LM-ProPower Technical manual

Advertisement

Table of Contents
loading

Summary of Contents for LM ProPower UltraLED

  • Page 1 LM-ProPower Technical manual...
  • Page 2 Web: www.lm-dental.com Copyright Copyright 2014 LM-Instruments Oy. All rights reserved. The contents of this manual may be changed without notice. No part of this manual may be reproduced in any form or by any means without permission in writing from LM-Instruments Oy.
  • Page 3: Table Of Contents

    Content Troubleshooting 1.1 UltraLED with tap water connector 1.2 UltraLED with medicament/water bottle 1.3 AirLED with tap water connector 1.4 AirLED with medicament/water bottle 1.5 CombiLED with tap water connector 1.6 CombiLED with medicament/water bottle Cause Diagnosis and Cause Codes Repair instructions 3.1 Removing the cover casing 3.2 Replacing the scaler handpiece tubing...
  • Page 4: Troubleshooting

    1. Troubleshooting 1.1 UltraLED with tap water connector Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 4 B. Lights are flashing on the control panel Page 4 C. No tip vibration Page 5 D.
  • Page 5 3. Adjust the water control ring on the handpiece to fully open (dots aligned). 4. Try with another tip. 5. Use the LM-ProPower Fixer to check that the water hose inside the handpiece has not stuck in the autoclave. Insert the Fixer carefully into the handpiece from the hose connector side.
  • Page 6 G. Difficulty to remove the ErgoGrip from the handpiece 1. Grab the ErgoGrip gently near the lens and at the same time twist and slide it off the handpiece 2. If the problem still remains, replace the ErgoGrip H. Handpiece is overheated during use 1.
  • Page 7: Ultraled With Medicament/Water Bottle

    1.2 UltraLED with medicament/water bottle Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 7 B. Lights are flashing on the control panel Page 7 C. No tip vibration Page 8 D.
  • Page 8 3. Adjust the water control ring on the handpiece to fully open (dots aligned). 4. Try with another tip. 5. Use the LM-ProPower Fixer to check that the water hose inside the handpiece has not stuck in the autoclave. Insert the Fixer carefully into the handpiece from the hose connector side.
  • Page 9 G. Difficulty to remove the ErgoGrip from the handpiece 1. Grab the ErgoGrip gently near the lens and at the same time twist and slide it off the handpiece 2. If the problem still remains, replace the ErgoGrip H. Handpiece is overheated during use 1.
  • Page 10: Airled With Tap Water Connector

    1.3 AirLED with tap water connector Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 10 B. Lights are flashing on the control panel Page 10 C. Difficulty to remove the ErgoGrip Page 10 D.
  • Page 11 D. Weak or no LED light. 1. Note that there are two versions of the ErgoGrip with different lenses and light pattern and intensity: “ErgoGrip LED” and “ErgoGrip FocusLED”. 2. Check that lens in the ErgoGrip is clear and clean. Clean or replace if necessary. 3.
  • Page 12: Airled With Medicament/Water Bottle

    1.4 AirLED with medicament/water bottle Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 12 B. Lights are flashing on the control panel Page 12 C. Difficulty to remove the ErgoGrip Page 13 D.
  • Page 13 C. Difficulty to remove the ErgoGrip from the handpiece 1. Grab the ErgoGrip gently near the lens and at the same time twist and slide it off the handpiece 2. If the problem still remains, replace the ErgoGrip D. Weak or no LED light 1.
  • Page 14: Combiled With Tap Water Connector

    1.5 CombiLED with tap water connector Type of problem Description A. Unit is not responding and no lights are lit on the control Page 14 panel B. Lights are flashing on the control panel Page 14-15 C. Difficulty to remove the ErgoGrip Page 15 D.
  • Page 15: No Tip Vibration

    2. Check that the tip is not worn out. Use a tip check card to evaluate wearing or try an unused tip. Use original LM-Instruments tips for reliable performance. 3. If you have an additional handpiece, try replacing the handpiece.
  • Page 16: Insufficient Or No Water Flow

    3. Adjust the water control ring on the handpiece to fully open (dots aligned). 4. Try with another tip. 5. Use the LM-ProPower Fixer to check that the water hose inside the handpiece has not stuck in the autoclave. Insert the Fixer carefully into the handpiece from the hose connector side.
  • Page 17: Combiled With Medicament/Water Bottle

    1.6 CombiLED with medicament/water bottle Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 17 B. Lights are flashing on the control panel Page 17-18 C. Difficulty to remove the ErgoGrip Page 18 D.
  • Page 18 2. Check that the tip is not worn out. Use a tip check card to evaluate wearing or try an unused tip. Use original LM-Instruments tips for reliable performance. 3. If you have an additional handpiece, try replacing the handpiece.
  • Page 19 4. Try with another tip. 5. Use the LM-ProPower Fixer to check that the water hose inside the handpiece has not stuck in the autoclave. Insert the Fixer carefully into the handpiece from the hose connector side. (To avoid the hose getting stuck, the water adjustment on the handpiece should be set to fully open before sterilization in autoclave.)
  • Page 20: Cause Diagnosis And Cause Codes

