Page 2
At the core of our values is an indelible need to support our customers with the best experience possible. In a world that can seem so large, Fungilab is ready to bridge the gap with a creativity and originality that will unite us unlike ever before.
Fungilab guarantees the satisfactory operation of the viscometers and its accessories if there have not been any unauthorized adjustments to the mechanical pieces, the electronic components and the software.
connection! Verify that the voltage and the frequency match with the specifications for the viscometer. Before turning on the machine, let it sit for some time so that it acclimates to the surrounding temperature in order to avoid a short-circuit caused by condensation. The fluctuations of the power source should not surpass ±10 % of the nominal voltage 3.
5. Specifications Speeds: 0.01 – 250 RPM Temperature Sensing Range: -40 °C to 300 °C (-40 °F to 572 °F) USB A Port for use with temperature probe Viscosity Accuracy: ±1.0 % of full scale range The use of accessory items will have an effect on the measurement accuracy Viscosity Repeatability: ±0.2 % of full scale range...
Page 8
Other solvents that generally provide good results are acetone, gasoline, or any solvent with a high percentage of alcohol For the use of any other solvent, make sure that it does not corrode the spindles or the protector. The spindles are made in AISI 316. Warning: Handle the volatile and flammable solvents with proper cautions.
(directive 89/655/CEE). Fungilab recommends using the carry-case provided with the equipment for making any kind of delivery. Please, keep the carry-case in a safe location. In case of transporting the equipment or during long storage periods, always use the carry-case by placing each part as shown in the picture.
Page 10
Standard spindles Model L: L1, L2, L3, L4 Models R and H: R2, R3, R4, R5, R6, R7 Fig 1. Viscometer in its carry-case VPAD Manual 10/47...
9. Equipment Description Fig. 2 Frontal view of the equipment 1. Levelling knob 2. Nut 3. Control tablet Column Front cover LED display Spindle guard Spindle Temperature probe 10. Base (viscometer stand) VPAD Manual 11/47...
Page 12
Fig. 3 Back view of the equipment 1. Power switch 5. Air ventilation 2. Power cable slot 6. USB Temperature probe 3. Warning Label connector 4. Serial number label Fig. 4. Equipment identification label Description of the equipment identification label: 1.
9.1 Equipment set-up Remove all of the parts from the carry-case. Note the figure below (fig 5). Correctly place the three height adjustable knobs (B) on the T-shaped base (A). Mount the fastening column (C) with the holding screws (D) at the base (A). ...
10. Important rheological information To obtain precise results it is necessary to know the most important rheological properties of the sample. Newtonian fluids The viscosity of these fluids does not depend on the shear rate meaning that at any speed the viscosity is the same.
Page 15
NOTE: The turbulent behaviour of a fluid can produce falsely high results in viscosity tests. Normally, turbulent behaviour is due to an excessively high rotation speed in relation to the viscosity of the sample (see detailed Warning further on). FACTORS AFFECTING VISCOSITY There are many variables that affect the rheological properties of products, so it is very important to take the following factors into account.
Page 16
Examine each spindle before using it. If it’s damaged or eroded in such a way that its dimensions are changed, it will provide false results for your viscosity reading. The spindle protector (provided with every Fungilab rotational viscometer) protects the spindle and the viscometer axle and it is important for the reading of low viscosities with standard spindles.
Page 17
The viscosity reading must be executed under laminar flow condition, not turbulent flow conditions. The first consideration is linked to the precision of the instruments. All of the FUNGILAB rotational viscometers guarantee a precision of () 1% from the bottom of any spindle/rotational speed combination scale.
Page 18
Turbulent flow: “non-linear” flow lines. Impossible to predict the exact movement of the fluid. Very quick. For rotational systems, this means that the fluid’s movement must be circumferential. When the internal forces of a fluid end up being too great, the fluid can become a turbulent flow, in that the particles that make it up become unpredictable, making it impossible to analyse it with standard mathematical models.
11. Accessories 11.1. Low viscosity adapters (LCP and LCP/B) Low viscosity adapters (LCP and LCP/B) are not included in the standard delivery. Any of these two versions (with or without thermostatic flow jacket) must be ordered as an additional accessory. Both LCP and LCP/B accessories are supplied complete with a spindle.
The piece named G has two possible holes for the upper screw. The top hole is a Universal hole to screw our low viscosity adapter to other brand viscometers. The bottom hole is to screw to Fungilab pieces. NOTE: Before starting with the measurements, make sure the viscometer is correctly balanced (check it with the bubble level).
