Fluke 87 IV Series Service Manual page 29

True rms multimeter
Hide thumbs Also See for 87 IV Series:
Table of Contents

Advertisement

Step
Function
42
AC uA
43
DC mA
44
DC mA
45
DC mA
46
DC mA
47
DC uA
48
DC uA
49
DC uA
50
DC uA
51
AC A
52
AC A
53
DC A
54
DC A
55
DC A
56
DC A
57
AC A
1.
Do not rely on autoranging for all tests. Some test steps will require manual ranging.
2.
Must allow for source and meter settling before recording reading.
3.
Wait at least 60 seconds for source and meter to settle before recording reading.
Using Relative Mode (REL) to zero residual reading (apply 0 Ω and allow UUT to settle for 45 seconds before
4.
)
pushing REL
.
Table 4. Performance Test Steps (cont.)
1
Range
Input Level
5000.0 uA
1.5 mA
50.000 mA
0 A
50.000 mA
50 mA
50.000 mA
-50 mA
400.00 mA
-329.00 mA
500.00 uA
500 uA
-500.00 uA
-500 uA
5000 uA
-5 mA
500.00 uA
Short
5.0000 A
250 mA
5.0000 A
250 mA
5.0000 A
0 A
5.0000 A
5 A
-5.0000 A
-5 A
10.000 A
-6 A
10.000 A
6 A
Frequency or
Display
Model
Lower
Limit
1 kHz
1488.2
-0.010
49.915
-49.915
-328.49
498.55
-498.55
-4987.3
-0.02
1 kHz
0.2422
20 kHz
0.2310
-0.0010
4.9740
-4.9740
-5.968
1 kHz
5.905
Performance Tests
2
Upper
Limit
1511.8
0.010
50.085
-50.085
-329.51
501.45
-501.45
-5012.7
0.02
0.2578
0.2690
0.0010
5.0260
-5.0260
-6.032
6.095
21