The finger pinch gauge can be used to measure pinch strength. It is
calibrated in pounds and kilograms of force.
Apply pinch force at the pinch groove while holding the pinch gauge
between your thumb and finger(s). When force is applied farther
toward the tip the reading will be slightly higher. When force is applied
farther toward the rear the reading will be slightly lower.
The gauge must be "zeroed" before each pinch test. Grasp the knurled
ring of the dial indicator and rotate it until the zero on the dial indicator
is directly under the black pointer.
The red maximum pointer must be reset before each pinch test.
Rotate the small knurled knob on top of the dial indicator in a
counterclockwise direction until it rests against the black pointer at the
zero marking. The red maximum pointer will remain at the subject's
maximum reading until it is reset.
Use the pinch gauge to perform the three basic pinch tests:
• Tip Pinch - thumb tip to index fingertip
• Key Pinch -thumb pad to lateral aspect of middle phalanx of index
finger
• Palmar Pinch - thumb pad to pads of the index and middle fingers
Standard Hydraulic
Pinch Gauge
(12-0235)

Advertisement

loading

Summary of Contents for Baseline 12-0235 Standard

  • Page 1 Hydraulic Pinch Gauge Instruction Manual Standard 12-0235 LiTE® 12-0226 12-0222 HD® 12-0239 HiRes® ER™ HiRes® 12-0228 Standard Hydraulic ER™ Digital 12-0237 Pinch Gauge (12-0235) The finger pinch gauge can be used to measure pinch strength. It is calibrated in pounds and kilograms of force. Apply pinch force at the pinch groove while holding the pinch gauge between your thumb and finger(s).
  • Page 2 Grip Dynamometer Norms for Adult Grip Strength (kgs) Average Performance of all Subjects on Grip Strength (kilograms) - Test results Women hand mean low-high mean low-high dominant 54.88 9.34 1.72 41.3 - 75.7 31.93 6.58 1.27 20.9 - 43.1 20-24 non-dominant 47.40 9.89...
  • Page 3 TIP PINCH NORMS (KGS) NORMS FOR ADULT PINCH STRENGTH Tip Pinch strength performance of all subjects (kilograms) women hand mean low-high mean low-high dominant 8.16 1.36 4.9 - 10.4 5.03 0.95 3.6 - 7.2 20-24 non-dominant 7.71 1.04 5.4 - 14.9 4.76 0.77 3.6 - 6.3...
  • Page 4 KEY PINCH NORMS (KGS) NORMS FOR ADULT PINCH STRENGTH Key Pinch strength performance of all subjects (kilograms) women hand mean low-high mean low-high dominant 11.79 1.59 9.5 - 15.4 7.98 0.91 6.4 - 10.4 20-24 non-dominant 11.25 1.54 8.6 - 14.1 7.35 0.95 5.9 - 10.4...
  • Page 5 PALMAR PINCH NORMS (KGS) NORMS FOR ADULT PINCH STRENGTH Palmar Pinch strength performance of all subjects (kilograms) women hand mean low-high mean low-high dominant 12.07 2.40 8.2 - 20.4 7.80 1.04 6.4 - 10.4 20-24 non-dominant 11.66 2.63 6.8 - 19.1 7.39 1.27 5.0 - 10.9...