ThermoForma 3940 Operating And Maintenance Manual

29 cu. ft. capacity reach-in incubator

Advertisement

Quick Links

Models 3940 and 3949
29 cu. ft. Capacity
Reach-In Incubator
Operating and Maintenance Manual
Manual No: 7033940 Rev. 14

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 3940 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ThermoForma 3940

  • Page 1 Models 3940 and 3949 29 cu. ft. Capacity Reach-In Incubator Operating and Maintenance Manual Manual No: 7033940 Rev. 14...
  • Page 2 Models 3940 and 3949 ___________________________________________________________________________ Read This Instruction Manual. Failure to read, understand and follow the instructions in CAUTION this manual may result in damage to the unit, injury to operat- ing personnel, and poor equipment performance. Contains Parts and Assemblies...
  • Page 3 Models 3940 and 3949 ______________________________________________________________________________Safety Important operating and/or maintenance instructions. Read the accompanying text carefully. Ce symbole attire l'attention de l'utilisateur sur des instructions importantes de fonctionnement et/ou d'entretien. Il peut être utilisé seul ou avec d'autres symboles de sécurité. Lire attentivement le texte d'accompagnement.
  • Page 4 Models 3940 and 3949 _________________________________________________________________________Service...
  • Page 5: Table Of Contents

    Models 3940 and 3949 ___________________________________________________________________Table of Contents Table of Contents Section 6 - Specifications ......6 - 1 Section 1 - Receiving .
  • Page 6: Section 1 - Receiving

    Models 3940 and 3949 ________________________________________________________________Installation and Set-Up Section 1 - Receiving 1.1 Preliminary Inspection 1.3 Responsibility for Shipping Damage This incubator was thoroughly inspected and carefully For products shipped F.O.B. Marietta, Ohio, the responsi- packed prior to shipment and all necessary precautions were bility of Thermo Forma ends when the merchandise is loaded taken to ensure its safe arrival.
  • Page 7: Section 2 - Installation And Set-Up

    Models 3940 and 3949 ________________________________________________________________Installation and Set-Up The water inlet is the 1/8” FPT connection located on the Section 2 - Installation and Set-Up rear top left corner of the incubator. Water inlet pressure must not exceed 40 PSI. a manual shut-off valve should be installed between the main water supply and the incubator.
  • Page 8: Remote Data Output

    Note: When the chamber temperature control setpoint is For Model 3940, plug the provided 10 ft. power cord with changed, the overtemp safety thermostat must be reset to the NEMA 14-20 plug into the power outlet connection on the accommodate the change.
  • Page 9: Setting The Undertemp Safety Thermostat

    Models 3940 and 3949 ________________________________________________________________Installation and Set-Up If the battery is weak or not connected, the green LED will 2.7 Setting the Undertemp Safety Thermostat flash. If power is lost to the cabinet, the LED will also flash. When replacing the 9-volt battery, use only an alkaline style Allow the chamber temperature and humidity to stabilize, battery.
  • Page 10: Offset Calibration

    Models 3940 and 3949 ________________________________________________________________Installation and Set-Up 11. Press the Mode key once, then the Up Arrow once. 2.9 Offset Calibration “InPt” will appear in the upper display, and “set” will be in the lower display. Press the Mode key until “CAL 1”...
  • Page 11 Models 3940 and 3949 ________________________________________________________________Installation and Set-Up 3940 (WATLOW 982) CONFIGURATION RECORD CUSTOMER: JOB NUMBER: UNT SERIAL NUMBER: CONTROL TYPE: Temperature PREPARED BY: DATE COMPLETED BY: DATE Switch Configuration: Main Boards Input 1 Board Output 3 Option board Jumper switch setting:...
  • Page 12 Models 3940 and 3949 ________________________________________________________________Installation and Set-Up 3940 (WATLOW 982) CONFIGURATION RECORD CUSTOMER: JOB NUMBER: UNT SERIAL NUMBER: CONTROL TYPE: Humidity PERPARED BY: DATE: COMPLETED BY: DATE: Switch Configuration: Main Boards Input 1 Board Output 3 Option board Jumper switch setting:...
  • Page 13: Honeywell Recorder (Optional)

