make changes to these menu items as desired  by rotating and pressing the trim knob.    9. Select  “Done”.  The  monitor  will  present  the  message  “All  changes  made  to  the  power‐up defaults will be in effect the next time the monitor is turned on.” Before  turned off.  10. Turn off the monitor, and then turn on the monitor again.    Note: The monitor must be powered off upon selecting “Done” to save any changes  into the monitor, and then the changes made to the Power on defaults will be in  effect next time the monitor is powered up.    YM6000 Service Manual...
  • Page 43: Table 9. Service Menu

    System Test  LCD Test    Alarm Audible Test    Tone Audible Test    Backup RAM Clear  NIBP Test  Pressure Sensor Accuracy Test    Air Leakage Test    Inflation Speed Test    Deflations Speed Test  Demo Mode  On, Off  EtCO  Calibration  Zero Calibration, Calibration  Done  The monitor will be powered off upon selecting  “Done”, then any changes in the service menu will  be in effect next time the unit is powered up.    Saved setting on Power off  If the save settings on power off is set to Custom, the monitor does not save the  settings when the monitor is power off. Therefore, the first settings when the monitor  is powered up become the power‐up defaults in the next power‐on cycling.    If the save settings on power off is set to Back Up, the monitor saves the current  settings when the monitor is powered off. The saved settings become the power‐on  defaults in the next power‐on cycling.  If the save settings on power off is set to Default, the factory default settings become  the power‐up defaults in the next power‐on cycling.    YM6000 Service Manual...
  • Page 44 IEC60601‐1‐8  High priority  ~976 Hz  ~976 Hz  7 beeps in 2 sec  10 beeps in 15 sec  Medium priority  ~697 Hz  ~697 Hz  2 beeps in 1 sec  3 beeps in 15 sec  Low priority  ~488 Hz  ~488 Hz  1 beeps in 5 sec  1 beeps in 30 sec    AC Line Frequency  The monitor supports AC line frequency both 50 Hz and 60 Hz. Select either 50Hz or  60 Hz for an appropriate AC line.    Unit Configuration  The unit configuration of each parameter can be set in the service menu.      Language Setting    The selected language will be used for all the text shown on the display; and it will be  effective the next time the monitor is powered up.    Date Format  The date format is selectable; Year/Month/Day, Month/Day/Year, Day/Month/Year    Alarm Reminder Tone  The monitor generates the alarm reminder tone at the preset interval to remind the  user that the audible alarm is suspended. The interval can be set to Off, 3 or 10  minutes via the service menu. If Off is selected, the reminder tone will be disabled.  YM6000 Service Manual...
  • Page 45 Discharges  may  not  be  reset,  but  they  will  be  reset  to  zero  when  a  new  CPU  module is installed.    System Test    These  menus  provide  to  facilitate  performing  verification  testing  for  the  overall  system.      For a detailed procedure, refer to Performance Verification section.    Table 11. System Test  Tests  Description  LCD Test  tests the LCD display.  Alarm Audible Test  tests the alarm tones.  Tone audible Test  tests the HR/PR tones, the key beeps and the completion  sounds.  Backup RAM Clear  is used to set the factory defaults.    YM6000 Service Manual...
  • Page 46: Demo Mode

    80, 100, 120, 140mmHg (13.3, 16.0, 17.3,  18.7 kPa)  BP On Alarm  On, Off  Off  Off  Off  NIBP Interval  Off, Cont, 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45,  Off  Off  Off  60, 90,120 ,180 minutes  NIBP SYS High Alarm Limits  adult/pediatric  160 mmHg  120 mmHg  90 mmHg  35 to 270 mmHg (4.7 to 36.0 kPa)  (21.3 kPa)  (16.0 kPa)  (12.0 kPa)  neonatal  45 to 130 mmHg (6.0 to 17.3 kPa)  (5 mmHg, 0.7kPa steps)  adult/pediatric  NIBP SYS Low Alarm Limits  90 mmHg  70 mmHg  40 mmHg  30 to 265 mmHg (4.0 to 36.3 kPa)  (12.0 kPa)  (9.3 kPa)  (5.3 kPa)  neonatal  40 to 125 mmHg (5.3 to 16.7 kPa)  YM6000 Service Manual...
  • Page 47 IBP 2 DIA High Alarm Limits  90 mmHg  70 mmHg  60 mmHg  ‐45 to 300 mmHg (‐6 to 40 kPa)  (12.0 kPa)  (9.3 kPa)  (8.0 kPa)  (5 mmHg, 0.7kPa steps)  adult/pediatric/neonatal  IBP 2 DIA Low Alarm Limits  50 mmHg  40 mmHg  20 mmHg  ‐50 to 295mmHg (‐6.7 to 39.3 kPa)    (6.7 kPa)  (5.3 kPa)  (2.7 kPa)  (5 mmHg, 0.7kPa steps)  adult/pediatric/neonatal  IBP 2 MEAN High Alarm  110 mmHg  90 mmHg  70 mmHg  ‐45 to 300 mmHg (‐6 to 40 kPa)  Limits  (14.7 kPa)  (12.0 kPa)  (9.3 kPa)  (5 mmHg, 0.7kPa steps)    adult/pediatric/neonatal  IBP 2 MEAN Low Alarm  60 mmHg  50 mmHg  30 mmHg  YM6000 Service Manual...
  • Page 48 0 to 2.4 kPa (Adult/Neo) (0.3 kPa steps)  0 %  0 %  0 %  0 to 2.3 % (Adult/Neo) (0.3 % steps)  Temperature          TEMP1 High Alarm Limits  15.1 to 45.0 °C (0.1°C steps)  39.0 °C  39.0 °C  39.0 °C  59.1 to 113.0 °F (0.1°F steps)  (102.2 °F)  (102.2 °F)  (102.2 °F)  TEMP1 Low Alarm Limits  15.0 to 44.9 °C (0.1°C steps)  36.0 °C  36.0 °C  36.0 °C  59.0 to 112.8 °F (0.1°F steps)  (96.8 °F)  (96.8 °F)  (96.8 °F)  TEMP2 High Alarm Limits  15.1 to 45.0 °C (0.1°C steps)  39.0 °C  39.0 °C  39.0 °C  59.1 to 113.0 °F (0.1°F steps)  (102.2 °F)  (102.2 °F)  (102.2 °F)  YM6000 Service Manual...
  • Page 49 한국어  (Korean),  中文  (Chinese),  English, Français (French), Deutsch  (German), Italiano (Italian),  日本語 Language*  English  (Japanese), Português (Portuguese),  Español (Spanish)  :  Note An asterisk (*) by a parameter in the above table indicates that the parameter can only be changed by authorized personnel  as described in the monitor service manual  Note:  Asterisks  (**)  by  a  parameter  in  the  above  table  indicate  the  print  settings  when  the  optional  printer  is  installed  in  the    monitor   YM6000 Service Manual...
  • Page 50 44 Service Menu and Factory Default Settings             This page is intentionally left blank. YM6000 Service Manual...
  • Page 51: Firmware Download

