Page 4
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyright or patents and does not convey any license under the patent rights or copyright of Mindray, or of others.
Page 5
Contents of this manual are subject to change without prior notice. All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this manual.
Page 6
Mindray or repairs by people other than Mindray authorized personnel.
Page 7
Service Mindray maintains a network of service representatives and factory-trained distributors. Prior to requesting service, perform a complete operational check of the instrument to verify proper control settings. If operational problems continue to exist, In North America contact the Service Department at (800) 288-2121, ext: 8116 for Technical Support or (201) 995-8000 for assistance in determining the nearest field service location.
Page 8
FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage. NOTE Provides application tips or other useful information to ensure that you get the most from your product. V60 Anesthetic Vaporizer Operator’s Manual...
(and further indicated by labels of different colors). Do not use the vaporizer if the vaporizer is filled with any agent other than the agent specified on the front label. This anesthetic vaporizer is not suitable for use in an MRI environment. V60 Anesthetic Vaporizer Operator’s Manual...
1.1.3 Notes NOTE Keep this manual close to the anesthetic vaporizer so that it can be obtained conveniently when needed. This manual describes all features and options. Your anesthetic vaporizer may not have all of them. V60 Anesthetic Vaporizer Operator’s Manual...
Laboratory (NRTL) per OSHA. U.S. Federal Law restricts this device to sale by or on the order of a Rx only physician or other practitioner licensed by state law to use or order the use of this device. V60 Anesthetic Vaporizer Operator’s Manual...
The Key Filler system is in compliance with ISO5360. Quik-Fil system complies with the performance data of ISO5360. Mindray recommends that the output concentration is monitored through an anesthetic gas monitoring device in compliance with ISO80601-2-55 to detect any hazardous output values.
V60 Anesthetic Vaporizer is an unheated, calibrated anesthetic vaporizer used for evaporating liquid anesthetic agents and delivering mixed gas of controlled concentration to an anesthetic delivery system. The V60 Anesthetic Vaporizer is available in models specific for use with Isoflurane and models available specific for use with Sevoflurane. WARNING The anesthetic vaporizer is intended to be operated only by licensed clinicians and qualified anesthesia personnel who have received adequate training in its use.
When the anesthetic delivery system is connected to multiple vaporizers, the interlock systems of the vaporizers ensure that only one vaporizer can be switched on at any one time while the others remain off and in a locked state. Interlock device V60 Anesthetic Vaporizer Operator’s Manual...
The openings in the plug-in adapter on the vaporizer fit onto the pins on the plug-in connector on the anesthetic delivery system. Plug-in adapter and pin Plug-in connector To connect/disconnect the vaporizer, the control dial must be at the “0” setting indicating locked status. Control dial at the “0” setting V60 Anesthetic Vaporizer Operator’s Manual...
WARNING Before operation, check if the interlock device is fully functional. A malfunction in the interlock system can endanger the patient by overdosing or mixing anesthetic agents. V60 Anesthetic Vaporizer Operator’s Manual...
The filling system is used to fill and drain the specific anesthetic agent. The filling system has a liquid level indicator which displays filling level with the maximum and minimum levels marked. Key Filler system Filler control Keyed block Maximum mark Minimum mark Locking knob V60 Anesthetic Vaporizer Operator’s Manual...
Page 23
Quik-Fil system Quik-Fil specific filling system Drainage screw Maximum mark Filler cap Minimum mark Drainage outlet V60 Anesthetic Vaporizer Operator’s Manual...
Page 24
FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
Ensure adequate ambient ventilation when filling the vaporizer. Before use, check that the correct anesthetic agent is used. For instance, check the name of anesthetic agent and color mark on the vaporizer and the anesthetic agent bottle. Isoflurane Purple Sevoflurane Yellow V60 Anesthetic Vaporizer Operator’s Manual...
4.2.1 Key Filler System The filling steps of V60 Isoflurane Key Filler Vaporizer are described below. If the vaporizer is connected to the anesthetic delivery system, fresh gas flow can remain as set. Turn the control dial clockwise back to the “0” position until the “0” button pops up.
Page 27
CAUTION If the connection between the filling adapter and anesthetic agent bottle is not leak-tight, anesthetic agent may escape. NOTE If a new unopened anesthetic agent bottle is partly empty, there may be a leak. V60 Anesthetic Vaporizer Operator’s Manual...