    Customer Service Form indicating cause code C-001 working handpiece the handpiece is faulty. and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty handpiece with the RMA number to LM for replacement.
  • Page 21 Fill in Customer Service Form indicating cause code C-003 and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty electronic unit for replacement. 4. If above actions do not solve the problem,...
  • Page 22 Cause code Repair / Remedy instructions working handpiece the handpiece is faulty. and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty handpiece for replacement. 2. If there is no Light with the working...
  • Page 23 Replace o-rings. Fill in a Customer Service Form indicating handpiece and bottle cap and connectors cause code C-021 and submit to LM fax +358 2 454 6444. when applicable). C-999 Level 1 SP 3. If replacing the o-rings does not solve the...
  • Page 24 Fill in Customer Service Form indicating cause code and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty coil for replacement. 8. If above actions do not solve the problem, C-008 Unidentified...
  • Page 25 C-014 and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty part with the RMA number to LM for replacement. 3. If replacing the AirLED tubing does not...
  • Page 26 C-015 request authorization to return the device to of air flow and submit to LM fax +358 2 454 6444. After obtaining an LM-Instruments for repair. RMA number from LM, return the entire device with accessories and the RMA number to LM for repair.
  • Page 27 C-014 and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty part with the RMA number to LM for replacement. 4. If replacing the AirLED tubing does not...
  • Page 28 Repair / Remedy instructions request authorization to return the device to no/low water flow. submit to LM fax +358 2 454 6444. After obtaining an RMA LM-Instruments for repair . number from LM, return the entire device with accessories and the RMA number to LM for repair.
  • Page 29: Repair Instructions

    3 Repair Instructions Warning! Some tests and measurements require that the power cable is plugged in with the casing removed. This is only allowed for legally authorized electricians and extra caution must be taken because of exposed live parts. 3.1 Removing the cover and casing Start with disconnecting the power cable before removing cover and casing.
  • Page 30 Unlock the tail from the connector. Make sure the tail is not twisted when reassembling the device. Unlock the handpiece holder by moving the metal plate upwards Remove the plate Twist the handpiece holder upwards and remove...
  • Page 31 Remove the support list by pulling it upwards Expand the casing and slide the chassis towards the front After sliding the chassis all the way to the front, lift it upwards with front part first. Reassemble in reverse order.
  • Page 32: Replacing The Scaler Handpiece Tubing

    3.2 Replacing the scaler handpiece tubing First remove cover and casing, see instruction section 1 Disconnect the handpiece wires, note the colors. From left: 1 - Blue, 2 - Red, 3 - Black and Green together, White at separate connector. Disconnect the water tubing (clear) from the magnetic valve (pos 2).
  • Page 33: Measuring Voltage From The Control Circuit Board To The Scaler Electronics Unit

    3.3 Measuring the voltages from the control circuit board to the scaler electronics unit Remove cover and casing, see instruction section Control Circuit Board Scaler Electronics Unit Measure the output voltage from control circuit board to scaler electronics unit. Pin 5-6 and pin 5-10 (see pictures above), reading should be 10 +/- 0,5 Vdc when running at working mode 3 and with foot control at maximum power.
  • Page 34: Replacing The Scaler Electronic Unit

    Measure the power supply to the scaler pin 1-2. Reading should be aprox. 30 Vdc. If all measurements were as above, the control circuit board is functioning properly and giving the correct control signals to the scaler electronics unit. But the scaler electronic unit is probably malfunctioning and must be replaced, see Section 4.
  • Page 35 If applicable (CombiLED), dismount the left magnetic valve coil to access the mounting screw. Loosen all screws from electrical connectors and slide out the scaler electronics unit. Reassemble in reverse order.
  • Page 36: Replacing The Compressor

    3.5 Replacing the compressor First remove cover and casing, see instruction section 1. Disconnect the yellow air tubing. Cut the cable ties. Disconnect the electrical wires at the connector on the control board (note the polarity). Old Design Replace the compressor and reassemble in reverse order. Note that the compressor should not be in contact with the chassis to avoid vibration noise and short circuiting the connectors of the compressor.
  • Page 37: Cleaning A Magnetic Valve

    3.6 Cleaning a magnetic valve First remove cover and casing, see instruction section 1. Magnetic valves function: 1: Polisher water 2: Scaler water 3: Polisher cleaning air 4: Polisher air Locate the valve according to picture above. Dismount the coil and the valve plug. Inspect and clean.
  • Page 38: Opening A Stuck Waterline In The Scaler Handpiece

