Mindray V60 Operator's Manual
Hide thumbs Also See for V60:
Table of Contents

Advertisement

Quick Links

V60 Anesthetic Vaporizer
Operator's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the V60 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Mindray V60

  • Page 1 V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 3 © 2015 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved. Operator's Manual issue date November, 2015. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 4 SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyright or patents and does not convey any license under the patent rights or copyright of Mindray, or of others.
  • Page 5 Contents of this manual are subject to change without prior notice. All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this manual.
  • Page 6 Mindray or repairs by people other than Mindray authorized personnel.
  • Page 7 Service Mindray maintains a network of service representatives and factory-trained distributors. Prior to requesting service, perform a complete operational check of the instrument to verify proper control settings. If operational problems continue to exist, In North America contact the Service Department at (800) 288-2121, ext: 8116 for Technical Support or (201) 995-8000 for assistance in determining the nearest field service location.
  • Page 8 FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 9: Table Of Contents

    6.5 Moving when Filled ......................6-6 7 Cleaning and Disinfecting ....................7-1 7.1 Cleaning .......................... 7-1 7.2 Disinfecting ........................7-2 8 User Maintenance ......................8-1 8.1 Repair Policy ........................8-1 8.2 Maintenance Schedule ....................8-1 V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 10 A.3 Operating Range ......................A-2 A.4 Performance Specifications ................... A-2 A.5 Product Configurations ....................A-3 A.6 Flow Range ........................A-3 B Accessories List ........................ B-1 C Symbols and Terminology ..................... C-1 C.1 Symbols .......................... C-1 C.2 Terminology ........................C-1 V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 11: Safety

    Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage. NOTE Provides application tips or other useful information to ensure that you get the most from your product. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 12: Warnings

    (and further indicated by labels of different colors). Do not use the vaporizer if the vaporizer is filled with any agent other than the agent specified on the front label. This anesthetic vaporizer is not suitable for use in an MRI environment. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 13: Cautions

    1.1.3 Notes NOTE Keep this manual close to the anesthetic vaporizer so that it can be obtained conveniently when needed. This manual describes all features and options. Your anesthetic vaporizer may not have all of them. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 14: Anesthetic Vaporizer Symbols

    Laboratory (NRTL) per OSHA. U.S. Federal Law restricts this device to sale by or on the order of a Rx only physician or other practitioner licensed by state law to use or order the use of this device. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 15: The Basics

    The Key Filler system is in compliance with ISO5360. Quik-Fil system complies with the performance data of ISO5360. Mindray recommends that the output concentration is monitored through an anesthetic gas monitoring device in compliance with ISO80601-2-55 to detect any hazardous output values.
  • Page 16: Intended Use

    V60 Anesthetic Vaporizer is an unheated, calibrated anesthetic vaporizer used for evaporating liquid anesthetic agents and delivering mixed gas of controlled concentration to an anesthetic delivery system. The V60 Anesthetic Vaporizer is available in models specific for use with Isoflurane and models available specific for use with Sevoflurane. WARNING The anesthetic vaporizer is intended to be operated only by licensed clinicians and qualified anesthesia personnel who have received adequate training in its use.
  • Page 17: Rear View

    2.3.2 Rear View Handle for locking lever Interlock system Locking pin V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 18: Configuration Differences

    2.4 Configuration Differences Model Sevoflurane Key Filler Isoflurane Key Filler Sevoflurane Quik-Fil Vaporizer Vaporizer Vaporizer Anesthetic agent Sevoflurane Isoflurane Sevoflurane Filling system Key Filler system Key Filler system Quik-Fil system Picture Filler Accessory Drainage Accessory V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 19: Method Of Operation

    When the anesthetic delivery system is connected to multiple vaporizers, the interlock systems of the vaporizers ensure that only one vaporizer can be switched on at any one time while the others remain off and in a locked state. Interlock device V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 20: Plug-In Adapter/Plug-In Connector

    The openings in the plug-in adapter on the vaporizer fit onto the pins on the plug-in connector on the anesthetic delivery system. Plug-in adapter and pin Plug-in connector To connect/disconnect the vaporizer, the control dial must be at the “0” setting indicating locked status. Control dial at the “0” setting V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 21: Interlock Device

    WARNING Before operation, check if the interlock device is fully functional. A malfunction in the interlock system can endanger the patient by overdosing or mixing anesthetic agents. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 22: Filling System

