Concha-Column Performance Characteristics

Water Capacity:
Standard
Low Compliance
Comfort-Flo Low-Compliance:
Minimal Compliance column:
Column Compliance:
Standard:
Low-Compliance:
Comfort-Flo Low-Compliance:
Minimal Compliance:
Column Compressible Volume:
Standard:
Low-Compliance:
Comfort-Flo Low-Compliance:
Minimal Compliance:
Note: The amount of water in a Standard Concha-Column decreases as water is used,
increasing the compressible volume as water is consumed.

General System Characteristics

Pressure Drop or Resistance to Flow:
System Circuits, Concha Column and Water Reservoir shall be in compliance to ISO 8835-2 for
Pressure Drop (Resistance to Flow). See individual circuit "Instructions For Use" for Resistance
to Flow data.
System Pressures:
The ConchaTherm circuits are designed to exceed the ISO/CD 5367 standard of 60 cm/H
(6 +/- 0.3 kPa), and are tested to perform at 90 cm/H
System Leakage:
System leakage (Circuit, Concha-Column and Water Reservoir) will have a combined leakage of
less than 25 ml/minute at 60 cm/H
Noise Levels:
The humidifier noise level under normal operation conditions shall be less than 50 dB measured
1 meter from the device, except while audible alarms are active.
Electrical Characteristics:
Input Voltage and Power:
Leakage Current:
Dielectric Withstand:
Temperature Control:
Power Fuses:
Thermal Fuse:
Operating Environment Temperature:
Operating Temperature:
Note: Operating the humidifier at the extremes of the operating temperature may affect the
temperature output of the device and/or cause nuisance alarms. The humidifier works ideally
at a room temperature of 22 to 26 °C (72 to 79 °F).
193 ± 10 ml with full reservoir
40 ± 5 ml
40 ± 5 ml
40 ± 5 ml
1.12 ml/cm H
2.51 ml/cm H
0.25 ml/cm H
0.25 ml/cm H
0.2 ml/cm H
103 ml with full reservoir
268 ml with reservoir at "replace" mark
248 ml ± 12 ml
248 ml ± 12 ml
109 ml
O.
2
O (6 +/- 0.3 kPa) per ISO 5367, Annex D.
2
100-120 V ~, 50-60 Hz, 2.6 A (MAX)
200-240 V ~, 50-60 Hz, 1.9 A (MAX)
Less than 100 micro-amps
1,250 volts minimum for one (1) minute
Dual thermistor
Two 4 amp, 5 x 20mm time delayed, 250 V ~
152 °C non-re-settable
140 °C re-settable
18 to 29 °C (68 to 85°C)
O with full reservoir
2
O with reservoir at "replace" mark
2
O
2
O
2
O
2
O
2