M o d e l 3 0 0 0 / 5 0 0 a n d M i c r o 3 1 0 0 / 5 0 5 Vo l u m e t r i c I n f u s i o n P u m p Te c h n i c a l S e r v i c e M a n u a l...
Page 3
The pump must be referred to a suitably qualified technician or returned to Graseby Medical in order to have the fault rectified. WARNING: Before using the pump, it should be inspected for physical damage. The pump should not be used if damage is evident, and should be returned to service personnel for repair before being returned to use.
Page 4
22. WARNING: If a backup alarm sounds, the pump should be immediately removed from the patient and sent to be repaired by a Graseby Medical qualified technician. Failure to do so may cause patient injury or death.
Page 5
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Cautions Cautions tell you about dangerous conditions that can occur and cause damage to the pump if you do not obey all of the instructions in this manual.
Page 6
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . I s s u e A ( A p r i l 2 0 0 2 ) 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
Introduction Scope of this manual ...................... 1 - 1 Related manuals ......................1 - 1 Graseby service contacts ....................1 - 1 Product Overview ......................1 - 2 System description ......................1 - 2 Optional labels ........................ 1 - 2 Software versions ......................
Page 10
Vo l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . About Snapshot screens ....................
Page 11
G r a s e b y M e d i c a l L t d . Vo l u m e t r i c I n f u s i o n P u m p s Chapter 7 - Maintenance Procedures Maintenance Procedures ....................
Page 12
Vo l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Chapter 9 - Illustrated Parts List Illustrated Parts List for pump Serial numbers from 3000 to 60,000 ......................
Page 13
Summary of hardware connection and handshaking ........... 12 - 7 Example Commands ....................12 - 8 Interfacing with pump software version 0.68 and above ..........12 - 9 Model 3000/500 3100/505 Computer Interface ............12 - 9 Protocol Definition ......................12 - 9 Character ........................12 - 9 Format - version 0.68 and above ................
Page 14
I n t r o d u c t i o n C h a p t e r 1...
Related manuals Chapter 9 Illustrated Parts List Chapter 9 Illustrated Parts List, Graseby service contacts UK service address USA service address 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 ) 1 —...
Model 3000/500 Micro 3100/505 Note: Graseby Volumetric Infusion Pumps are also known as the Model 500 and Micro 505. With the exception of the numbering the 3000/500 and the 3100/505 versions of the pumps are identical. Optional labels Software versions Chapter 4, Volumetric Pump Menus.
Page 17
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s About Version 0.71 Upgrading to Version 0.71 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
Page 18
S p e c i f i c a t i o n C h a p t e r 2...
60,000 From bottom of drip chamber to top of pump Model 3000/500 Micro 3100/505 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 ) 2 —...
Page 20
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Model 3000/500 Micro 3100/505...
Page 21
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Power Over-current protection Voltage...
Page 22
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Primary and Secondary Infusions Rate range Model 3000/500 Range Increment Micro 3100/505...
Page 23
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Volume to be infused range Model 3000/500 Range Increment...
Page 24
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Rate range Model 3000/500 Range Increment...
Page 25
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Symbols used on the pump Front panel symbols Inside battery door symbols...
Trumpet curves Model 3000/500 Graseby Standard Adult Set, 8C-820 Micro 3100/505 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 ) 2 —...
Page 28
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . I s s u e A ( A p r i l 2 0 0 2 ) 2 —...
Page 29
O v e r v i e w o f P u m p O p e r a t i o n & I n i t i a l C h e c k s C h a p t e r 3...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Exterior of the 500/3000 pump - front (Revised case design shown here) Indicators and displays...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Exterior of the pump - rear (Revised case design shown here) 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . How the pump is powered AC power connection Battery and portable operation...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s How the pump works Instruction Manual Technical User Manual...
Page 35
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Pressure / occlusion monitoring Pumps with revision 0.64 and higher software Occlusion...
Page 36
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Backlight for Message Display Backlight on AC Backlight on Battery...
Page 37
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . LED/segment monitoring I s s u e A ( A p r i l 2 0 0 2 ) 3 —...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Power up tests and checks On/Off Cold boot...
Page 39
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Options On/Off Silence...
Page 40
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s End of cold boot - special infusion modes Warm boot insistent...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Power monitoring On AC Mains Instruction Manual, Security on the...
Page 43
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . WARNING: Correct management of battery charging, as described in this documentation is essential to ensure that the pump can operate...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Volumetric Pump Menus Which version of software is in the pump? Introduction to menus...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . About the Technician menu Common parameters on Technician and Biomedical menus: Entering the Technician menu...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Summary of Technician menu parameters Options 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Using Technician parameters Options Pump Identifier...
