CHANGE OF COMPRESSOR DELAY TIME

The delay time of high and low stage side compressor can be changed to reduce the load on the power
line and to facilitate the start-up (reset) of the freezer after power failure.
The example in the table is based on the assumption that the delay time is changed to 4 minutes.
(The delay time is set in 2 minutes at the factory.)
Note:
• The delay time should be same for high stage side and low stage side compressors.
• The setting range for delay time is between 2 and 15 minutes. The cool down of chamber temperature
may be slow when the setting of delay time is over 5 minutes, depending on the installation environment.
There is no need changing the delay time when the capacity of power source is enough.
Table 8. Changing procedure for delay time (change from 2 minutes to 4 minutes)
Description of operation
1
Press numerical value shift key for
2
5 seconds.
Set the figure to F05 with the
3
numerical value shift key.
4
Press SET key.
Set the figure to 004 with the
5
numerical value shift key.
6
Press SET key.
• The compressor starts to operate with the delay time set by the above procedure at the time of power
on or after power failure. For the low stage side compressor, however, its start up may be behind the
delay time because the compressor is affected by the chamber temperature and temperature of cascade
condenser installed in the freezer.
Key operated
The current chamber temperature is
----
displayed.
The first digit is flashed.
When pressed, the figure of settable
digit changes.
The current delay time is displayed.
SET
The first digit is flashed.
When pressed, the figure of the first
digit changes.
The delay time is memorized and
SET
the current chamber temperature is
displayed.
23
Indication after operation