Specifications; Pulse Oximeter; Principle Of Operation; Spo (Oxygen Saturation) - Novametrix Medical Systems 515B Service Manual

Hide thumbs Also See for 515B:
Table of Contents

Advertisement

10
10.1
10.1.1
10.1.2
10.1.3
10.1.4
10.1.5
10.2
10.2.1
Rev. 01

Specifications

Below are specifications for the Novametrix Model 515B/C Pulse Oximeter. These
specifications are listed for informational purposes only, and are subject to change without
notice.

Pulse Oximeter

Principle of Operation

• Red/Infrared absorbtion
SpO
(Oxygen Saturation)
2
• Range: 0-100%
• Accuracy: (for 1 standard deviation or 68% of sample distribution)
± 2% SpO
(for 80-100% SpO
2
unspecified for 0-79%
• Display Resolution: 1%
• Averaging: 8 seconds
• Audio: Pitch of pulse tone varies with SpO

Pulse Rate

• Range: 30-250 beats per minute (bpm)
• Accuracy: (1 standard deviation), ± 1% of full scale
• Display Resolution, 1 bpm
• Averaging time: 8 seconds

Sensors

• Reusable Y-Sensor™ (can be sterilized and used with all patient populations) and
reusable adult finger sensor

Plethysmogram (Model 515C only)

• Pulsatile waveform

General Specifications

Alerts

• Limits: Automatic and adjustable limits for SpO
• Audio: Adjustable volume, 2 min. silence or OFF (LED indicators)
• Visual: Flashing numerics upon violated limit(s) & red "Alert Bar" with limit (high or
low) violation indicator
)
2
value
2
and Pulse Rate
2
Model 515B/C Service Manual
43

Advertisement

Table of Contents
loading

This manual is also suitable for:

515c

Table of Contents