Page 52
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s KVO rate entry Occlusion default setting Occlusion Alarm Setting...
Page 53
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Volume Over Time Dose-Rate Calculation DRC Recovery...
Page 54
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Service Date Micrograms Options...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . About the Biomedical menu System faults Entering the Biomedical menu...
Page 56
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Using the Biomedical menu - after a fault has occurred 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 )
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s About Snapshot screens Using PC history software to interpret snapshot screens To display Snapshot Screen-a...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Decoding Snapshot screen-a Top row of snapshot screen-a: Example shows...
Page 60
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Snapshot screen-a - system state codes list continues ...
Page 61
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . System state codes, continued 4 —...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Decoding Snapshot screen-b Note 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
Page 63
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Converting Snapshot screen-b values from decimal to binary 4 —...
Page 64
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Chart A Pump Main Modes Bit 7 Bit 6...
Page 65
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Chart D Pump sensor flags and occlusion alarm setting Bit 7...
Page 66
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Chart F - Battery Voltage Analog-to-Digital Value...
Page 67
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Chart G - State Code 33 - Corrupted Triple Redundant variable in NV RAM Variable Identifier Chart G - State Code 35 - Undefined interrupt...
Page 68
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Chart G - State Code 93 To convert the GGG value: To interpret the result...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Using Biomedical Parameters Options Time and date display...
Page 70
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Charge History Pressure display parameter (on battery power) 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
Page 71
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Pressure display - calibration mode Field Range...
Page 72
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Baud rate Options On/Off...
Page 73
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Model number and software version Audio alarm test non-insistent...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Audio alarm specifications Enabling special programming functions Enable/disable rate taper...
Page 75
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Enable/disable volume over time Enable/disable dose-rate calculation 4 —...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s About the Service Functions menu Entering the Service Functions menu Occlusion Alarm Setting...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Cassette housing and door insistent insistent...
Page 82
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Figure 5.2 Cassette housing and door 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Cam housing switch and safety clip switch Cam housing Stepping motor...
Page 84
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 ) 5 —...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Cams, pistons and valves of the pumping mechanism Pumping cycle description (Figs 5.3, 5.4 and 5.5) I s s u e A ( A p r i l 2 0 0 2 )
Page 86
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s [13] [14] [15]...
Page 87
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . valve opens to allow fluid flow to upper upper piston draws...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Empty bag contacts No flow above pump...
Page 90
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Photo sensor Over pressure valve and swing arm Back pressure regulation...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Pole clamp and IV pole Modular connection system Attaching a single pump to an IV pole...
Page 92
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Attaching more than one pump to an IV pole (Revised case design shown here) Inspecting the modular connections...
Page 93
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . How to connect pumps together Note: As you tighten the screw on the pole clamp, the actuator arm adjusts the modular connection hook and disconnect button.
Page 94
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Moving an IV pole with pumps connected Disconnecting modular connected pumps WARNING:...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Maintenance Procedures Introduction 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Recommended test equipment Recommended troubleshooting equipment I s s u e A ( A p r i l 2 0 0 2 )
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Periodic cleaning Recommended cleaning agents Cleaning...
Page 101
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Cleaning agents NOT recommended for use Activated dealdehydes Alcohol solutions...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Chemical decontamination procedures Recommended yearly checks and tests I s s u e A ( A p r i l 2 0 0 2 )
Page 103
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Carry out lubrication Recommended lubricant Check pole clamp screws...
Page 104
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Recommended annual functional and accuracy tests WARNING: Do not perform these tests while the pump is in use on a patient,...
Page 105
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s 3.
Page 106
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . 4.
Page 107
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s 6.
Page 108
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . 10.
Page 109
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Hold Note: Occlusion...
Page 110
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Silence 13.
Page 111
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s non- insistent Hold...
Page 112
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Method 2 I s s u e A ( A p r i l 2 0 0 2 ) 7 —...
Page 113
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 ) 7 —...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Troubleshooting Troubleshooting, Chapter 4 WARNING: Unplug the AC power cord before opening the pump housing to avoid risk of electric...
Page 116
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Table 7.2 Error code related troubleshooting chart Code Error/event Possible cause/solution...
Page 117
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Table 7.3 Symptom/cause troubleshooting chart Symptom Possible cause/solution...
