8. It is recommended that every five (5) years the electric heater be replaced.
9. It is recommended that once a month a qualified person checks the solid-state relay
to confirm the solid-state relay is running correctly to avoid breakdowns.
Note: The electrical heater will continuously heat in the event the solid-state relay
breaks down, leading to the potential of a serious hazard.
10. Immediately power 'OFF' the electrical heater if the electrical heater over temperature
alarm alerts / sounds. If the above is not resolved fully power 'OFF' and unplug from the
main power supply and arrange a service technician to investigate and repair.
! ATTENTION (May affect operational performance or service life)
11. The overall safe operation of the instrument is the responsibility of the owner of the
instrument and their assigned operator(s), who in turn are responsible for ensuring the
user manual guidance is applied to, optimise the safety and protection of personnel and
the Instrument before, during and after lyophilizer operation.
12. Timely maintenance of the instrument MUST be maintained to ensure continued safe
operation, good instrument condition thus optimising the instruments service life.
13. Only accredited and qualified professional repair technicians can open the instrument or
conduct required repairs. Persons performing repairs on the instrument other than those
selected or approved by the Company shall operate to void any warranty contained
hereinabove for the product.