    1. Turn the monitor off.  2. Turn the monitor on by pressing the Power button.  3. The  monitor  displays  the  bar  while  the  checksum  for  the  flash  memory  is  in  progress.  4. Press  NIBP  start/stop  button  and  Return  button  simultaneously  before  filled  the  bar completely.  5. The monitor will display the firmware download screen as shown below figure 13.    Figure 13. Firmware downloading display  YM6000 Service Manual...
  • Page 52: Figure 14. Firmware Downloading Display

      Table 15. Completion codes  Code  Description  S. 1. 3  Boot download completed.  S. 2. 3  Main download completed.    Note: If any problem during Firmware downloading, the error code will be displayed  in the box located on the center of the screen. Refer to Firmware Download in  Troubleshooting section.    10. After completion of downloading, turn the monitor off.  11. Disconnect the LAN cable from the monitor.  12. After a few seconds, turn the monitor on again.  13. Perform the tests specified in Performance Verification section.    Note:  When  a  new  firmware  downloading  is  completed,  the  monitor  is  set  to  the  factory default.  YM6000 Service Manual...
  • Page 53: Troubleshooting

    If  the  trouble  symptom  persists,  swap  back  the  replacement  PCB  assembly  with  the  suspected  malfunctioning  PCB  assembly  (the  original  PCB  assembly  that  was  installed when you started troubleshooting) and continue troubleshooting as directed  in this section.    Obtaining Replacement Parts  Mediana Technical Service Department provides technical assistance information and  replacement  parts.  To  obtain  replacement  parts,  contact  Mediana.  Refer  to  the  part  names and part numbers listed in Spare Parts section.  YM6000 Service Manual...
  • Page 54: Troubleshooting Guide