Page 118
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Symptom Possible cause/solution Back Up...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Replacing the LCD Identify the LCD type 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
Page 120
Replacement components for EL type display Part name Part numbers for model Contents 3000/500 3100/505 I s s u e A ( A p r i l 2 0 0 2 ) 7 — 2 4 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Reinitialising the pump after replacing the NVRAM (IC18) NV RAM U18 Part Numbers CPU board initialisation for Version 0.53 software...
Page 122
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . CPU board initialisation for Version 0.53 software continued Options...
Page 123
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s CPU board initialisation for Version 0.63 software continued Special Functions...
Page 124
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . CPU board initialisation for Version 0.63 software continued Options...
Page 125
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s CPU board initialisation for Version 0.63 software continued Resetting or verifying the Rev to Pos Range...
Page 126
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . CPU board initialisation for Version 0.65 / 0.67 / 0.71 software continued Special Functions...
Page 127
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Technician Menu (Version 0.71 software only) Total Volume Options...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Door assembly removal Figure 8.2 Door disassembly...
Page 133
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Door Rear - reassembly procedure for CE marked pumps ONLY, with serial numbers 3000 to 59,999 and for ALL versions of pumps with serial numbers from 60,000...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s CPU/Power supply board removal WARNING: Unplug the AC power cord before opening the pump housing to avoid risk of electric shock, which could result in death or serious injury.
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . I/O board assembly removal WARNING: Unplug the AC power cord before opening the pump housing to avoid risk of electric shock, which could result in death or serious injury.
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Display board assembly removal Troubleshooting WARNING: Unplug the AC power cord before opening the pump housing to avoid risk of electric...
Page 137
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Membrane switch replacement (Original case design) WARNING: Unplug the AC power cord before opening the pump housing to avoid risk of electric...
Page 138
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Membrane switch replacement (Revised case design) WARNING: Unplug the AC power cord before opening the pump housing to avoid risk of electric...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Cassette housing disassembly Safety clip mechanism disassembly Reassembly...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Calibration and adjustment of components Cam housing switch adjustment Figure 8.8 Cam switch adjustment...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Voltage Checks Power supply voltage check 16V DC check - power cord connected...
Page 148
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s 12V DC check - battery D11 Cathode TP7 (CPU/Power supply)
Page 149
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . 5.0V DC check - power cord connected The front housing must be removed to verify the 5.0V DC voltage.
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 3000 to 59,999 CAUTION...
General assembly of the volumetric pump Item Description Part number 3000/500 3100/505 When ordering a Keypad Overlay, also order an Overlabel if one is shown for your country in the list below: Note: If no Overlabel is listed for your country, then no Overlabel is required.
Page 154
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 3000 to 59.999 Figure 1 General Assembly of the Volumetric pump 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
Page 155
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 3000 to 59,999 9 —...
Page 156
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 3000 to 59,999 Item Description...
Page 157
G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 3000 to 59,999 Item Description Part number 3000/500 3100/505 See Figure 5 see Figure 6 See Figure 4 Note: If a replacement LCD is required and the Display Board is fitted with inverter U401, then the Display Board must be modified and the Display Conversion Kit should be ordered.
Page 158
Item Description Part number 3000/500 3100/505 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 )
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 3000 to 59,999 Pump assembly Item...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 3000 to 59,999 Rear housing assembly Item...
Page 161
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 3000 to 59,999 Item Description...
Page 162
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 3000 to 59,999 Item Description...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 3000 to 59,999 Display board assembly Item...
Page 164
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 3000 to 59,999 Item Description...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 3000 to 59,999 CPU power supply/RS232 board assembly Item...
Page 166
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 3000 to 59,999 Figure 5 CPU power supply/RS232 board assembly 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 3000 to 59,999 I/O board assembly Item...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for Serial numbers from 60,000 CAUTION Because the circuit boards used in the pumps primarily consist...
Illustrated Parts List for pump Serial numbers from 60,000 only General assembly of the volumetric pump Item Description Part number 3000/500 3100/505 see Figure 8 see Figure 9 9 — 1 8 I s s u e A ( A p r i l 2 0 0 2 )
Page 170
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 60,000 only Figure 7 General Assembly of the Volumetric pump 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l...
Page 171
V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 60,000 only Item Description Part number 3000/500 3100/505 See Figure 11 see Figure 12 See Figure 10...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 60,000 only Pump assembly Item...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 60,000 only Rear housing assembly Item...
Page 174
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 60,000 only Figure 9 Rear housing assembly - labels 9 —...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 60,000 only Display board assembly Item...
