SAFETY WARNING Low temperatures are present in this equipment. Extreme care should be taken. DO NOT let bare skin come into contact with metal surfaces. Bright Instruments...
SAFETY WARNING Extremely sharp knives/blades. Use knife/blade guards at all times. Use correct tools for removal and insertion of knives/blades. DO NOT leave knives/blades laying around. Place knives/blades not in use, into box/wallet provided. Bright Instruments...
Your attention is therefore drawn to the of the dangers and observe the following following precautions: warnings: a. Do not leave the microtome unattended with an exposed knife in position. Remove the knife, or cover it with the guards provided. Bright Instruments...
Open the window and place a beaker of ensuring that users are aware of information. containing 10ml of ammonia SG.880 in the this situation. chamber. Close the window. e. Leave for one hour. The cryostat is now decontaminated.essential that it be decontaminated. Bright Instruments...
Page 8
For further information regarding alternative decontamination procedures please refer to ‘Safe Working and the Prevention of Infection in Clinical Laboratories’ , ISBN 0 11 885446 1. Bright Instruments...
Seller’s liability under this clause shall be in lieu of any warranty or condition implied by law as to the quality or fitness for any particular purpose of such goods. Bright Instruments...
DESPATCH) and Bright Instrument Company Limited immediately. The microtome is supplied already installed in the chamber, carefully remove the packing from around the microtome and the anti-roll Bright Instruments...
(Section 2.1.3) switching on. We recommend the instrument is left standing for at least eight hours, and preferably overnight, before switching on. Bright Instruments...
Set button, then press Up or Down buttons until desired temperature is set. After 5 seconds the actual chamber temperature will show. QUICK FREEZER TEMPERATURE DEFROST CHAMBER TEMPERATURE To review set temperature at any time press Starlet 2212 and release Set button. Instruments Bright Instruments...
If the blade has by scraping it with a wooden or plastic been used only at one end, it may be reversed scraper or wiping it with a cloth dampened and refitted as described above. with alcohol. Bright Instruments...
IMPORTANT: Always push the operating ANGULAR lever/hand wheel down slightly before changing thickness setting. DO NOT SCALE ! Bright Instruments IF IN DOUBT - ASK! Object Holder (34) 2.2.6 OBJECT HOLDER Cryo-M-Bed before inserting in the carrier (head)(12) and First apply a coating of embedding compound...
To adjust the height of the ARP turn screw Figure 12a. Figure 12b. (84) Upward adjustment can be made with the ARP against the blade; downward adjustment must Render be made with it away from the blade. If the Anti Roll Plate Render Magnaplate Anti-Roll Plate Bright Instruments...
16µm. d. After cutting an acceptable section, lock h. Clean the blade surface using the bristle the specimen at the bottom by using the hand brush with an upward motion. wheel lock, or by pushing the lever back. Bright Instruments...
See notes on Decontamination (section 3.6). minute defrosts. d. Do not extend the defrost cycle time beyond thirty minutes, as this could cause the entire chamber to warm up above 0ºC. Bright Instruments...
Determine which fuse is (48/72) to correct the gap. blown and replace it with one of exactly the f. Set pawl in the rewind position. same type and rating. Bright Instruments...
“Code microbiological and clinical departments. is fitted. of Practise for the Prevention of Infection in Clinical Laboratories and Post-mortem Rooms”, a. Defrost the cryostat completely. ISBN 0 11 320464 7 b. Remove the microtome (Section 3.3.3) Bright Instruments...
Page 21
Carry out a similar procedure on the microtome. g. Ensure microtome and chamber are dry. h. Apply low temperature grease sparingly to screw threads and re-install microtome in the chamber. i. The cryostat is now ready to be switched back on. Bright Instruments...
Specimen loose on holder? Has the specimen dehydrated? Freezing technique too slow? Clamping screw too loose? Cutting temperature too warm? Cutting temperatures too warm / cold? Anti-roll plate adjusted too high? Tissue expansion due to block warming up? Bright Instruments...
6.7 SECTIONS CURLING 6.9 CUTTING TEMPERATURES 6.7.1 SECTIONS CURL OVER ROLL PLATE Standard textbooks on histological technique give table of recommended cutting Anti-roll plate too low? temperatures for different tissues. However, in most cases the following guidelines will prove adequate. Bright Instruments...