    Models 3940 and 3949_________________________________________________________________Start-Up and Operation Section 3 - Start Up and Operation 2.10 Honeywell Recorder (optional) The Honeywell, DR 4500 Recorder is a one to four-chan- nel microprocessor-based, circular chart recorder. The recorder is capable of recording both temperature and humidity and printing alphanumeric chart data on blank heat- sensitive chart.
  • Page 14 Models 3940 and 3949_________________________________________________________________Start-Up and Operation Overtemp Safety Control, Indicator Light, and Audible Dehumidify Switch and Indicator (Figure 3-3) Alarm (Figure 3-2) The dehumidify switch will enable the refrigeration The overtemp safety thermostat should be set system if the refrigeration switch is currently off. When slightly above the operating temperature of the incubator.
  • Page 15: Setting Incubator Operating Temperature

    Models 3940 and 3949_________________________________________________________________Start-Up and Operation 3.2 Setting the Incubator’s Operating Temperature a. Removing the mechanical lockout (Figure 3-5) Make sure the unit is turned The Watlow temperature off. Press Press controller’s upper numerical Wearing a grounding wrist display shows the actual temperature strap or maintaining constant con- inside the incubator chamber.
  • Page 16 3940 (WATLOW 982) CONFIGURATION RECORD CUSTOMER: JOB NUMBER: UNT SERIAL NUMBER: CONTROL TYPE: Temperature PREPARED BY: DATE COMPLETED BY: DATE Switch Configuration: Main Boards Input 1 Board Output 3 Option board Jumper switch setting: Form B Software Configuration: Operations Menus...
  • Page 17 3940 (WATLOW 982) CONFIGURATION RECORD CUSTOMER: JOB NUMBER: UNT SERIAL NUMBER: CONTROL TYPE: Humidity PERPARED BY: DATE: COMPLETED BY: DATE: Switch Configuration: Main Boards Input 1 Board Output 3 Option board Jumper switch setting: Form B Software Configuration: Operations Menus...
  • Page 18 HONEYWELL TRULINE CONFIGURATION RECORD SHT 1 OF 4 CUSTOMER: JOB NUMBER: UNIT SERIAL #: CONTROL TYPE: PREPARED BY: DATE: COMPLETED BY: DATE: GROUP FUNCTION VALUE OR GROUP FUNCTION VALUE OR PROMPT PROMPT SELECTION PROMPT PROMPT SELECTION TUNING1 PROP BD or CHART CHRTSPD 7 DAY...
  • Page 19 HONEYWELL TRULINE CONFIGURATION RECORD SHT 2 OF 4 GROUP FUNCTION VALUE OR GROUP FUNCTION VALUE OR PROMPT PROMPT SELECTION PROMPT PROMPT SELECTION PEN4 PEN4 DISABLE INPUT4 DECIMAL PEN4IN UNITS CHART4HI ENGUNITS CHART4LO IN4TYPE PEN4ON XMITTER PEN4OFF IN4HI MAJORDIV IN4LO MINORDIV CUTOFF4 RNG4TAG INPTCOMP...
  • Page 20 HONEYWELL TRULINE CONFIGURATION RECORD SHT 3 OF 4 GROUP FUNCTION VALUE OR GROUP FUNCTION VALUE OR PROMPT PROMPT SELECTION PROMPT PROMPT SELECTION CONTROL2 PID SETS ALARMS A1S1 VAL SW VALUE A1S2 VAL SP SOURC A1S1 TYPE NONE RATIO A1S2 TYPE NONE BIAS A1S1 HL...
  • Page 21 HONEYWELL TRULINE CONFIGURATION RECORD SHT 4 OF 4 GROUP FUNCTION VALUE OR GROUP FUNCTION VALUE OR PROMPT PROMPT SELECTION PROMPT PROMPT SELECTION EVNT MSG EVENT 1 NONE MESSAGE 1 POSITION 1 EVENT 2 NONE MESSAGE 2 POSITION 2 EVENT 3 NONE MESSAGE 3 POSITION 3...
  • Page 22 Avoid spraying cleaner on the CO sensor. • • Use sterile, distilled or demineralized water. Do not use powdered gloves for tissue cultures. Millcreek Road, Box 649 • Marietta, Ohio 45750 USA • 740-373-4763 http://www.service@thermoforma.com USA and Canada 888-213-1790 • Telefax: 740-373-4189 •...
  • Page 23 Incubators ______________________________________________________________________________ Preventive Maintenance Preventive Maintenance for 3911/3920/3940/3980 Series Incubators Refer to Action Daily Weekly Yearly Manual Section Inspect door latch, hinges and door gasket seal. 3.5 (3911), 3.4 (3920), Check the air exchange ventilator caps for adjustment; open or closed as required 3.5 (3940), 3.17 (3980)
  • Page 24: Air Exchange Ventilator Caps

