The user of the microINR System can help prevent electromagnetic
interference by maintaining a minimum distance between other
portable/mobile RF communications equipment (transmitters)
and the microINR System at least 30 cm (about 12 inches).
Immunity to proximity fields from RF wireless
communications equipment:
Band (MHz)
Service
380-390
TETRA 400
430-470
GMRS 460, FRS 460
704-787
LTE Band 13, 17
GSM 800/900, TETRA 800,
800-960
iDEM 820, CDMA 850,
LTE Band 5
GSM 1800; CDMA 1900;
1700-1990
GSM 1900; DECT;
LTE Band 1, 3, 4, 25; UMTS
Bluetooth, WLAN,
2400-2570
802.11 b/g/n,
RFID 2450, LTE Band 7
5100-5800
WLAN 802.11 a/n
Additional Recommendations:
The microINR should not be used adjacent to or stacked with
other equipment and that if adjacent or stacked use is
necessary, the microINR should be observed to verify normal
operation in the configuration in which it will be used.
Use only the supplied charger provided by the manufacturer or
you may damage the Meter. Other cables and accessories may
negatively affect EMC performance.
Immunity
Test Level (V/m)
27
28
9
28
28
28
9
20
Need help?
Do you have a question about the microINR and is the answer not in the manual?