Content Troubleshooting 1.1 UltraLED with tap water connector 1.2 UltraLED with medicament/water bottle 1.3 AirLED with tap water connector 1.4 AirLED with medicament/water bottle 1.5 CombiLED with tap water connector 1.6 CombiLED with medicament/water bottle Cause Diagnosis and Cause Codes Repair instructions 3.1 Removing the cover casing 3.2 Replacing the scaler handpiece tubing...
1. Troubleshooting 1.1 UltraLED with tap water connector Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 4 B. Lights are flashing on the control panel Page 4 C. No tip vibration Page 5 D.
Page 5
3. Adjust the water control ring on the handpiece to fully open (dots aligned). 4. Try with another tip. 5. Use the LM-ProPower Fixer to check that the water hose inside the handpiece has not stuck in the autoclave. Insert the Fixer carefully into the handpiece from the hose connector side.
Page 6
G. Difficulty to remove the ErgoGrip from the handpiece 1. Grab the ErgoGrip gently near the lens and at the same time twist and slide it off the handpiece 2. If the problem still remains, replace the ErgoGrip H. Handpiece is overheated during use 1.
1.2 UltraLED with medicament/water bottle Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 7 B. Lights are flashing on the control panel Page 7 C. No tip vibration Page 8 D.
Page 8
3. Adjust the water control ring on the handpiece to fully open (dots aligned). 4. Try with another tip. 5. Use the LM-ProPower Fixer to check that the water hose inside the handpiece has not stuck in the autoclave. Insert the Fixer carefully into the handpiece from the hose connector side.
Page 9
G. Difficulty to remove the ErgoGrip from the handpiece 1. Grab the ErgoGrip gently near the lens and at the same time twist and slide it off the handpiece 2. If the problem still remains, replace the ErgoGrip H. Handpiece is overheated during use 1.
1.3 AirLED with tap water connector Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 10 B. Lights are flashing on the control panel Page 10 C. Difficulty to remove the ErgoGrip Page 10 D.
Page 11
D. Weak or no LED light. 1. Note that there are two versions of the ErgoGrip with different lenses and light pattern and intensity: “ErgoGrip LED” and “ErgoGrip FocusLED”. 2. Check that lens in the ErgoGrip is clear and clean. Clean or replace if necessary. 3.
1.4 AirLED with medicament/water bottle Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 12 B. Lights are flashing on the control panel Page 12 C. Difficulty to remove the ErgoGrip Page 13 D.
Page 13
C. Difficulty to remove the ErgoGrip from the handpiece 1. Grab the ErgoGrip gently near the lens and at the same time twist and slide it off the handpiece 2. If the problem still remains, replace the ErgoGrip D. Weak or no LED light 1.
1.5 CombiLED with tap water connector Type of problem Description A. Unit is not responding and no lights are lit on the control Page 14 panel B. Lights are flashing on the control panel Page 14-15 C. Difficulty to remove the ErgoGrip Page 15 D.
2. Check that the tip is not worn out. Use a tip check card to evaluate wearing or try an unused tip. Use original LM-Instruments tips for reliable performance. 3. If you have an additional handpiece, try replacing the handpiece.
3. Adjust the water control ring on the handpiece to fully open (dots aligned). 4. Try with another tip. 5. Use the LM-ProPower Fixer to check that the water hose inside the handpiece has not stuck in the autoclave. Insert the Fixer carefully into the handpiece from the hose connector side.
1.6 CombiLED with medicament/water bottle Type of problem Description A. Unit is not responding and no lights are lit on the control panel Page 17 B. Lights are flashing on the control panel Page 17-18 C. Difficulty to remove the ErgoGrip Page 18 D.
Page 18
2. Check that the tip is not worn out. Use a tip check card to evaluate wearing or try an unused tip. Use original LM-Instruments tips for reliable performance. 3. If you have an additional handpiece, try replacing the handpiece.
Page 19
4. Try with another tip. 5. Use the LM-ProPower Fixer to check that the water hose inside the handpiece has not stuck in the autoclave. Insert the Fixer carefully into the handpiece from the hose connector side. (To avoid the hose getting stuck, the water adjustment on the handpiece should be set to fully open before sterilization in autoclave.)
Customer Service Form indicating cause code C-001 working handpiece the handpiece is faulty. and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty handpiece with the RMA number to LM for replacement.
Page 21
Fill in Customer Service Form indicating cause code C-003 and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty electronic unit for replacement. 4. If above actions do not solve the problem,...
Page 22
Cause code Repair / Remedy instructions working handpiece the handpiece is faulty. and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty handpiece for replacement. 2. If there is no Light with the working...
Page 23
Replace o-rings. Fill in a Customer Service Form indicating handpiece and bottle cap and connectors cause code C-021 and submit to LM fax +358 2 454 6444. when applicable). C-999 Level 1 SP 3. If replacing the o-rings does not solve the...
Page 24
Fill in Customer Service Form indicating cause code and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty coil for replacement. 8. If above actions do not solve the problem, C-008 Unidentified...
Page 25
C-014 and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty part with the RMA number to LM for replacement. 3. If replacing the AirLED tubing does not...
Page 26
C-015 request authorization to return the device to of air flow and submit to LM fax +358 2 454 6444. After obtaining an LM-Instruments for repair. RMA number from LM, return the entire device with accessories and the RMA number to LM for repair.