    The filling system is used to fill and drain the specific anesthetic agent. The filling system has a liquid level indicator which displays filling level with the maximum and minimum levels marked. Key Filler system Filler control Keyed block Maximum mark Minimum mark Locking knob V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 23 Quik-Fil system Quik-Fil specific filling system Drainage screw Maximum mark Filler cap Minimum mark Drainage outlet V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 24 FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 25: Filling And Draining

    Ensure adequate ambient ventilation when filling the vaporizer. Before use, check that the correct anesthetic agent is used. For instance, check the name of anesthetic agent and color mark on the vaporizer and the anesthetic agent bottle. Isoflurane Purple Sevoflurane Yellow V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 26: Key Filler System

    4.2.1 Key Filler System The filling steps of V60 Isoflurane Key Filler Vaporizer are described below. If the vaporizer is connected to the anesthetic delivery system, fresh gas flow can remain as set. Turn the control dial clockwise back to the “0” position until the “0” button pops up.
  • Page 27 CAUTION If the connection between the filling adapter and anesthetic agent bottle is not leak-tight, anesthetic agent may escape. NOTE If a new unopened anesthetic agent bottle is partly empty, there may be a leak. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 28 Turn the locking knob counter clockwise. Remove the keyed block. Push the keyed end of the filling adapter into the opening of the filling system until it is fully and properly seated. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 29 Tighten the locking knob by turning it clockwise. CAUTION If the connection between the filling adapter and filling system is not leak-tight, anesthetic agent may escape. Raise the anesthetic agent bottle upside down. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 30 14. Put the keyed block back into the opening of the filling system. 15. Tighten the locking knob. 16. Unscrew the filling adapter from the anesthetic agent bottle. 17. Tighten the cap of the anesthetic agent bottle even if it is completely empty. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 31: Quik-Fil System

    Screw the Quik-Fil adapter firmly into the anesthetic agent bottle. CAUTION If the connection between the filling adapter and anesthetic agent bottle is not leak-tight, anesthetic agent may escape. Agent-specific filling cannot be assured when bottles without collars used. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 32 Tighten the filler cap. Failure to do so may cause fresh gas and anesthetic agent to escape when the vaporizer is turned on. Unscrew the filling adapter from the anesthetic agent bottle. 10. Tighten the cap of the anesthetic agent bottle even if it is completely empty. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 33: Draining The Vaporizer

    Do not reuse the anesthetic agent drained from the vaporizer. 4.3.1 Key Filler System The draining steps of V60 Isoflurane Key Filler Vaporizer are described below. Place the vaporizer upright or properly mount it so that all the anesthetic agent can drain out.
  • Page 34 Turn the locking knob counter clockwise. Remove the keyed block. Push the keyed end of the filling adapter into the opening of the filling system until it is fully and properly seated.. V60 Anesthetic Vaporizer Operator’s Manual 4-10...
  • Page 35 13. Put the keyed block back into the opening of the filling system. 14. Tighten the locking knob. 15. Unscrew the filling adapter. 16. Tighten the cap of the anesthetic agent bottle even if it is completely empty. V60 Anesthetic Vaporizer Operator’s Manual 4-11...
  • Page 36: Quik-Fil System

    Rotate Drainage funnel Depress Anesthetic agent bottle CAUTION If the connection between the filling adapter and anesthetic agent bottle is not leak-tight, anesthetic agent may escape. V60 Anesthetic Vaporizer Operator’s Manual 4-12...
  • Page 37 To prevent overflow, if necessary, close the drainage knob, replace the full bottle with an empty bottle and continue the drainage process. Close the drainage screw clockwise. Tighten the filler cap. V60 Anesthetic Vaporizer Operator’s Manual 4-13...
  • Page 38: Blowing Off The Vaporizer

    5% and flush for 5 hours at 5 L/min Air or for 2 hours at 10 L/min Air. Ensure that inspiratory port, expiratory port, and manual bag port are occluded and that the gas scavenging system is functional and active. V60 Anesthetic Vaporizer Operator’s Manual 4-14...
  • Page 39: Checks Before Use

    We recommend monitoring oxygen concentration continuously and setting at least a low alarm limit to detect insufficient oxygen supply. Plugs on the plug-in adapter of the vaporizer need to be removed and saved when opening the package. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 40: Setting Checks

