Figure 2-1: PGC5000 Electronics compartment open

Legend: PGC5000 Electronics compartment open
Item
Description
1
DBDID amplifier assembly
2
DBDID power supply board
3
Flange nut
2.1

Detector removal

WARNING – Bodily injury. High voltage present. Make sure that power to the PGC5000 has
been disconnected before removing the detector.
To remove the detector from the PGC5000, perform the following steps:
Disable the isothermal oven temperature zone and allow the oven to cool to room temperature.
Power down the oven.
Disconnect the column, cell vent line and the reactor gas inlet from the cell body. See items 1-
3 in
Figure 2-2.

Figure 2-2: Disconnect lines

1
2
3
6 | DB DI D | 8 92 J0 09 MN A E
1
2
3