    1. When  AC  power  cord  is  connected  to  the  monitor,  AC  indicator  on  the  front  panel is not lit.  1‐1 Ensure power cord is plugged into the appropriate AC outlet of voltage and  frequency.  1‐2 Replace the power supply assembly.  1‐3 Ensure the key wire is connected to the key board assembly and the main  board. If the connection is good, replace the key board assembly.    2. The monitor fails to power‐up when the power button is pressed.  2‐1 Ensure power cord is plugged into operational AC outlet of the appropriate  voltage and frequency. Ensure AC power indicator is lit. If indicator is not lit,  replace the power supply assembly.  2‐2 Check fuses located on the power supply assembly above AC inlet receptacle.  Replace fuses if necessary.  2‐3 Inside monitor, check the main cable and ensure that it is connected to the  main board and the power supply assembly.  2‐4 Ensure the key board assembly is connected to the main board. If connection  is good, replace the key board assembly.  2‐5 If the problem persists, replace the main board.    3.   The monitor turns on, then sounds an alarm and shuts off and no error code is  displayed.  3‐1 Replace the main board.  YM6000 Service Manual...
  • Page 55 If  the  action  requires  replacement  of  a  PCB  assembly  or  module, refer to Disassembly Guide.      1. LCD screen is totally black after system powers‐up.  1‐1 Ensure the inverter wire is connected to the LCD assembly.  1‐2 Ensure the LCD cable is connected to the LCD assembly and the main board.  1‐3 If the problem persists, replace LCD assembly.  1‐4 If the problem persists, replace the main board.    2. LCD screen is illuminated, but no data is visible after system powers‐up.  2‐1 Ensure LCD cable is connected to the main board and LCD assembly.  2‐2 If the problem persists, replace the LCD cable.  2‐3 If the problem persists, replace the main board.  2‐4 If the problem persists, replace the LCD assembly.    3. Display values are missing or erratic.  3‐1 If the measurement cables are connected, ensure that the cables are properly  connected.  3‐2 If the problem persists, replace the measurement cables.  3‐3 If the problem persists, replace LCD cable.  3‐4 If the problem persists, replace the main board.  YM6000 Service Manual...
  • Page 56 1‐1 Moisture or spilt liquids can cause an alarm not to sound. Allow the monitor  to dry thoroughly before using.  1‐2 If the problem persists, replace the main board.    2. Audible alarm does not sound.  2‐1 Check alarm silence status.  2‐2 Check alarm volume setting in Setup menu.  2‐3 Ensure the speaker is connected to the power supply assembly.  2‐4 If problem persists, replace speaker.  2‐5 Check the main cable between main PCB and the power supply assembly.  2‐6 If problem persists, replace the power supply assembly.  2‐7 If problem persists, replace the main board.    3. The monitor responds to button press, but key press tone fails to sound.  3‐1 Check the key beep volume setting in Setup menu.  3‐2 Ensure speaker is connected to main the power supply assembly.  3‐3 If problem persists, replace speaker.  3‐4 Check the main cable between the main board and the power supply  assembly.  3‐5 If problem persists, replace the power supply assembly.  3‐6 If problem persists, replace the main board.  YM6000 Service Manual...
  • Page 57 The  following  are  symptoms  of  problems  relating  to  NIBP  functions  and  recommended actions. If the action requires replacement of a PCB or module, refer to  Disassembly Guide.    1. The cuff does not inflate.  1‐1 Ensure the cuff hose is not folded and NIBP tube inside of the monitor is not  blocked or kinked.  1‐2 If problem persists, replace the NIBP module.  1‐3 If problem persists, replace the main board.    2. Alarm Message C11 or E03  2‐1 Ensure the cuff hose is not folded and NIBP tube inside of the monitor is not  blocked or kinked.  2‐2 If problem persists, replace the NIBP module.  2‐3 If problem persists, replace the main board.    3. The NIBP value is not reliable.  3‐1 Perform the pressure sensor accuracy test and air leakage test.  3‐2 If problem persists, replace the NIBP module.  YM6000 Service Manual...
  • Page 58   Respiration    The  following  are  symptoms  of  problems  relating  to  Respiration  functions  and  recommended actions. If the action requires replacement of a PCB or module, refer to  Disassembly Guide.    1. Respiration signal (imRR) is noisy or the value is not displayed.  1‐1 Ensure ECG accessories are connected properly or the electrode is not  contaminated.  1‐2 If problem persists, replace the ECG board.  1‐3 If problem persists, replace the main board.    2. The monitor does not measure the respiration (awRR).  2‐1 Ensure the hose inside the monitor  2‐2 If problem persists, replace the EtCO  module    2‐3 If problem persists, replace the main board.  YM6000 Service Manual...
  • Page 59 Guide.    1. The EtCO  is not displayed at LCD  1‐1 Please  remind  that  it  takes  3  minutes  to  warm  up  the  EtCO   module  after  turning on the monitor.  1‐2 Please go to the Main screen of setup icon and set the display window to 6‐ch  Wave.  1‐3 Check the accessory is properly connected to the monitor and measurement  circuit.  1‐4 If the problem persists, replace the EtCO  accessories.  1‐5 If the problem persists, replace the EtCO  module.      YM6000 Service Manual...
  • Page 60 Table 17. Alarm Messages and Check Items Alarm Messages  Check Items  NIBP: Check cuff (C11)  Cuff pressure did not increase enough even after activating the  pump for more than 30 seconds (adult). There is a possibility that  a cuff hose is disconnected, or a cuff may not be wrapped around  an arm. Check cuff and cuff hose.  This error possibly occurs in case of large cuffs that are wrapped  around loosely.  When the error still occurs even after checking above, there is a  possible air leakage from a ruptured cuff.  Replace it with a new one.  NIBP: Check cuff / Patient  Blood pressure could not be measured even after cuff pressure  (C12)  decreased. It is possibly because pulse was not strong enough for  measurement, or because change of pulse amplitude could not  be obtained. Check whether cuffs are not wrapped around thick  clothing. After wrapping cuffs around property, measure again.  When the error occurs in the initial measurement in continuous  mode, the second measurement will start unless Stop button is  pressed.  NIBP: Cuff excessive  Measurement failed because of patient movement during  artifact (C13)  measurement. Tell the patient to stay still, then, measure again.  When it occurs in the initial measurement in continuous mode,  the second measurement will start unless Stop button is pressed. NIBP: Cuff insufficient  Measurement failed because of insufficient pressurizing. There is  YM6000 Service Manual...