Page 27
C-014 and submit to LM fax +358 2 454 6444. After obtaining an RMA number from LM, return the faulty part with the RMA number to LM for replacement. 4. If replacing the AirLED tubing does not...
Page 28
Repair / Remedy instructions request authorization to return the device to no/low water flow. submit to LM fax +358 2 454 6444. After obtaining an RMA LM-Instruments for repair . number from LM, return the entire device with accessories and the RMA number to LM for repair.
3 Repair Instructions Warning! Some tests and measurements require that the power cable is plugged in with the casing removed. This is only allowed for legally authorized electricians and extra caution must be taken because of exposed live parts. 3.1 Removing the cover and casing Start with disconnecting the power cable before removing cover and casing.
Page 30
Unlock the tail from the connector. Make sure the tail is not twisted when reassembling the device. Unlock the handpiece holder by moving the metal plate upwards Remove the plate Twist the handpiece holder upwards and remove...
Page 31
Remove the support list by pulling it upwards Expand the casing and slide the chassis towards the front After sliding the chassis all the way to the front, lift it upwards with front part first. Reassemble in reverse order.
3.2 Replacing the scaler handpiece tubing First remove cover and casing, see instruction section 1 Disconnect the handpiece wires, note the colors. From left: 1 - Blue, 2 - Red, 3 - Black and Green together, White at separate connector. Disconnect the water tubing (clear) from the magnetic valve (pos 2).
3.3 Measuring the voltages from the control circuit board to the scaler electronics unit Remove cover and casing, see instruction section Control Circuit Board Scaler Electronics Unit Measure the output voltage from control circuit board to scaler electronics unit. Pin 5-6 and pin 5-10 (see pictures above), reading should be 10 +/- 0,5 Vdc when running at working mode 3 and with foot control at maximum power.
Measure the power supply to the scaler pin 1-2. Reading should be aprox. 30 Vdc. If all measurements were as above, the control circuit board is functioning properly and giving the correct control signals to the scaler electronics unit. But the scaler electronic unit is probably malfunctioning and must be replaced, see Section 4.
Page 35
If applicable (CombiLED), dismount the left magnetic valve coil to access the mounting screw. Loosen all screws from electrical connectors and slide out the scaler electronics unit. Reassemble in reverse order.
3.5 Replacing the compressor First remove cover and casing, see instruction section 1. Disconnect the yellow air tubing. Cut the cable ties. Disconnect the electrical wires at the connector on the control board (note the polarity). Old Design Replace the compressor and reassemble in reverse order. Note that the compressor should not be in contact with the chassis to avoid vibration noise and short circuiting the connectors of the compressor.
3.6 Cleaning a magnetic valve First remove cover and casing, see instruction section 1. Magnetic valves function: 1: Polisher water 2: Scaler water 3: Polisher cleaning air 4: Polisher air Locate the valve according to picture above. Dismount the coil and the valve plug. Inspect and clean.
3.7 Opening a stuck waterline in the scaler handpiece First remove the scaler handpiece from the tubing. Set the water control ring to fully open (dot-dot). Gently insert the LM-ProPower Fixer tool into the handpiece from the connector side to open up the stuck water line.
3.8 Cleaning of clogged system in AirLED Connect a small tube approx. 40mm length to the powder outlet Fill the powder chamber with citric acid* up to the max level. Put it back on the unit and run the cleaning cycle. The air pressure will flow the citric acid trough the clogged tubing.
5 Technical data 5.1 Technical data CombiLED LM-Instruments Oy Manufacturer PL 88 (Norrbyn rantatie 8) FIN-21601 Parainen, FINLAND Model LM-ProPower CombiLED EN60601-1: Class 1, Type BF Classification 93/42 EU: Medical products, class IIa L x W x H 270 x 110 x 165 mm (without bottle and powder container)
5.2 Technical data UltraLED Scaler LM-Instruments Oy Manufacturer PL 88 (Norrbyn rantatie 8) FIN-21601 Parainen, FINLAND Model LM-ProPower UltraLED EN60601-1: Class 1, Type BF Classification 93/42 EU: Medical products, class IIa L x W x H 270 x 140 x 165 mm (without bottle)
5.3 Technical data AirLED Polisher LM-Instruments Oy Manufacturer PL 88 (Norrbyn rantatie 8) FIN-21601 Parainen, FINLAND Model LM-ProPower AirLED EN60601-1: Class 1, Type BF Classification 93/42 EU: Medical products, class IIa L x W x H 270 x 110 x 165 mm (without bottle and powder container)
6. Warranty Terms The following warranty terms apply to the sale of LM-Instrument Oy’s products (hereinafter “Products”) to a purchasing company or individual by LM-Instruments Oy (hereinafter “Manufacturer”). Manufacturer hereby warrants that the Products will be free from defects arising from faulty materials or workmanship for a period of twenty four (24) months from the date of purchase by a customer from Manufacturer’s authorized dealer (hereinafter “Authorized Dealer”).
3. For small equipment (i.e. power scalers, air polishers, sharpening devices as well as their parts) it is important to indicate the serial number of the device as well as the original LM- Instruments invoice number. In addition, a copy of the Authorized Dealer’s invoice to the end customer (or other proof of purchase) must be included.
Customer Service Form Please fill in and return to: LM-Instruments Oy PO Box 88 (Norrbyn rantatie 8) FI-21601 Parainen, Finland Fax +358 2 4546 444 service@lm-dental.com Customer identification Dealer name Contact person Date Phone Address Email Practice name Contact person...