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Illustrated Parts List for pump Serial numbers from 60,000 only CPU power supply/RS232 board assembly Item...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Illustrated Parts List for pump Serial numbers from 60,000 only I/O board assembly Item...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Electrical Diagrams Introduction This chapter contains the following diagrams:...
Page 180
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . CPU/Power Supply Board Audio Unit...
Page 181
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s CPU/Power Supply Board AC_ON RESET...
G r a s e b y M e d i c a l L t d . Volumetric Infusion Pumps Figure 10-4 Display board schematic 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 ) 1 0 —...
Volumetric Infusion Pumps G r a s e b y M e d i c a l L t d . Figure 10-5 CPU/Power supply schematic (sheet 1 of 3) I s s u e A ( A p r i l 2 0 0 2 ) 10 —...
Page 185
G r a s e b y M e d i c a l L t d . Volumetric Infusion Pumps Figure 10-6 CPU/Power supply schematic - CE (sheet 2 of 3) 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 ) 1 0 —...
Volumetric Infusion Pumps G r a s e b y M e d i c a l L t d . Figure 10-7 CPU/Power supply circuit (sheet 3 of 3) I s s u e A ( A p r i l 2 0 0 2 ) 10 —...
G r a s e b y M e d i c a l L t d . Volumetric Infusion Pumps Figure 10-8 Input/Output schematic 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 ) 1 0 —...
G r a s e b y M e d i c a l L t d . Volumetric Infusion Pumps Waveform Diagrams Introduction Oscilloscope waveforms at selected test points are shown in the following list of figures: Figure Title 11.1 Air detect - emitter, TP3 on I/O board 11.2...
Page 191
Volumetric Infusion Pumps G r a s e b y M e d i c a l L t d . 999 mL/h (99.9 mL/h on 3100/505) .5 V/Div 10 ms/Div Normal Triggering Positive Slope Figure 11.1 Air detect - emitter, TP3 on I/O board 11 —...
Page 192
Figure 11.2 Air detect - detector, TP2 on I/O board The wave form can be seen approximately every 5 seconds on the 3000/500 pumps, and approximately every 60 seconds on 3100/505 pumps. 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 ) 1 1 —...
Page 193
Volumetric Infusion Pumps G r a s e b y M e d i c a l L t d . 50.0 mL/h 10 V/Div 10 ms/Div Normal Triggering Positive Slope Figure 11.3 Motor drive pulses, TP7 on I/O board Motor Drive Pulses - 50.0 mL/h 11 —...
Page 194
G r a s e b y M e d i c a l L t d . Volumetric Infusion Pumps 100 mL/h (99.9 mL/h on the 3100/505 pumps) 10 V/Div 5 ms/Div Normal Triggering Positive Slope Figure 11.4 Motor drive pulses, TP7 on I/O board Motor Drive Pulses - 100 mL/h (99.9 on the 3100/505 pumps) 3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l I s s u e A ( A p r i l 2 0 0 2 )
Page 195
Volumetric Infusion Pumps G r a s e b y M e d i c a l L t d . 3000/500 pumps only: 900 mL/h 10 V/Div 5 ms/Div Normal Triggering Positive Slope Figure 11.5 Motor drive pulses, TP7 on I/O board Motor Drive Pulses - 900 mL/h 11 —...
Page 196
Pulse duration and time between pulses varies with pumping rate. Example above corresponds to a typical pulse width when pump is run at 999 mL/hr on 3000/500 pumps. For 505/3100 pumps the infusion can be run in the Diagnostic mode at 999 mL/hr...
Page 197
Negative Slope Figure 11.7 Interrupter detector, TP4, I/O board On the 3000/500 pumps the signal appears approximately every 3 seconds at 500 mL/h, and approximately every 15 seconds on the 3100/505 pumps, at 99.9 mL/h. 11 — 8 I s s u e A ( A p r i l 2 0 0 2 )
Page 198
R e m o t e M o n i t o r i n g & C o n t r o l o f t h e p u m p C h a p t e r 1 2...
Remote monitoring and controlling of the pump WARNING: While Graseby Medical Limited have taken all reasonable steps to ensure that the pump operates correctly while under remote control, it is the responsibility of the person who designs and implements the controlling device to ensure that the resulting system (pump and controlling device) is fit for its intended purpose.
V o l u m e t r i c I n f u s i o n P u m p s G r a s e b y M e d i c a l L t d . Summary of commands - Version 0.67 and below Commands for monitoring the pump - CM Command...
G r a s e b y M e d i c a l L t d . V o l u m e t r i c I n f u s i o n P u m p s Terminator Error Processing Summary of hardware connection and handshaking...
