Page 1
Instruction manual VWR A-1001/A-1101/A-1201 Sipper/Temperature Control System European Catalogue Number EU/UK/CH Plug: 634-1138 A-1101 CONTROLLER, SIPPER EU/UK/CH Plug: 634-1139 A-1001 CONTROLLER, SIPPER WITH CONSTANT-TEMPERATURE EU/UK/CH Plug: 634-1140 A-1201 CONTROLLER, CONSTANT-TEMPERATURE Version: 1.0.1 Issued: 16, June 2020...
Page 2
Legal Address of Manufacturer Europe VWR International bv Researchpark Haasrode 2020 Geldenaaksebaan 464 B-3001 Leuven + 32 16 385011 www.vwr.com Made in China...
Operation ..................................16 Temperature Control ............................16 Sipper ................................17 System settings ................................20 Technical service ................................21 Warranty ..................................21 Compliance with local laws and regulations ........................22 Equipment disposal ..............................22 Local VWR offices in Europe and Asia Pacific........................ 23...
Safety Information Please follow the guidelines below and read this manual in its entirety to ensure safe operation of the unit. We recommend against the use of A-1000 Constant Temperature/Automatic Injection System. Do not open the device. Disconnect the device from the mains supply before carrying out ...
The A-1001 series Constant Temperature/Automatic Injection Controller is specially developed to meet the needs of some applications of our UV/Vis spectrophotometer. It is divided into 3 models according to different functions. Model A-1001: Sipper with Temperature Control, model A-1101: Sipper, model A-1201: Temperature...
1 – Touch Screen 2 – Injection Pipe 3 – Peristaltic Pump 4 – USB Port 5 – Constant Temperature Control Port 6 – Serial Port 7 – Power Inlet 8 – Fuse Seat 9 – Power Switch 10 – In Hole 11 –...
Page 9
Connect the power cord Check to confirm instrument power switch is turned off. Plug the female of the power cord into the power inlet of the instrument and plug the male into the power outlet. 1 – Power inlet 2 – Power cord 3 –...
Page 10
Connect the injection pipe As shown in the figure, insert the output end of the injection pipe (label 1) through the in hole of the cover and insert the injection end of the flow cuvette (label 2). One end of the sample discharge pipe (label 5) is connected to the sample end of the flow cuvette (label 2), and the other end is passed through the sample hole of the cover plate.
CAUTION This symbol indicates the presence of high voltage and warns the user to proceed with caution. CAUTION This symbol indicates risks associated with hot surfaces. Product Specifications Model A-1001 A-1101 A-1201 Sipper with Temperature Function Sipper Temperature Control Control 10~90℃(0.1°C interval...
Getting Started The following chart describes the basic operation of the instrument. Turn the instrument power on Switch on the power on the back of the instrument to start. Note: Do not turn the instrument on and off continuously and quickly. Wait at least 30 seconds before turning the instrument back on, otherwise the electrical and mechanical systems may be damaged.
When modifying certain setting parameters, press and slide the icon to a position to change the setting value. Using the instrument Main Interface The main interface is divided into two parts, the left part is the temperature control area, and the right part is the injection control area. A-1001...
Page 15
A-1101 A-1201 Constant temperature control state Current temperature Set temperature Appointment time process Sipper control state Sipper speed Temperature control start Stop temperature control Recycle the sample in the flow cuvette after the end of the measurement Cleaning the flow cuvette Inhale the sample into the flow cuvette...
Stop sipper/recycling/cleaning System parameter setting Operation Temperature Control Start temperature control Main interface, Press the button Start on the left to pop up the temperature control < > parameter interface, press the button to set the temperature control parameters according to the requirements. Press the button OK , the temperature control module start working.
Slow: Rise/Lower temperature slowly, no overshoot (for biometrics) Medium: Rise/Lower temperature in medium speed, small overshoot Fast: Rise/Lower temperature quickly, larger overshoot. Reservation Waiting for a period of time to start rising/lowering temperature, range: 0.5~10h, 0.5h interval adjustment or set to "-" no reservation. Stop temperature control Main interface, if temperature control is turned on, press the button Stop on the left to end the temperature control.
Page 18
Time Single injection time, range: 1 ~ 300s, interval 1s adjustment or set to "-" continuous injection until manual stop. Speed Sample inhalation speed, range: 1 ~ 8ml / min, interval 0.5 ml / min adjustment. Recycling samples Place the discharge end of the silicone tube in the waste collection container; Main interface, press the button Recycling on the right, pop-up recycling control <...
Page 19
time (or press the button Stop ) end. Press the button Cancel to return to the main interface. Time Single cleaning time, range: 1 ~ 300s, interval 1s adjustment or set to "-" continuous injection until manual stop. Stop injection/recycling/cleaning Main interface, if the injection/recycling/cleaning is in progress, press the button Stop button on the right to end.
System settings The main interface, press the icon to enter the system settings interface, Set display brightness Press the icon , on the Brightness tab to decrease/increase the display brightness. Turn automatic standby on/off Standby mode means that the instrument automatically reduces the brightness of the display when there is no operation for a period of time to save energy and extend the life of the display.
(2) years from date of delivery. If a defect is present, VWR will, at its option and cost, repair, replace, or refund the purchase price of this product to the customer, provided it is returned during the warranty period.
The customer is responsible for applying for and obtaining the necessary regulatory approvals or other authorisations necessary to run or use the Product in its local environment. VWR will not be held liable for any related omission or for not obtaining the required approval or authorisation, unless any refusal is due to a defect of the product.