2-12μm in 2μm increments Instruments 4-24μm optional Dimensions H315 x W465mm(min)/ 546mm(max) x D740mm Packing Net weight: 40kg Gross weight: 48kg Packing dimensions: H400 x W920 x D550mm Shipping volume: 0.20m3 Weights vary according to specification of instrument Bright Instruments...
An environmental hazard cannot be excluded in the event Eff ects: of unprofessional handling or disposal. Instruments Office: Telephone: +44 (0) 808 168 9697 Web: brightinstruments.com Bright Instruments LTD Burnett House, Lakeview Court Email: sales@brightinstruments.com Member: Instruments Ermnie Business Park, Huntingdon Cambridgeshire, PE29 6UA...
Page 29
Advice on common storage: Do not store together with Oxidising agents. Further information In the event of fi re, cool tanks with water spray. Instruments Bright Instruments LTD Office: Telephone: +44 (0) 808 168 9697 Web: brightinstruments.com Burnett House, Lakeview Court Email: sales@brightinstruments.com...
Page 31
MSDS are available for professional users on request. Instruments Office: Telephone: +44 (0) 808 168 9697 Web: brightinstruments.com Bright Instruments LTD Burnett House, Lakeview Court Email: sales@brightinstruments.com Member: Instruments Ermnie Business Park, Huntingdon Cambridgeshire, PE29 6UA...
Page 32
Not in compliance with Ems number F-C,S-V published by ERMA New Zealand. inventory. Marine pollutant Instruments Office: Telephone: +44 (0) 808 168 9697 Web: brightinstruments.com Bright Instruments LTD Burnett House, Lakeview Court Email: sales@brightinstruments.com Member: Instruments Ermnie Business Park, Huntingdon Cambridgeshire, PE29 6UA...
Page 33
An environmental hazard cannot be excluded in the event of unprofessional handling or disposal. Instruments Office: Telephone: +44 (0) 808 168 9697 Web: brightinstruments.com Bright Instruments LTD Burnett House, Lakeview Court Email: sales@brightinstruments.com Member: Instruments Ermnie Business Park, Huntingdon...
Will not evaporate or dissolve Air: DO NOT allow to enter drainage systems, rivers or waterways. Instruments Office: Telephone: +44 (0) 808 168 9697 Web: brightinstruments.com Bright Instruments LTD Burnett House, Lakeview Court Email: sales@brightinstruments.com Member: Instruments Ermnie Business Park, Huntingdon Cambridgeshire, PE29 6UA...
Page 35
Regulatory information: This product is a preparation and is NOT classifi ed according to EEC Guideline. Instruments Office: Telephone: +44 (0) 808 168 9697 Web: brightinstruments.com Bright Instruments LTD Burnett House, Lakeview Court Email: sales@brightinstruments.com Member: Instruments Ermnie Business Park, Huntingdon...
Page 40
Figure 4. Coarse Advance Figure 5. Section CALE ! Control Thickness Control BT - ASK! 14. Ratchet Wheel 28. Micron Indexing Dial 17. Pawl 39. Blade Holder Clamp Knob STARLET MICROTME Bright Instruments...
Page 41
Figure 11. Trimming DO NOT SCALE ! IF IN DOUBT - ASK! 17. Pawl 12. Carrier Head 200. Rotary Main Shaft 14. Ratchet Wheel 203. Coupling Joint 17. Pawl 208. Main Shaft Coupler 20. Clamp Screw HAFT ASSEMBLY STARLET MICROTME Bright Instruments...
Page 42
Figure 16. Pawl Adjustment 16. Spacer 73. Pawl Spring Backing Plate 13. Pawl Arm 17. Pawl 17. Pawl 97. Screw 14. Ratchet Wheel 18. Pivot Lock Screw 19. Pawl Spring 107. Roll Pin Clearance Gap: 2mm + 0.2 - 0.0 Bright Instruments...
(Refer to Fig.11) handwheel version and replace the following Drive Shaft Coupler part numbers in Fig 3, 6, 29, 30, 31, 32, 33, 39, Drive Mechanisim 41, 47, 74, 91, 92 and 95. Bright Instruments DO NOT SCALE! IF IN DOUBT ASK!