    Anesthetic delivery system). Flush the breathing system with fresh gas before connecting a patient. WARNING Do not operate the vaporizer until all checks have been successfully performed. All repairs must be carried out by qualified service personnel. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 41: Basic Operations

    Only use anesthetic vaporizer with anesthetic delivery systems that are suitable according to ISO 8835-4. If the Mindray V60 vaporizer is connected to anesthetic delivery systems from other manufacturers, it is the responsibility of the operator to ensure that all technical specifications of the vaporizer and the anesthetic delivery systems are met.
  • Page 42 Switch the vaporizers off when one or more than one vaporizers have been on the manifold of the anesthetic delivery system, before mounting additional vaporizers. Set the control dial to the “0” position. Hold the vaporizer in vertical position with both hands and lower gently onto the anesthesia machine plug-in connector. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 43 Locking lever in locked position Connect additional vaporizers on the anesthetic delivery system as needed: For a 2 vaporizer anesthesia delivery system, ensure the interlock pins of both vaporizers are in direct contact with each other. Direct contact V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 44: Adjusting The Concentration Of Anesthetic Agent

    If the concentration cannot be set, do not force the control dial. Check that all other vaporizers connected are in “0” position and that the interlock device is operational. Stop use of the vaporizer if the control dial loosens or is damaged. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 45: Switching Off The Vaporizer

    (2) Keep within the permissible temperature and humidity range. (3) Observe use-by date of the anesthetic agent. If the vaporizer has to be removed from the anesthetic delivery system, see 6.4 Disconnecting the Vaporizer and 6.5 Moving when Filled. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 46: Disconnecting The Vaporizer

    When the vaporizer is detached from the anesthetic delivery system and transported separately, the control dial must remain at the “0” position. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 47: Cleaning And Disinfecting

    Do not allow the cleaning solutions to get into the gas inlet, gas outlet, or filling system. CAUTION Liquids other than the anesthetic agents specified for the vaporizer get into the vaporizer, may cause injury to the patient. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 48: Disinfecting

    Sodium hypochlorite solution (10% available chlorine) Super Sani-Cloth (0.5% Quaternary ammonium chloride and 55% Isopropyl alcohol) WARNING Do not sterilize the vaporizer and accessories. This may damage the vaporizer and may cause incorrect output concentration. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 49: User Maintenance

    NOTE All repairs must be carried out by qualified service personnel. Replace damaged parts with components manufactured or sold by Mindray. After repair, test the unit to make sure that it complies with the manufacturer’s published specifications. Contact Mindray for service assistance.
  • Page 50: Checking The Concentration

    ±0.25 vol.% or ±20% rel., +0.35/-0.25 vol.% or +30/-20% rel., whichever is greater whichever is greater Test result If the corrected measured value is within the permissible range of output concentration, the vaporizer can be put into operation. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 51: Checking The Filling System

    The locking lever is undamaged and not buckled. The interlock device is undamaged, guides easily and cannot be removed. Two interlock pins are present. Seals are undamaged. Manufacturer’s plate on the back of the vaporizer is present and secure. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 52 FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 53: Troubleshooting

    Repair by trained service personnel if the vaporizer is defective. The vaporizer is incorrectly If necessary, re-install the vaporizer or installed or the plug-in adapter is have it repaired by trained service damaged. personnel. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 54 The vaporizer is overfilled. Drain the vaporizer to the maximum incorrect filling level is mark and check the concentration. shown in the sight glass. Sight glass display is faulty. Repair by trained service personnel. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 55: Filling And Draining Related Faults And Remedies

    Contact us immediately. There is foreign body between Remove foreign body. the plug-in connector and plug-in adapter. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 56 FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 57: Storage And Transport

    Do not transport the vaporizer with anesthetic agent filled, or it may cause incorrect output concentration. NOTE Liquid anesthetic agents and filled vaporizers are subject to Hazardous Goods Regulations. These regulations do not apply to the residues of anesthetic agents left in the wick after draining. V60 Anesthetic Vaporizer Operator’s Manual 10-1...
  • Page 58 FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual 10-2...
  • Page 59: Theory Of Operation

    Fresh gas enters by the inlet D. Some of the fresh gas is routed through the vaporizing chamber G, and charged with anesthetic agent in soaked wick F. The rest of the fresh gas is routed past the airway I and through the temperature compensator E. V60 Anesthetic Vaporizer Operator’s Manual 11-1...
  • Page 60 B, finally flows out from the outlet C. The vaporizing chamber G is completely shut off from the gas flow by valves J and K. No anesthetic-agent can escape from the vaporizing chamber G. V60 Anesthetic Vaporizer Operator’s Manual 11-2...
  • Page 61: Calibration