  • Page 61 Motion artifact or external vibrations possibly affected cuffs.  Check for patient movement and if the cuff is free from stays still  and cuffs are free from outside vibration, then, measure again.  NIBP: Cuff pressure  Cuff pressure exceeded more than 300 mmHg (adult) during  failure (C19)  measurement.  There is a possibility that the patient moved during  measurement or strong pressure from outside might be added to  cuffs. Considering above, measure again.  NIBP: Cuff weak pulse  Amplitude of pulse obtained from cuffs are too weak.  (C20)  This error possibly occurs when cuffs are wrapped around  loosely in ASO patients or when cuffs are wrapped around thick  clothing.  Wrap cuffs around properly, then, measure again.  NIBP: Check cuff, hose  Patient to be measured, and cuff size used, do not match.  and mode (C21)  This error may occur if the blood pressure measurement mode  setting is incorrect, if the cuff has been wrapped tightly in the  adult mode, loosely in the neonatal mode or if the arm has been  bent during measurement.  Check the measurement mode setting and application of the cuff,  and measure again.  NIBP: Internal error (E03)  NIBP module error  BPM pressure sensor fault.  Pump operated for ten seconds, however pressure does not  change. Check the connection of the cuff hose.  NIBP: Internal error (E07)  Reboot the monitor. If the problem persists, cease use  immediately and contact qualified service personnel or your  local supplier.  YM6000 Service Manual...
  • Page 62   sensor, or blood flow at the sensor site may be unsatisfactory.  Check the condition of the patient and fitting of the sensor, and  measure again.  : Internal error.  A problem with the SpO  measurement has been detected. The   measurement function does not operate. If switching  power OFF/ON has no effect it is possible that a fault has  occurred. Stop using the monitor immediately and contact  qualified service personnel or your local supplier.   TEMP{n}: Internal error.  An internal circuit fault has been detected. If switching power  OFF/ON has no effect it is possible that a fault has occurred in  the monitor. Cease use immediately.  CAPNO: Internal error.  A problem with the capnography measurement function has  been detected. The capnography measurement function does not  operate. If switching power OFF/ON has no effect it is possible  that a fault has occurred. Cease use immediately.  CAPNO: Sensor error.  The connector may be damaged or a fault may have occurred  within the gas unit or main unit. Cease use immediately.  SYSTEM: Critically low‐ Connect the AC power cord of the monitor to the AC main to  battery condition.  recharge the battery.  SYSTEM: Real time clock  Reboot the monitor. If the problem persists, cease use  error.  immediately and contact qualified service personnel or your  local supplier.  SYSTEM: WDT error.  Reboot the monitor. If the problem persists, cease use  immediately and contact qualified service personnel or your  YM6000 Service Manual...
  • Page 63 Check whether electrodes are correctly attached and electrodes  are new and wet. Confirm the patient’s skin is clean.  {label}: Unable to zero  Could not zero calibrate pressure. Check that the transducer is  calibration.  open to the atmosphere, and check the three‐way tap. As it is  also possible that the measured pressure incorporates noise,  check the measurement circuit.  {label}: Out of range.  A value outside the measurement range was obtained. As the  transducer has been subject to abnormal pressure, check the  measurement circuit.  : Check sensor.  Sensor not connected. If connected, the cable or connector may  be damaged. Replace with a new cable. If replacing the cable has  no effect the problem may be within the device. In this case,  cease use immediately.  : Sensor failure.  A problem with the SpO  sensor has been detected. The SpO   measurement function does not operate. The possible cause is a  connection failure of the SpO  sensor and the extension cable, or  a failure of the sensor or cable. Reconnect the sensor and  extension cable or replace them with new ones. If the problem  doesn’t clear up after carrying out the remedies above or  switching power OFF/ON, a grave fault can develop, Cease the  use of the sensor immediately.  : Module reset.  A problem with the SpO  measurement has been detected. The   measurement function does not operate. If switching  power OFF/ON has no effect it is possible that a fault has  occurred. Cease use immediately.  TEMP{n}: Out of range.  A measure reading outside the measurement range was  YM6000 Service Manual...
  • Page 64 Check whether cuffs are not wrapped around thick clothing,  whether the patient is moving and whether there is external  vibration. When it occurs in the initial measurement in  continuous mode, the second measurement will start unless Stop  button is pressed.  NIBP: Retry, Cuff  Blood pressure could not be measured because oscillation graph  irregular pulses (C15)  was not normal. There is a possibility that patient movement or  external vibration interrupted the measurement. Check whether  a patient stays still and cuffs are free from outside vibrations,  then, measure again. When it occurs in the initial measurement  in continuous mode, the second measurement will start unless  Stop button is pressed.  NIBP: Retry, Cuff motion  Blood pressure could not be measured because noises  artifact (C16)  interrupted pulse waveform signal. There is a possibility that  patient movement or external vibration interrupted the  measurement. Confirm the patient is not moving and the cuff is  free of external vibration, then, measure again. When it occurs in  the initial measurement in continuous mode, the second  measurement will start unless Stop button is pressed.  NIBP: Retry, Cuff time‐ Pulse waveform signal more than 160 beats are detected during  out, over 160 pulses (C18)  measurement. There is a possibility that noise, patient movement  or external vibrations. Confirm the patient is not moving and the  cuff is free of external vibration, then, measure again.  NIBP: Retry, Cuff pressure  Cuff pressure exceeded more than 300mmHg during  failure (C19)  measurement. There is a possibility that the patient moved  during measurement or strong pressure from outside might be  YM6000 Service Manual...
  • Page 65 In case the recorder door is open, close the door.  paper.  In case the recorder paper is empty, insert new paper and close  the door.  SYSTEM: Abnormally  The monitor has been abnormally shut down last time.    shut down last time.  When power is lost for less than 30 seconds, the monitor  will preserve the current settings and trend data restored  automatically before the power loss. However, if the  power loss is over 30 seconds, the monitor will be back to  the previous user settings (or the factory default settings)  as per the ‘save settings on power off’ in the service menu. Contact qualified personnel in your facility or your local  supplier.  SYSTEM: No recorder  The recorder is not installed in your monitor. If required, contact  installed.  your local supplier.  Note:  If  the  alarm  message  still  appears,  contact  your  local  supplier  for  advice  on  remedial action.      YM6000 Service Manual...
  • Page 66 60 Troubleshooting             This page is intentionally left blank.  YM6000 Service Manual...
  • Page 67: Disassembly Guide