    3.7 Opening a stuck waterline in the scaler handpiece First remove the scaler handpiece from the tubing. Set the water control ring to fully open (dot-dot). Gently insert the LM-ProPower Fixer tool into the handpiece from the connector side to open up the stuck water line.
  • Page 39: Cleaning Of Clogged System In Airled

    3.8 Cleaning of clogged system in AirLED Connect a small tube approx. 40mm length to the powder outlet Fill the powder chamber with citric acid* up to the max level. Put it back on the unit and run the cleaning cycle. The air pressure will flow the citric acid trough the clogged tubing.
  • Page 40: Spare Parts

    4 Spare parts 4.1 CombiLED LM-ProPower CombiLED for tap LM-ProPower CombiLED with bottle Water Code Description Amount Code Description Amount 4518111 Chassis 4517991 Bottle Holder Blind 4518121 Mounting Plate 4518111 Chassis 4518131 Attachment plate 4518121 Mounting Plate 5608741 Power Board...
  • Page 43: Ultraled Scaler

    4.2 UltraLED Scaler LM-ProPower UltraLED with bottle LM-ProPower UltraLED for tap water Code Description Amount Code Description Amount 4518011 Polisher Holder Blind 4517991 Bottle Holder Blind 4518111 Chassis 4518011 Polisher Holder Blind 4518121 Mounting Plate 4518111 Chassis 4518131 Attachment plate...
  • Page 46: Airled Polisher

    4.3 AirLED Polisher LM-ProPower AirLED with bottle LM-ProPower AirLED for tap water Code Description Amount Code Description Amount 4518111 Chassis 4517991 Bottle Holder Blind 4518121 Mounting Plate 4518111 Chassis 4518131 Attachment plate 4518121 Mounting Plate 5608741 Power Board 4518131 Attachment plate...
  • Page 49: Recommended Spare Part Stock

    4.4 Recommended spare part stock Code Description Amount UltraLED AirLED CombiLED 15608821 Foot control 2 Pcs 15610261 AirLED polisher tubing 2 Pcs 15609541 Medicament bottle 2 Pcs 15610331 AirLED powder container 2 Pcs 16003762 Foot control cable 2 Pcs 15610221 Scaler electronic unit 2 Pcs 15608751 Cover CombiLED 2 Pcs...
  • Page 50: Technical Data

    5 Technical data 5.1 Technical data CombiLED LM-Instruments Oy Manufacturer PL 88 (Norrbyn rantatie 8) FIN-21601 Parainen, FINLAND Model LM-ProPower CombiLED EN60601-1: Class 1, Type BF Classification 93/42 EU: Medical products, class IIa L x W x H 270 x 110 x 165 mm (without bottle and powder container)
  • Page 51: Ultraled Scaler

    5.2 Technical data UltraLED Scaler LM-Instruments Oy Manufacturer PL 88 (Norrbyn rantatie 8) FIN-21601 Parainen, FINLAND Model LM-ProPower UltraLED EN60601-1: Class 1, Type BF Classification 93/42 EU: Medical products, class IIa L x W x H 270 x 140 x 165 mm (without bottle)
  • Page 52: Airled Polisher

    5.3 Technical data AirLED Polisher LM-Instruments Oy Manufacturer PL 88 (Norrbyn rantatie 8) FIN-21601 Parainen, FINLAND Model LM-ProPower AirLED EN60601-1: Class 1, Type BF Classification 93/42 EU: Medical products, class IIa L x W x H 270 x 110 x 165 mm (without bottle and powder container)
  • Page 53: Warranty Terms

    6. Warranty Terms The following warranty terms apply to the sale of LM-Instrument Oy’s products (hereinafter “Products”) to a purchasing company or individual by LM-Instruments Oy (hereinafter “Manufacturer”). Manufacturer hereby warrants that the Products will be free from defects arising from faulty materials or workmanship for a period of twenty four (24) months from the date of purchase by a customer from Manufacturer’s authorized dealer (hereinafter “Authorized Dealer”).
  • Page 54: Return Shipment Policy

    3. For small equipment (i.e. power scalers, air polishers, sharpening devices as well as their parts) it is important to indicate the serial number of the device as well as the original LM- Instruments invoice number. In addition, a copy of the Authorized Dealer’s invoice to the end customer (or other proof of purchase) must be included.
  • Page 55: Customer Service Form

    Customer Service Form Please fill in and return to: LM-Instruments Oy PO Box 88 (Norrbyn rantatie 8) FI-21601 Parainen, Finland Fax +358 2 4546 444 service@lm-dental.com Customer identification Dealer name Contact person Date Phone Address Email Practice name Contact person...