11.1.2 Dismounting and cleaning Unscrew the spindle of the viscometer axis and lower the spindle slowly in the sample container (K). Caution on removing the spindle, it can be hot due to a previous high temperature experiment. Remove Adapter (G) from metallic glass. ...
11. 2. Small sample adapters APM and APM/B NOTE: Small sample adapters (APM and APM/B) are not included in the standard delivery. Any of these two versions (with or without thermostatic flow jacket) must be ordered as an additional accessory. APM and APM/B accessory are not supplied with a spindle.
Page 23
The piece named G has two possible holes for the upper screw. The top hole is a Universal hole to screw our small sample adapter to other brands viscometer. The bottom hole is to screw Fungilab pieces. NOTE: Before starting with the measurements, make sure the viscometer is correctly balanced (check it with the bubble level).
11. 2. 2 Dismounting and cleaning Unscrew the spindle of the viscometer axis and lower the spindle slowly in the sample container (K). Caution on removing the spindle, it can be hot due to a previous high temperature experiment. ...
Page 25
R Model or H & TR spindles Spindle Shear rate [ s ] *) Sample volume [ ml ] Container 0.93 x RPM 0.34 x RPM 10.4 TR10 0.28 x RPM 11.0 TR11 0.25 x RPM 13.5 TR12 0.48 x RPM TR12C TR13 0.40 x RPM...
Temperature range of circulation jacket and thermostatic flow conditions: Permitted temperature range: -10 a +100°C (14 a 212 °F) Use a thermostatic bath with demineralised water or refrigeration special liquid. Change the liquid form the thermostatic bath regularly. Recommended flow: 15 l/min. Materials: ...
Page 27
11. 3. 1 Heldal unit mounting Fig. 16 Heldal unit set in the viscometer 1. Lower stop ring 9. Base 2. Rib joint 10. Levelling knobs 3. Displacement command 11. Heldal motor unit 4. Viscometer fastening bolt 12. Knobbed fastening rib 5.
Page 28
Important: Do not fasten the stop rings to the fastening ribs (12) too tightly. They are plastic pieces and they can be damaged. Both stopper rings (upper and lower) look exactly the same and can be changed. Place the Heldal motor (11) in the fastener (8) while pressing the displacement command (3). ...
12. Model/Spindle correspondence tables Standard Spindles + R1 (Table 1): Viscometer model Spindle VPAD L VPAD R VPAD H SPECIAL SPINDLES FOR APM ADAPTER (Table 2): Viscometer model Spindle VPAD L VPAD R TR10 TR11 TR12 TR13 VPAD H TR10 TR11 TR12 TR13...
Page 30
SPECIAL HELDAL SPINDLES (Table 3): Viscometer model Spindle VPAD L VPAD R VPAD H SPECIAL SPINDLES FOR LCP ADAPTER (Table 4): Viscometer model Spindle VPAD L LCP/SP VPAD R LCP/SP SPECIAL VANE SPINDLES (Table 5): Viscometer model Spindle VPAD L VPAD R VPAD H VPAD Manual...
15. VPAD L Special aerial spindle selection table Maximum guideline values in cP (mPa·s) RPM / SP 100K 400K 250K 200K 125K 7.5K 2.5K 1.5K 2.5K 1.2K Table 10. ATTENTION: K Indicates thousands. Example: 7.8K = 7800 M Indicates millions Example: 1.56M = 1560000 NOTE: It is not recommended to work with viscosity values of less than 15% of the lower part of the selected scale.
17. VPAD L LCP Adaptor table Maximum guideline values in cP (mPa·s) RPM/SP 1.2K Table 12. ATTENTION: K Indicates thousands. Example: 7.8K = 7800 M Indicates millions Example: 1.56M = 1560000 NOTE: It is not recommended to work with viscosity values of less than 15% of the lower part of the selected scale.
22. VPAD R LCP Adaptor table Maximum guideline values in cP (mPa·s) 21.3K 12.8K 6.4K 44.2K 3.2K 2.5K 2.1K 1.6K 1.2K Table 17. ATTENTION: K Indicates thousands. Example: 7.8K = 7800 M Indicates millions Example: 1.56M = 1560000 NOTE: It is not recommended to work with viscosity values of less than 15% of the lower part of the selected scale.
When the serial number is incorrect or it does not suit with the written in the warranty. FUNGILAB’s sole obligation shall be to repair or to replace any part(s) that prove defective within the warranty period and shall not be liable for consequential damages resulting from the use of its products.