Cleaning
Following the cleaning instructions after each use will prevent any remaining medication in the nebulizer from drying and
causing the device to nebulize ineffectively.
To prevent possible risk of infection due to contamination, follow these instructions:
Clean the nebulizer before initial operations.
Clean the nebulizer after each aerosol treatment.
ALWAYS use fresh cleaning solution.
DO NOT towel dry the nebulizer components after cleaning.
NOTE: Your physician and/or an Invacare dealer may specify certain cleaning procedures. If so, follow their recommendations. If not, clean
according to the following instructions:
1.
Fill two plastic containers or bowls, one for washing and the other for soaking:
A. Washing: Hot water/detergent solution.
B. Soaking: Hot water/vinegar solution (one part vinegar to three parts water).
2.
Remove the nebulizer and mouthpiece from the tubing.
3.
Disassemble the nebulizer chamber by turning counterclockwise and separating.
4.
Thoroughly wash the nebulizer and mouthpiece in hot water/detergent solution.
5.
Rinse the nebulizer and mouthpiece with clear hot tap water.
6.
Soak the nebulizer and mouthpiece in the hot water/vinegar solution for 30 minutes.
7.
Rinse with hot tap water and allow pieces to air dry completely.
NOTE: If using medical disinfectant cleaners, follow the manufacturer's instructions carefully.
NOTE: Keep the outer surface of the air tubing dust-free by wiping regularly.
NOTE: The nebulizer air tubing does not have to be washed internally because only filtered air passes through the tubing.
Cleaning the Mouthpiece
1.
Boil the mouthpiece in water for a maximum of five minutes or autoclave at a maximum of 121°C for ten minutes.
2.
Allow the mouthpiece to air dry completely.
Nebulizer Replacement
The nebulizer is designed for single patient use.
The nebulizer MUST be replaced after a long period of inactivity or when it is obstructed by dry medication, dust, etc.
The reusable nebulizer should last from six months to one year with general use. The disposable nebulizer should last
30 days. For the nebulizer that is disposable, it is recommended that an extra nebulizer be kept at all times.
Nebulizer
WARNING
2