    CAUTION:  Observe  ESD  (electrostatic  discharge)  precautions  when  working  within the unit.      General    This section describes disassembly procedures with detailed disassembly instructions  and illustrations. Disassembly Sequence Flow Chart that is used to access replaceable  parts of the monitor is illustrated in Figure 15. The boxes on the flow chart represent  the  various  components  or  sub‐assemblies.  A  complete  listing  of  the  available  spare  parts and part numbers is in Spare Parts section. Follow the reverse sequence of the  disassembly procedures for reassembly.    The monitor can be disassembled down to all major component parts, including:  PCB assemblies  acquisition modules, the power supply assembly & the LCD assembly  battery  cables &wires  cases  printer    The following tools are required:  small, Phillips‐head (+) screwdriver  medium, Phillips‐head (+) screwdriver  needle‐nose pliers  YM6000 Service Manual...
  • Page 68: Figure 15. Disassembly Sequence Flow Chart

    NIBP Module Bracket Speaker C3-1 Printer Module Assembly Key Board AC Inlet Knob LCD Window Encorder LCD Guide Alarm / Power LED Wondow ECG Connector Temperature Connectors Water Trap Receiver IBP Conector Rubber Foot Figure 15. Disassembly Sequence Flow Chart      YM6000 Service Manual...
  • Page 69: Replacement Level Supported

    1. After  step  D2,  remove  2  AC  main  fuses  (F1,  F2:  250V/6.3A)  out  of  the  socket  if  required.  2. Remove 2 battery fuses (F3, F4: 250V/6.3A) out of the socket if required.  3. Replace (a) new fuse(s).    4. Reassembly the monitor.      Battery Replacement    This  section  describes  the  steps  to  remove  the  battery  from  the  monitor  for  replacement without disassembling the main case of the monitor    1. After step A, replace new batteries.  2. Reassemble the monitor.  YM6000 Service Manual...
  • Page 70: Monitor Disassembly

    This section describes the steps to remove battery and the handle and to separate the  front and rear case assemblies.      T0105 T0104 M6010 T0106 Figure 16. Battery and Handle Disassembly  Table 18. Part Descriptions – Battery and Handle  Part Codes  Descriptions  T0106  Battery Cover  M6010  Battery  T0104  Handle A  T0105  Handle B    A. Battery disassembly  1. Remove 2 screws (3 × 6) fastening the battery cover on the bottom of the monitor.  2. Remove the battery cover.  3. Remove a battery carefully from the monitor.    B. Handle disassembly  1. Pull the handle B out from the handle A.  2. Remove 2 screws (4 × 10) on the handle A.  3. Remove the handle A from the rear case.    YM6000 Service Manual...
  • Page 71: Figure 17. Monitor Disassembly

        Disassembly Guide 65 T0114 T0112 T0113 Figure 17. Monitor Disassembly   Table 19. Part Descriptions – Monitor  Part Codes  Descriptions  T0112  Side cap top  T0113  Side cap bottom  T0114  Side cap stopper    Before the steps C and D  1. Remove side caps top, bottom and stopper on the monitor.  2. Remove 6 screws (4 × 10) on the monitor.  3. Separate the front case assembly from the rear case assembly.  4. Disconnect the SMPS wire connected to the main board  5. Disconnect the main cable connected to the SMPS.  6. Disconnect the NIBP hose connected to the NIBP connector.  YM6000 Service Manual...
  • Page 72: Front Case Disassembly (C)

    Table 20. Part Descriptions – Side connector holder and Main board  Part Codes  Descriptions  T0107  Side connector holder  W0114  Main cable / 24 pin    C1. Side connector holder disassembly  1. Remove 2 screws (3 × 6) on the side connector holder.  2. Separate the side connector holder from the front case assembly.    C2. Main board disassembly  1. Disconnect  the  LCD  cable,  the  inverter  wire,  the  key  wire,  the  encoder  wire,  ECG  connector wire and two temperature wires connected to the main board.  2. Remove the main cable from the main board.  3. Remove 11 screws (3 × 6) on the main board.  4. Separate the main board from the front case assembly.  YM6000 Service Manual...
  • Page 73: Figure 19. Front Case Disassembly - Cpu Module, Spo