Volumetric Infusion Pumps Interfacing with pump software version 0.68 and above The information contained in this section applies to Graseby Model 3000/500 3100/505 pumps with main CPU versions greater than or equal to 0.68 Model 3000/500 3100/505 Computer Interface The Graseby Model 3000/500 3100/505 pumps are designed to interface with a computer that supports an RS-232 serial port.
Graseby Medical L t d Volumetric Infusion Pumps Command Record Format - version 0.68 and above A command record consists of a string of characters sent from a host to the pump that are in a specific order and follow the rules listed here.
Page 209
Graseby Medical L t d Volumetric Infusion Pumps CM (Computer Monitor) commands are used for monitoring pump information. The Return Record Format section gives details of parameters and return information. The pump shall respond, when queried, to the following “CM” or monitor-only commands:...
Page 210
Graseby Medical L t d Volumetric Infusion Pumps The pump will also respond to the following “CI” or interface/control commands: CI (Interface) Command Code Name Parameter Needed Silence Key Keypress Infused Volume Clear ü Time and Date Set ü Pump On or Off.
Graseby Medical L t d Volumetric Infusion Pumps Record Terminator The last character of any record transmitted by the host or pump will be an ASCII ">" (3E hex). The use of this character is not allowed in the command parameter area.
Page 214
Graseby Medical L t d Volumetric Infusion Pumps verify that the return code is 'Command Accepted' and that the rate value returned is indeed the rate that was expected). Conversely, if the return code is not ‘Command Accepted’, the host should ignore the command data (DDD) returned because it may be invalid or non- existent.
Page 215
Graseby Medical L t d Volumetric Infusion Pumps Byte Bits Description represent the current alarm number. A value of zero implies no alarm. In general, an alarm will start an audible signal and change the standard message to an alarm specific message.
Page 216
Graseby Medical L t d Volumetric Infusion Pumps The response to all accepted CI commands will include the volume infused, volume limit, rate, and alarm bytes formatted as described in Table 1. 12 - 18 Issue 1(April 2002) 3000/3100 and 500/505 Service Manual...
Graseby Medical Ltd. Volumetric Infusion Pumps Pump Powered Down - AC/Battery The Rate and VTBI values are always retained over a power cycle, regardless of the power source. The occlusion pressure setting will always be defaulted upon power down, regardless of the power source.
Volumetric Infusion Pumps Graseby Medical L t d Response Time-outs and Error Processing It shall be the host's responsibility to monitor pump responses, detect a no response time-out and detect all other problems inherent in serial communication. Error detection and recovery are the responsibility of the host.
Graseby Medical Ltd. Volumetric Infusion Pumps Example Commands Version 0.68 and above Command Command Sent by Pump Response: Description: Host: Data query <09Dxxxx> <20RAAAAAAaaaa.abbbb.bccc.cXXXX> Enable CI commands <0BI60xxxx> <20RAAAAAAaaaa.abbbb.bccc.cXXXX> with 60 s timeout Set Limit to 1000 ml <0DL1000xxxx> <20RAAAAAAaaaa.a1000.0ccc.cXXXX>...
Page 226
S I M S G r a s e b y L t d . V o l u m e t r i c I n f u s i o n P u m p s Index Symbols adjustment 8 - 14 3000/3100 disassembly 8 - 13...
Page 227
V o l u m e t r i c I n f u s i o n P u m p s S I M S G r a s e b y L t d . pump 8 - 1 charging 4 - 23 Disconnect clearing 4 - 30...
Page 228
S I M S G r a s e b y L t d . V o l u m e t r i c I n f u s i o n P u m p s Linkage system 5 - 1 Optional labels 1 - 2 Lubrication 7 - 7 Oscilloscope waveforms 11 - 1...
Page 229
V o l u m e t r i c I n f u s i o n P u m p s S I M S G r a s e b y L t d . Software version 1 - 2 0.71 features 1 - 3 Quick Rate Change parameter 4 - 5 on Biomed menu 4 - 26...
Page 230
S I M S G r a s e b y L t d . V o l u m e t r i c I n f u s i o n P u m p s Valve integrity test 7 - 11 Version 0.53 initialisation 7 - 23 Version 0.63 initialisation 7 - 24 Version 0.65 / 0.67 / 0.71 initialisation 7 - 27...
Page 231
The details given in this Manual are correct at the time of going to press. The company, however, reserves the right to improve the equipment shown. For further information, please contact your local distributor or Graseby Medical direct on +44 (0)1923 246434 Graseby Medical Ltd.