1 Safety 1.1 Intended use The D-H 22 S manual blasting unit is intended for use in dental laboratories: − for removing the residuals of embedding materials and oxides on cast components after their being unbedded, − for leveling and smoothing of cast component surfaces and −...
Refers to particular ways of operation or handling, the non-adherence of which can lead to malfunctioning, damages or other kinds of trouble. Danger Refers to dangerous situations, in which injuries can occur. D-H 22 S - 16.02.2002 / Vers.: 1...
− Foot switch with cable and plug ..........no. 022 SM 514 − 2 m PVC fabric hose, blue Ø 8.2 x 6 with insert sleeve for quick-action coupling and swivel nut................no. 022 SM 579 D-H 22 S - 16.02.2002 / Vers.: 1...
2.3 Identification and description of the unit components (See figures 1 and 3 on pages 5 and 7) The D-H 22 S manual blasting unit can be operated with all standard laboratory polish blasting media and special fused alumina. The pressure regulator (6) adapts the blasting force to the situation at hand.
Page 7
33 Front floor compartment 15 Mains connection socket 34 10 screws for front floor compartment (230 V/50 Hz) 16 Compressed air connection 50 Fuse for transformer (blasting chamber illumination) 18 Blasting chamber illumination (24 V) D-H 22 S - 16.02.2002 / Vers.: 1...
The working pressure is controlled by manometer (5). Note The working pressure should not be set to more than 5 bar. 9. The sanding jet (exits the injector) is released by actuating the foot switch. D-H 22 S - 16.02.2002 / Vers.: 1...
150 A (150 µm) Caution The unit is designed for the use of HARNISCH+RIETH blasting agents. We do not grant guarantee for malfunctioning or damages which result from the use of other kinds of blasting agents. 1. Open front flap (2).
(See figures 1 and 3, on pages 5 and 7) Caution The D-H 22 S manual blasting unit should only be operated with the ultra-fine filter (7) and filter bag (14) properly inserted and free of damages. 1. Open the door to the filter compartment (4) by screwing off the two star grips (11).
DIN 24 184). 1. Open the door to the filter compartment (4) by screwing out the two star grips (11). 2. Press the plastic connection (41) from the nozzle (40) using both hands. D-H 22 S - 16.02.2002 / Vers.: 1...
2. Let out blasting agent by taking off the sand outlet lid (13). Note While doing so, it is also advisable to take out and clean the insert sieve (8) (see section “4.3“). D-H 22 S - 16.02.2002 / Vers.: 1...
8. Check the suction basket (31) in the lower part of the sand basin for dirt particles and, if necessary clean it (for further details see “Letting out blasting agent“, section “4.4“). D-H 22 S - 16.02.2002 / Vers.: 1...
Before opening the front floor compartments (33) switch off unit at main switch Danger (12) and pull the mains supply plug. − Removing the 10 screws (34) opens the front floor compartment. D-H 22 S - 16.02.2002 / Vers.: 1...
: Facility for separating health endangering dusts with re-circulation of the clean air ZH 1/487 into the work rooms. A technical documentation is available. The operation instructions belonging to the machine are also available. Director of the Quality Control Department Winterbach, 06th September, 1995 D-H 22 S - 16.02.2002 / Vers.: 1...