    Descriptions  P1031  Main board  P1033  CPU module  M0009   module  P1072  IBP board    C2‐1. CPU module disassembly  1. .After step C2, remove 4 screws (3 × 4) on the CPU module.  2. Remove the CPU module from the main board.  3. Remove 8 screws (3 × 4) to separate 4 supporters (5mm) from the main board.    C2‐2. SpO  module disassembly  1. .After step C2, remove 3 screws (3 × 6) on the SpO  module.  2. Remove the SpO  module from the main board.  3. Remove 6 screws (3 × 4) to separate 3 supporters (12mm) from the main board.    C2‐3. IBP board disassembly  4. .After step C2, remove 2 screws (3 × 6) on the IBP board..  1. Remove the IBP board from the main board.  2. Remove 2 screws (3 × 4) to separate 2 supporters (6mm) from the main board.    YM6000 Service Manual...
  • Page 74: Figure 20. Front Case Disassembly -Bracket And Key Board

    68 Disassembly Guide W0117 P1032 T1009 W0115   Figure 20. Front Case Disassembly –Bracket and Key board    Table 22. Part Descriptions –Bracket and Key board  Part Codes  Descriptions  T1009  Rubber Key  W0115  Key wire  P1032  Key board  W0117  LCD cable    C3. Bracket disassembly  1. Remove 7 screws (3 × 6) on the bracket.  2. Separate the bracket from the front case assembly.    C4. Key board disassembly  1. Remove the rubber key from the key board.  2. Remove 5 screws (3 × 6) on the key board.  3. Remove the key board from the bracket.  4. Remove the key wire form the key board.  YM6000 Service Manual...
  • Page 75: Figure 21. Front Case Disassembly - Slcd And Bracket

        Disassembly Guide 69 T4112 M4022   Figure 21. Front Case Disassembly – sLCD and Bracket    Table 23. Part Descriptions – LCD and Bracket  Part Codes  Descriptions  M4022  LCD  T4112  Bracket      C3‐1. LCD disassembly  1. After step C3, remove 4 screws (3 × 6) on the LCD.  2. Remove the LCD from the bracket.  YM6000 Service Manual...
  • Page 76: Figure 22. Front Case Disassembly - Knob, Lcd Window, Encoder Assembly And Connectors

    Figure 22. Front Case Disassembly – Knob, LCD window, Encoder assembly and  Connectors    Table 24. Part Descriptions – Knob, LCD window, Encoder assembly and Connectors  Part Codes  Descriptions  T0096  Front case  T0100  Knob  T0115  LCD window  X1012  Encoder board assembly  W0126  Encoder wire  T0097  LCD guide  T0098  Alarm LED window  T0099  Power LED window  E4111  ECG connector / 6 pin  X1010  Temperature assembly / 3 pin  X1011  Temperature assembly / 4 pin  E4118  Water trap receiver assembly  W0153  IBP wire / 4 pin  W0154  IBP wire / 5 pin  T1006  Rubber foot    YM6000 Service Manual...
  • Page 77   C9. Alarm LED window and Power LED window disassembly  1. Remove the alarm LED window form the front case.  2. Remove the power LED window from the front case.    C10. ECG connector disassembly  1. Remove 4 tapping screws (3 × 8) on the ECG connector.  2. Remove the ECG connector from the front case.    C11. Temperature connector disassembly  1. Remove two nuts (14mm) fastening the temperature connector from the front case.  2. Remove two temperature connectors from the front case.    C12. Water trap receiver disassembly  1. Pull the water trap receiver straight out to remove from the front case assembly.    C13. IBP connector disassembly  2. Remove 2 screws (3 × 8) on the IBP connector (4 pin).  1. Remove the IBP connector form the front case.  2. Remove 2 screws (3 × 8) on the IBP connector (5 pin)  3. Remove the IBP connector from the front case.    C14. Rubber foot disassembly  1. Remove two rubber foots from the front case.    YM6000 Service Manual...
  • Page 78: Rear Case Disassembly (D)

    72 Disassembly Guide Rear Case Disassembly (D)    This section describes the items that may be removed on the rear case assembly.    M0013 W0155   Figure 23. Rear case disassembly – EtCO  module    Table 25. Part Descriptions – EtCO  module  Part Codes  Descriptions  M0013  EtCO  module  W0155  EtCO  wire / 4 pin    D1. EtCO  module disassembly  1. Remove the EtCO  wire from the EtCO  module and the SMPS.  2. Remove 4 screws (3 × 6) on the EtCO  module.  3. Remove the EtCO  module from the SMPS.  YM6000 Service Manual...
  • Page 79: Figure 24. Rear Case Disassembly - Speaker

        Disassembly Guide 73 E9008   Figure 24. Rear case disassembly – Speaker    Table 26. Part Descriptions – Speaker  Part Codes  Descriptions  E9008  Speaker    D2. SMPS case disassembly  1. Remove 4 screws (3 × 6) on the SMPS case A.  2. Disconnect the AC inlet wire and the printer wire connected to the SMPS.  3. Separate the SMPS case from the rear case assembly.    D3. Speaker disassembly  1. Disconnect the speaker wire connected to the SMPS.  2. Remove 2 tapping screws (3 × 8) on the speaker.  3. Remove the speaker from the SMPS case A.    YM6000 Service Manual...
  • Page 80: Figure 25. Rear Case Disassembly -Smps Case, Smps And Nibp Module

