4. Using a 5/64-inch (2 mm) Allen (i.e., hexagonal) wrench, secure the four
supplied M3 X 0.5 hex-head shoulder screws to the back of the Core AP as
shown in Figure 2-6. Torque-tighten each shoulder screw to 8-inch lbs.
5. Mount the Core AP to the wall plate by inserting the head of each shoulder
screw into the four mounting holes provided in the wall plate and then sliding
the Core AP down so that the shoulder screw heads are securely in the grooves
of the wall plate.
6. Using a Torx T10 screw driver, secure the Core AP to the wall plate by screwing
the supplied M3.5 X 0.6 T-10 Torx Head security screw into the PEM nut in the
wall plate as shown in Figure 2-7. Torque-tighten the screw to 10-inch lbs. Note
that this safety screw is used to prevent the Core AP from sliding out of the wall
plate.
IntelliVue Smart-hopping 1.4 GHz Access Point Installation Guide
Mounting the IntelliVue Smart-hopping Core AP to a Wall (CA Earthquake Rated)
Figure 2-7: Securing the Core AP to its Quick Release Wall Plate
Insert M3.5 Security Screw
through this View Port and
Tighten Using a Torx T10
Screwdriver.