stacked with other equipment. If adjacent or
stacked use is necessary, the device should be
observed to verify normal operation in the config-
uration in which it will be used.
Performance
Measurement accuracy
Description
When the glucose concentration is
>4.2mmol/L (75mg/dL), the sensor
accuracy deviation does not exceed
±20%;
When the glucose concentration is
≤4.2mmol/L (75mg/dL), the accuracy
deviation does not exceed ±20mg/dL