    Models 3940 and 3949_________________________________________________________________Start-Up and Operation Section 4 - Routine Maintenance Press the Mode key to scroll through the menus until the upper and lower displays show “2 Loc” (Figure 3-11). Press the down arrow key until the 2 becomes 0 (zero) as shown in 4.1 Cleaning the Incubator...
  • Page 25: Section 5 - Service

    Models 3940 and 3949 _____________________________________________________________________________Service Section 5 - Service 11. Pull the entire assembly out of the panel, and remove the unit. CAUTION 12. Replace the thermostat and probe. Contains Parts and Assemblies Note: Reseal the probe access hole with Permagum and retie...
  • Page 26: Removing The Top Section

    Models 3940 and 3949 _____________________________________________________________________________Service Note: When raising the top section, notice the gasket seal at the 5.5 Removing the Top Section top opening of the incubator chamber. The gasket must be cor- rectly positioned when reinstalling the top on the chamber.
  • Page 27: Checking The Steam Generator Safety Thermostat Calibration

    Models 3940 and 3949 _____________________________________________________________________________Service 5.7 Cleaning and Adjusting the Humidity Steam 7. Carefully pry off the brass top of the float control reser- Generator, P/N 505087 voir. Clean and disinfect it and the connecting tube between the reservoir and the steam generator being Depending upon the quality of water used in the humidifi- careful to not damage the float assembly.
  • Page 28 Models 3940 and 3949 _____________________________________________________________________________Service Section 6 - Specifications 4. Using a clamp-on ammeter, establish that current is flowing in the blue wire going to the heater connectors. Refer to Figure 5-2. Temperature Control ±0.1°C @ +37°C (98.6°F) 5. Proper current is about 3.6 amps. If the current is not Range 0°C (32°F) to +60°C (140°F)
  • Page 29 Models 3940 and 3949 _________________________________________________________________________Spare Parts Section 7 - Spare Parts Humidity Part No. Description Control ±5% RH Range Above ambient to 95% 290163 RTD Temperature Sensor Sensor Bulk polymer 230066* Fuse, Ceramic 10A 350V Controller Electronic, direct set in % RH...
  • Page 30 Models 3940 and 3949 _________________________________________________________________________Spare Parts 7 - 2...
  • Page 31 Models 3940 and 3949 _________________________________________________________________________Spare Parts MODEL 3949 - 220V MAINS DISCONNECT 1/8” FPT WATER INLET BACK VIEW 7 - 3...
  • Page 33 Models 3940 and 3949 __________________________________________________________________Electrical Schematics 9 - 1...
  • Page 34 Models 3940 and 3949 __________________________________________________________________Electrical Schematics 9 - 2...
  • Page 35 Models 3940 and 3949 __________________________________________________________________Electrical Schematics 9 - 3...
  • Page 36 Models 3940 and 3949 __________________________________________________________________Electrical Schematics 9 - 4...
  • Page 37 Models 3940 and 3949 __________________________________________________________________Electrical Schematics 9 - 5...
  • Page 38 Models 3940 and 3949 __________________________________________________________________Electrical Schematics 9 - 6...
  • Page 42 Thermo Forma Millcreek Road, P.O. Box 649 Marietta, Ohio 45750 U.S.A. Telephone (740) 373-4763 Telefax (740) 373-4189...