PULSE-OXIMETER
PulsefitĀ®
BO-750/BO-750BT
Instructions

Table of contents

General information ................................................................ 2
Warning and precaution ........................................................... 3
Are the following items included? ............................................. 5
Inserting battery ...................................................................... 6
Making measurement ............................................................... 8
Performing data communication (BO-750BT) ......................... 12
Error displays and troubleshooting ......................................... 14
Care and maintenance ............................................................ 16
Specifications ........................................................................ 17
Technical description ............................................................. 18
Warranty and service ............................................................. 20

Advertisement

Table of Contents
loading

Summary of Contents for Nissei Pulsefit BO-750

  • Page 1: Table Of Contents

    PULSE-OXIMETER Pulsefit® BO-750/BO-750BT Instructions Table of contents General information ..............2 Warning and precaution ............3 Are the following items included? ..........5 Inserting battery ..............6 Making measurement ............... 8 Performing data communication (BO-750BT) ......12 Error displays and troubleshooting ......... 14 Care and maintenance ............
  • Page 2: General Information

    General information BO-750 is designed to measure %SpO and pulse rate accurately, quickly and simply at a fingertip using high precision optical components and sophisticated analogue-digital circuits. Haemoglobin in blood turns to oxyhaemoglobin in the lungs to carry oxygen to body cells and turn itself to deoxyhaemoglobin after transferring oxygen to the body cells and returning to the lungs.
  • Page 3: Warning And Precaution

    Warning and precaution Do not use BO-750 along with, near, in, or for, • MRI (MR), electrosurgical unit, defibrillator, mobile phone, RF communication equipment, or hyperbaric oxygen treatment devices • an explosive environment such as where flammable anaesthetics exist or inside oxygen chamber • infant or neonate (BO-750 is designed for adults.) Measurement can be affected by or unreliable readings may result from • lotions, nail polish and unclean fingertip...
  • Page 4 Before starting measurement, make sure that • both LED and photo sensor surfaces are clean and clear. • finger is clean. Do not use BO-750 on injured or wounded fingers. • batteries have enough power for long term measurement. Exhausted batteries may cause unexpected interruption of measurement during operation.
  • Page 5: Are The Following Items Included

    Are the following items included? Check if the following items are included. If anything is missing, contact your distributor. Pulse oximeter Photo sensor Finger insertion section Display Light source Battery cover Light shield Display switching button Clip Hole for strap attachment Be careful not to make a mistake with the strap attachment.
  • Page 6: Inserting Battery

    Inserting battery Slide the battery cover at the back of the device in the direction of the arrow to open. Put in the single AAA alkaline battery (LR03) as shown on the device and close the battery cover. When inserting or removing the battery, push the negative side of the battery against the spring.
  • Page 7 About the battery Replace the battery when the battery mark ( appears. Measurement cannot be taken when the indicator changes from flashing to lit. The battery mark which appears when all indicators are lit does not mean that the battery is due for replacement. • Dispose used batteries in a proper manner in accordance with the regulations of each municipality.
  • Page 8: Making Measurement

    Making measurement Measurements cannot be taken correctly in the following situations Hand and device are cold When cold, blood vessels constrict and blood flow deteriorates, making measurement impossible. • If your fingertip is cold, warm your finger by massaging it to improve blood flow first before measuring.
  • Page 9 Hold the clip on the device, open the finger insertion section, and fit it on your finger so that the display is on the fingernail side. Insert your finger fully all the way in so that it contacts the light source and photo sensor on the inside of the device.
  • Page 10 Amplitude intensity Pulse rate The intensity of the amplitude of the detected pulse is /min indicated in 4 levels. Unit of pulse rate (bpm) Stability mark Bluetooth mark Indicates that the measurement *BO-750BT only condition is favorable (See page 12) Blood oxygen saturation %SpO2 level...
  • Page 11 Measurement is not possible if the backlight turns orange and appears. For information about the error display, see page 14. When the device is removed, the measurement ends automatically and the power supply is switched off after about 8 sec. Auto display switching Invert the device to automatically switch the display.
  • Page 12: Performing Data Communication (Bo-750Bt)

    • Android 4.3 or higher (equipped with Bluetooth 4.0 or higher) 2. Download NISSEI HealStyle from the Apple App Store or Google Play Store. 3. S t a r t N I S S E I H e a l S t y l e a n d p e r f o r m u s e r...
  • Page 13 Using the app, the BO-750BT can manage the measurement results with a smart phone. See the NISSEI website for more information on how to run the app. http://www.nissei-kk.co.jp/english/ • If the connection is not smooth, remove your finger from the main unit to end the measurement once before re-connecting.
  • Page 14: Error Displays And Troubleshooting

    Error displays and troubleshooting Indication Cause Countermeasure Signal is not detected Re-install the device so that your finger is in contact with the light source and photo sensor. An object is stuck in the Remove the stuck object. finger insertion section P u l s e r a t e c a n b e More information from the pulse measured but the blood wave detected is required in the...
  • Page 15 Indication Cause Countermeasure Initial test error: A finger Check that there is nothing in the was inserted or an object insertion section, and start the was stuck in the finger measurement after the initial test insertion section during error display disappears. the initial test.
  • Page 16: Care And Maintenance

    Care and maintenance Calibration No calibration or adjustment is required for the life of the product. Cleaning Wipe stain or dirt with soft cloth damped with neutral detergent or isopropyl alcohol. Remove batteries before cleaning. Handling and Storage Because the unit includes precision parts, care should be taken to avoid extreme temperature variations, humidity, shock, dust, and direct sunlight.
  • Page 17: Specifications

    Information on the range of the peak wavelengths and maximum optical output power of the light emitted by the pulse oximeter can be useful to a clinicians performing photodynamic therapy. Contact Nissei Healthcare (UK) Ltd., our EC representative for any further information.
  • Page 18: Technical Description

    Technical description BO-750 complies with the EMC, electromagnetic compatibility, standard, IEC60601-1-2. Refer to the tables below for specific information regarding compliance to the standard. BO-750, as a medical electrical equipment, needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided below. Portable and mobile RF communications equipments can affect the device. The use of accessories other than those specified in this manual may result in increased emissions or decreased immunity of the device. BO-750 should not be used adjacent to or stacked with other equipment.
  • Page 19 Table 4 - Guidance and manufacturer’s declaration - electromagnetic immunity - BO-750 is intended for use in the electromagnetic environment specified below. The customer or the user of BO-750 should assure that it is used in such an environment. Immunity test I E C 6 0 6 0 1 t e s t Compliance Electromagnetic environment - guidance level level Portable and mobile RF communications equipment should be...
  • Page 20: Warranty And Service

    Warranty and service NISSEI warrants the product for two years from the date of purchase for functionality and accuracy without charge for inspection, adjustment, repair and labour. Evidence of date of purchase is required for warranty. However, this warranty does not cover defects resulting from, damage caused by wear or misuse, damage caused by unauthorised repair or modification or damage caused by natural disaster, violent action or war.