    M2012  SMPS  1  W0113  SMPS wire  1  T4127  Ground terminal  1  M0015  NIBP module  1  W0165  NIBP cable  1    D2‐1.SMPS case A, B disassembly  1. After step D2, remove 3 screws (3 × 8) on the SMPS case B.  2. Remove the SMPS case B from the SMPS case A.    D2‐2. SMPS disassembly  1. After step D2, remove 7 screws (3 × 6) on the SMPS.  2. Remove the SMPS from the SMPS case A.  3. Remove the SMPS wire from the SMPS.  4. Remove the ground terminal from the SMPS.    D2‐3. NIBP module disassembly  4. After D2, remove the remove 4 screws (3 × 6) on the NIBP module.  5. Remove the NIBP module from the SMPS case A  YM6000 Service Manual...
  • Page 81: Figure 26. Rear Case Disassembly - Printer Module And Ac Inlet

    W0124   Figure 26. Rear case disassembly – Printer module and AC inlet    Table 28. Part Descriptions – Printer module and AC inlet  Part Codes  Descriptions  Qty  T0101  Rear case  1  X4104  Printer module  1  W0118  Printer wire  1  W0124  AC inlet  1    D4. Printer module disassembly  1. Remove the printer wire from the printer module.  2. Open the printer door to separate the printer module from the rear case.  3. Remove 3 screws (3 × 6) on the printer module.  4. Separate the printer module from the rear case.    D5. AC inlet disassembly  1. Remove the wire connected to the AC inlet.  2. Pull out the AC inlet to separate from the rear case.    YM6000 Service Manual...
  • Page 82 76 Disassembly Guide             This page is intentionally left blank.  YM6000 Service Manual...
  • Page 83: Spare Parts

    Spare Parts   WARNING:  Follow  local  government  ordinances  and  recycle  instructions    regarding disposal or recycling of device components, including batteries.      General    Spare parts, along with part numbers, are shown in Table  . “Item No.” corresponds  to the circled callout numbers in Figure  .      Obtaining Replacement Parts    Mediana Technical Service Department provides technical assistance information and  replacement parts. To obtain replacement parts, contact Mediana. Refer to parts by the  part names and part numbers.    YM6000 Service Manual...
  • Page 84: Figure 27. Ym6000 Exploded View - Spare Parts

    78 Spare Parts Figure 27. YM6000 Exploded View – Spare Parts  YM6000 Service Manual...
  • Page 85 30  M2012  SMPS  31  E9008  Speaker  32  T0101  Rear case  33  W0124  AC inlet wire  34  W0165  NIBP cable / 16 pin  35  M0015  NIBP module  36  M0009   Module  37  M0013  EtCO  module  38  P1072  IBP board  39  T0098  Alarm LED window  40  T0097  LCD guide  41  T0099  Power LED window  YM6000 Service Manual...
  • Page 86 Bracket  45  W0153  IBP wire / 4 pin  46  W0154  IBP wire / 5 pin  47  W0155  EtCO  wire / 4 pin  48  E4118  Water trap receiver assembly  49  W0115  Key wire / 12 pin    50  T0111  Printer cover    Note:  You  can  use  the  assembly  kit  (SS8000‐0)  supplied  by  the  manufacturer  for  screws, washers and supporters.    YM6000 Service Manual...
  • Page 87: Packing For Shipment

      To ship the monitor for any reason, follow the instructions in this section.    Pack the monitor carefully. Failure to follow the instructions in this section may result  in  loss  or  damage  not  covered  by  the  Mediana  warranty.  If  the  original  shipping  carton  is  not  available,  use  another  suitable  carton;  North  American  customers  may  call Mediana Technical Service Department to obtain a shipping carton.  Prior to shipping the monitor, contact your supplier or the Mediana office (Technical  Service  Department)  for  a  returned  goods  authorization  (RGA)  number.  Mark  the ...
  • Page 88: Repacking In A Different Carton

    82 Trends Repacking In a Different Carton    If the original carton is not available, use the following procedure to pack the monitor:    1. Place the monitor in a plastic bag.  2. Locate a corrugated cardboard shipping carton with at least 200 pounds per square  inch (psi) bursting strength.  3. Fill the bottom of the carton with at least 2 inches of packing material.  4. Place the bagged unit on the layer of packing material and fill the box completely  with packing material.  5. Seal the carton with packing tape.  6. Label the carton with the shipping address, return address, and RGA number, if  applicable.      YM6000 Service Manual...
  • Page 89: System Processing Description

    System Processing Description     System Overview    The  YM6000  monitor  is  a  multi‐function  monitor  for  use  on  adult,  pediatric  and  neonatal  patients;  ECG,  heart  rate,  non  invasive  blood  pressures,  arterial  oxygen  saturation, pulse rate, respiration rate and temperature.  In addition to monitoring and displaying the status of these physiological parameters,  the instrument performs various microprocessor‐programmed analytical functions;  Creating both visual and audible alarm signals when settable limits are violated;  Creating and displaying warning/error messages when conditions are detected that  would degrade or prevent valid measurements;  Creating and displaying graphical or tabular trend data;  Providing  input  to  an  optional  recorder  for  printout  of  current  or  tabular  trend ...
  • Page 90: Figure 29. Power Unit Block Diagram

    84 System Processing Description Unit Description    Power unit: consists of power entry module, power supply, battery charger, battery,  external DC input and DC/DC unit.    AC on Battery Level 9-19V 0.5A AC-DC AC 85-260V (Inrush 1.5A) 50/60Hz 5V 2.5A Battery NI-MH 10cell (9-16V) 12V 1.7A Charged for (Inrush 2A) 12 hours DC/DC 3800mAh Charger Thermistor FUSE 10kohm NTC DC/DC Charging cut-off at...
  • Page 91: Figure 32. Sound Unit Block Diagram