    If the temperature of the vaporizer before use is outside 10 to 40 , a time of 15 min/ to be allowed for temperature adjustment so that the concentration remains within the accuracy specified. V60 Anesthetic Vaporizer Operator’s Manual 11-3...
  • Page 62 Influence of Running Time). The diagrams show typical temperature dependence when operating with a 2 L/min flow of Air. If temperature is not within this range, the deviations are shown as following figures, despite continuing compensation: V60 Anesthetic Vaporizer Operator’s Manual 11-4...
  • Page 63: Influence Of Flow

    The diagrams show the influence of flow on the concentration delivered after 1 minute at 22 , 101.3 kPa when operating with Air. Isoflurane Set concentration Output concentration vol.% vol.% 0.6% Fresh-gas flow L/min V60 Anesthetic Vaporizer Operator’s Manual 11-5...
  • Page 64: Influence Of Gas Composition

    The higher the concentration set; The lower the gas flow; The more extreme the change of gas type. The extent of this dynamic deviation increases as gas flow increases, though the duration of the deviation will decrease. V60 Anesthetic Vaporizer Operator’s Manual 11-6...
  • Page 65: Influence Of Atmospheric Pressure

    Under no circumstances should the vaporizer ever be used at an atmospheric pressure and / or temperature at which the anesthetic agent could start to boil, as the concentration delivered will rise and be uncontrolled. V60 Anesthetic Vaporizer Operator’s Manual 11-7...
  • Page 66: Influence Of Fluctuations In Pressure

    E. Fresh gas flow of 1 L/min, concentration set of 6%, running time of 6 hours. F. Fresh gas flow of 4 L/min, concentration set of 5%, running time of 4 hours. G. Fresh gas flow of 1 L/min, concentration set of 5%, running time of 6 hours. V60 Anesthetic Vaporizer Operator’s Manual 11-8...
  • Page 67: Anesthetic Agent Consumption

    (L/min) output concentration (vol.%) Note: The rate of consumption of anesthetic agent depends primarily on flowrate and vapor output concentration. The figures are approximate and are intended for general guidance only. V60 Anesthetic Vaporizer Operator’s Manual 11-9...
  • Page 68 FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual 11-10...
  • Page 69: A Product Specifications

    Weight 6 ± 0.5kg (empty) Dimensions Height:239 ±3 mm Width: 120 ±3 mm Depth: 210 ±10 mm Filling volume 360 ml (dry wick) 300 ml (moist wick) 260 ml (between the minimum and maximum marks) V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 70: Operating Range

    Alone, freestanding During operation (fixed position) Pressure difference Difference between pressure range and ambient pressure on the -10 to 20 kPa vaporizer outlet Flow resistance(without connectors) at 10L/min Air at 22 switch on/off: <70cm H V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 71: Product Configurations

    A.5 Product Configurations Filling system Key Filler Quik-Fil Isoflurane vaporizer Sevoflurane vaporizer A.6 Flow Range Flow Range 0.2 to 15L/min 0.2 to 10L/min for concentrations >5 Vol.% V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 72 FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 73 Description Filling adapter 040-002707-00 Key Filler filling adapter for isoflurane vaporizer 040-002708-00 Key Filler filling adapter for sevoflurane vaporizer Quik-Fil filling adapter for sevoflurane (0605) 115-026747-00 Draining adapter Quik-Fil drainage funnel for sevoflurane vaporizer 040-000067-00 V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 74 FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 75: C Symbols And Terminology

    Percentage %rel Relative deviation from value in % Degree Celsius, unit of temperature Degree, unit of plane angle Kilogram, unit of mass Kilopascal, unit of pressure Hundred Pascal, unit of pressure Pascal, unit of pressure V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 76 While filling, adopt filling adapter to connect the anesthetic agent bottle and filler together. While draining, turn on draining valve by specific key, and adopt draining Quik-Fil adapter and drainage funnel to connect the anesthetic agent bottle and the anesthetic vaporizer to drain anesthetic agent. V60 Anesthetic Vaporizer Operator’s Manual...
  • Page 78 P/N 046-006516-00 3.0...