User manual <English>

Table of contents

General information ................................................................ 2
Warning and precaution ........................................................... 3
Are the following items included? ............................................. 5
Inserting battery ...................................................................... 6
Making measurement ............................................................... 8
Performing data communication (BO-750BT) ......................... 12
Error displays and troubleshooting ......................................... 14
Care and maintenance ............................................................ 16
Specifications ........................................................................ 17
Declaration for Directive 2014/53/EU: .................................... 19
Warranty and service ............................................................. 20

Advertisement

Table of Contents
loading

Summary of Contents for Nissei pulsFit BO-750

  • Page 1: Table Of Contents

    PULSE-OXIMETER BO-750/BO-750BT User manual <English> Table of contents General information ..............2 Warning and precaution ............3 Are the following items included? ..........5 Inserting battery ..............6 Making measurement ............... 8 Performing data communication (BO-750BT) ......12 Error displays and troubleshooting ......... 14 Care and maintenance ............
  • Page 2: General Information

    General information Intended Use BO-750 is designed to measure %SpO and pulse rate accurately, quickly and simply at a fingertip using high precision optical components and sophisticated analogue-digital circuits. The product is a designed for the professional use in medical institutions. The product is not intended to be used either outside the healthcare facility, during professional transportation of a patient outside the healthcare facility or home healthcare environment.
  • Page 3: Warning And Precaution

    Warning and precaution Do not use BO-750 along with, near, in, or for, • MRI (MR), electrosurgical unit, defibrillator, mobile phone, RF communication equipment, or hyperbaric oxygen treatment devices • an explosive environment such as where flammable anaesthetics exist or inside oxygen chamber •...
  • Page 4 Before starting measurement, make sure that • both LED and photo sensor surfaces are clean and clear. • finger is clean. Do not use BO-750 on injured or wounded fingers. • batteries have enough power for long term measurement. Exhausted batteries may cause unexpected interruption of measurement during operation.
  • Page 5: Are The Following Items Included

    Are the following items included? Check if the following items are included. If anything is missing, contact your distributor. Pulse oximeter Photo sensor Finger insertion section Display Light source Battery cover Light shield Display switching button Clip Hole for strap attachment Make sure that the strap is attached to the correct position.
  • Page 6: Inserting Battery

    Inserting battery Slide the battery cover at the back of the device in the direction of the arrow to open. Put in one AAA alkaline battery (LR03) as shown on the device and close the battery cover. When inserting or removing the battery, push the negative side of the battery against the spring.
  • Page 7 About the battery Replace the battery when the battery mark ( appears. Measurement cannot be taken when the mark changes from blinking to remaining indication. The battery mark which appears when all indicators are lit at the initial test does not mean that the battery is due for replacement. •...
  • Page 8: Making Measurement

    Making measurement Measurements cannot be taken correctly in the following situations Hand and device are cold When they are cold, blood vessels are constricted and blood flow is deteriorated. It causes the disturbance of measurement. • If your fingertip is cold, warm your finger by massaging etc. to improve the blood flow before making a measurement.
  • Page 9 Hold the device and pinch it like shown in the picture to open the finger insertion part and fit it on your finger with the display side of the device on the fingernail side. Insert your finger to the end so that the fingertip is centered at the light source and photo sensor inside the device.
  • Page 10 Amplitude intensity Pulse rate The intensity of the pulse amplitude is indicated in 4 /min levels. Unit of pulse rate (bpm) Stability mark Bluetooth mark Indicates that the measurement *BO-750BT only condition is favorable (See page 12) Blood oxygen saturation %SpO level Unit of blood oxygen...
  • Page 11 Measurement is not possible if the backlight turns to orange appears. For information about the error display, see page 14. When the device is removed, the measurement ends automatically and the power supply is switched off after about 8 sec. Auto display switching Invert the device to automatically switch the display.
  • Page 12: Performing Data Communication (Bo-750Bt)

    * Please make sure that the mobile computing platform (Smartphone etc.) connected to BO-750BT conforms to applied safety, EMC, and Wireless communication standards. 2. Download NISSEI HealStyle from the Apple App Store or Google Play Store. 3. Start NISSEI HealStyle and register the user information.
  • Page 13 The management of the measurement results with Smartphone is possible for BO-750BT by using the app. See the NISSEI website for more information on how to run the app. http://www.nissei-kk.co.jp/english/ • If you register the connection ID in the app, you will not need to enter the connection ID the next time you want to perform data communication.
  • Page 14: Error Displays And Troubleshooting

    Error displays and troubleshooting Indication Cause Countermeasure Signal is not detected * Re-install the device so that your finger is in contact with the light source and photo sensor. An object is stuck in the Remove the stuck object. finger insertion section Pulse rate More information from the pulse...
  • Page 15 Indication Cause Countermeasure A finger was inserted or Check that there is nothing in the an object was stuck in the insertion section, and start the finger insertion section measurement after the initial test during the initial test. error display disappears. Sensor is not working Contact your distributor.
  • Page 16: Care And Maintenance

    Care and maintenance Calibration No calibration or adjustment is required for the life of the product. Cleaning/Disinfection Remove batteries before cleaning or disinfecting the device. Wipe stain or dirt with soft cloth damped with neutral detergent or isopropyl alcohol, and then wipe it off with another soft cloth moistened with cold or tepid water.
  • Page 17: Specifications

    Specifications Operating principle Double wavelength lights absorption method Measurement result display Average of 4 pulses (automatic update at every pulse) %SpO 0 - 100% Measurement Range Pulse rate 30 - 240bpm %SpO ±2% (70%≤SpO ≤100%) * Accuracy Pulse rate ±3%/±1 digit (30 - 240bpm) * Power source One AAA alkaline (LR03) battery Rated voltage...
  • Page 18: Technical Description For Electromagnetic Disturbances

    Technical description for electromagnetic disturbances BO-750/BO-750BT complies with the Electromagnetic Disturbances standard, IEC60601-1-2:2014. As a medical electrical equipment, special precautions regarding the electromagnetic disturbances shall be taken at usage of the device according to the information provided below. • The device is not intended for use in environments where the intensity of electromagnetic disturbance is high, such as near active HF surgical equipment and MRI (magnetic resonance imaging) equipment etc.
  • Page 19: Declaration For Directive 2014/53/Eu

    Declaration for Directive 2014/53/EU: Hereby, NIHON SEIMITSU SOKKI CO., LTD. declares that BO-750BT is in compliance with Directive 2014/53/EU. The full text of the EU declaration of conformity is available at the following internet address: http://www.nissei-kk.co.jp/english/medical/pulse/ - 19 -...
  • Page 20: Warranty And Service

    Warranty and service NISSEI warrants the product for two years from the date of purchase for functionality and accuracy without charge for inspection, adjustment, repair and labour. Evidence of date of purchase is required for warranty. However, this warranty does not cover defects resulting from, damage caused by wear or misuse, damage caused by unauthorised repair or modification or damage caused by natural disaster, violent action or war.