        System Processing Description 85 Sound unit: consists of 8bit PIC micom, 2‐channel amplifiers and speaker.    RS-232 VOutR Alarm sound ARM9 MCU Audio interface PIC16F73 S3C2440A VOutL Amplifier Pulse tone sound   Figure 32. Sound Unit Block Diagram    Communication unit : 4‐channel UART  Asyncronous Serial Communication with NIBP Unit S3C2440A System BUS Asyncronous Serial Communication with Analog Control Unit UART TL16C554A Asyncronous Serial Communication with Printer Unit...
  • Page 92: Figure 36. Nibp Unit Block Diagram

    86 System Processing Description NIBP unit: measures non‐invasive blood pressure data.        Figure 36. NIBP Unit Block Diagram  EtCO unit: measures EtCO  data.  2        Figure 37. EtCO  Unit Block Diagram  ECG unit: measures electrocardiographic waveform data.      Figure 38. ECG Unit Block Diagram         ...
  • Page 93: Figure 39. Respiration Unit Block Diagram

        System Processing Description 87   Respiration unit: measures respiration rate data.  Patient Isolation Constant Isolation DC +/-5V DC +5V Isolation Current DC/DC Source Patient High Band Band Connector Photo Protection Pass Rectifier Pass Pass coulper Circuit Filter Filter Filter PIC16C74B Gain control   Figure 39. Respiration Unit Block Diagram ...
  • Page 94: Figure 42. Ibp Unit Block Diagram

    88 System Processing Description IBP unit: measures Invasive blood pressure waveform data.    Figure 42. IBP Unit Block Diagram      Analog  control  unit:  analog  circuit  control,  PIC16C74B  –  selection  of  ECG  channels and filter, size adjustment, verification of lead off, QRS, heart rate, etc.  Patient Isolation DC +5V Isolation DC +/-5V Isolation DC/DC ECG Filter Selection ECG Gain Control ECG Lead Off ECG Signal Saturation PIC16F74 Pace Pulse Detection Serial Communication...
  • Page 95: Ecg Processing

        System Processing Description 89 ECG Processing    The  measurement  of  the  skin  surfaces  electrocardiogram  is  based  on  the  electrical  signals  on  the  skin  surface,  produced  as  the  heart  muscle  contracts  and  relaxes.  The  signals  are  detected  by  electrodes  placed  on  the  patient  body.  The  information  on  heart activity carried by these signals varies with the placing of the electrodes. ...
  • Page 96 90 System Processing Description       ● Oscillometric Response  (Pressure Pulses)  Distole Systole         TIME           Systole ● Cuff Pressure  Distole   TIME     Overall Accuracy Discussion  Overall system accuracy shall be determined by considering various influences of the  pressure  sensor  accuracy,  motion  artifacts,  other  artifact  created  by  pressure  valve,  technical errors of electrical components, and the origin error of oscillometric method. ...
  • Page 97: Spo 2 Processing

        System Processing Description 91    Processing   Pulse oximetry works by applying a sensor to a pulsating arteriolar vascular bed. The  sensor contains a dual light source and photodetector. Bone, tissue, pigmentation, and  venous  vessels  normally  absorb  a  constant  amount  of  light  over  time.  The  arteriolar  bed  normally  pulsates  and  absorbs  variable  amounts  of  light  during  systole  and  diastole,  as  blood  volume  increases  and  decreases.  The  ratio  of  light  absorbed  at  systole ...
  • Page 98: Figure 44. Oxyhemoglobin Dissociation Curve

                      100 – (%carboxyhemoglobin + %methemoglobin)     Accuracy  The saturation (SpO ) accuracy specification and/or pulse rate (PR) performance were  analyzed by comparative oximetry performance (COPS) tests between YM6000 SpO   module  and  MP506  module  with  the  same  version  of  the  oximetry  algorithm  from  Nellcor. This was to demonstrate that the performance of YM6000 SpO  module was  equivalent to that of MP506, which had been validated during both standard motion,  combined  motion  and  cold‐induced  peripheral  vasoconstriction  (low  perfusion) ...
  • Page 99: Respiration Processing

    The  resistance  of  the  thermistor  is  measured  by  passing  a  current  through  it  and  measuring  the  voltage  developed  across it.  The  YM6000  monitor  is  designed  to  accept  the  signals  from  electrically  isolated  a  range of temperature probes from YSI‐400 series. The probes may be used for skin or  rectal  temperature  measurement.  Probes  are  furnished  with  a  standard  10‐feet  lead; ...
  • Page 100: Invasive Blood Pressure Processing

    94 System Processing Description Invasive Blood Pressure Processing       The pressure transducer is connected to a pressure line which, by means of a catheter  is  invasively  connected  to  the  patient  blood  stream.  The  force  of  movement  of  the  blood in the patient vessels in transported by the fluid column n the pressure line to  the transducer. This movement cause an electrical signal to be generated which is then  amplified to display the pressure wave and the numeric for the systolic, diastolic and  mean pressure values.    The  blood  pressure  is  influenced  by  the  respiratory  system.  This  occurs  in  spontaneous ...

Table of Contents