Page 28
Turn the locking knob counter clockwise. Remove the keyed block. Push the keyed end of the filling adapter into the opening of the filling system until it is fully and properly seated. V60 Anesthetic Vaporizer Operator’s Manual...
Page 29
Tighten the locking knob by turning it clockwise. CAUTION If the connection between the filling adapter and filling system is not leak-tight, anesthetic agent may escape. Raise the anesthetic agent bottle upside down. V60 Anesthetic Vaporizer Operator’s Manual...
Page 30
14. Put the keyed block back into the opening of the filling system. 15. Tighten the locking knob. 16. Unscrew the filling adapter from the anesthetic agent bottle. 17. Tighten the cap of the anesthetic agent bottle even if it is completely empty. V60 Anesthetic Vaporizer Operator’s Manual...
Screw the Quik-Fil adapter firmly into the anesthetic agent bottle. CAUTION If the connection between the filling adapter and anesthetic agent bottle is not leak-tight, anesthetic agent may escape. Agent-specific filling cannot be assured when bottles without collars used. V60 Anesthetic Vaporizer Operator’s Manual...
Page 32
Tighten the filler cap. Failure to do so may cause fresh gas and anesthetic agent to escape when the vaporizer is turned on. Unscrew the filling adapter from the anesthetic agent bottle. 10. Tighten the cap of the anesthetic agent bottle even if it is completely empty. V60 Anesthetic Vaporizer Operator’s Manual...
Do not reuse the anesthetic agent drained from the vaporizer. 4.3.1 Key Filler System The draining steps of V60 Isoflurane Key Filler Vaporizer are described below. Place the vaporizer upright or properly mount it so that all the anesthetic agent can drain out.
Page 34
Turn the locking knob counter clockwise. Remove the keyed block. Push the keyed end of the filling adapter into the opening of the filling system until it is fully and properly seated.. V60 Anesthetic Vaporizer Operator’s Manual 4-10...
Page 35
13. Put the keyed block back into the opening of the filling system. 14. Tighten the locking knob. 15. Unscrew the filling adapter. 16. Tighten the cap of the anesthetic agent bottle even if it is completely empty. V60 Anesthetic Vaporizer Operator’s Manual 4-11...
Rotate Drainage funnel Depress Anesthetic agent bottle CAUTION If the connection between the filling adapter and anesthetic agent bottle is not leak-tight, anesthetic agent may escape. V60 Anesthetic Vaporizer Operator’s Manual 4-12...
Page 37
To prevent overflow, if necessary, close the drainage knob, replace the full bottle with an empty bottle and continue the drainage process. Close the drainage screw clockwise. Tighten the filler cap. V60 Anesthetic Vaporizer Operator’s Manual 4-13...
5% and flush for 5 hours at 5 L/min Air or for 2 hours at 10 L/min Air. Ensure that inspiratory port, expiratory port, and manual bag port are occluded and that the gas scavenging system is functional and active. V60 Anesthetic Vaporizer Operator’s Manual 4-14...
We recommend monitoring oxygen concentration continuously and setting at least a low alarm limit to detect insufficient oxygen supply. Plugs on the plug-in adapter of the vaporizer need to be removed and saved when opening the package. V60 Anesthetic Vaporizer Operator’s Manual...
Anesthetic delivery system). Flush the breathing system with fresh gas before connecting a patient. WARNING Do not operate the vaporizer until all checks have been successfully performed. All repairs must be carried out by qualified service personnel. V60 Anesthetic Vaporizer Operator’s Manual...
Only use anesthetic vaporizer with anesthetic delivery systems that are suitable according to ISO 8835-4. If the Mindray V60 vaporizer is connected to anesthetic delivery systems from other manufacturers, it is the responsibility of the operator to ensure that all technical specifications of the vaporizer and the anesthetic delivery systems are met.
Page 42
Switch the vaporizers off when one or more than one vaporizers have been on the manifold of the anesthetic delivery system, before mounting additional vaporizers. Set the control dial to the “0” position. Hold the vaporizer in vertical position with both hands and lower gently onto the anesthesia machine plug-in connector. V60 Anesthetic Vaporizer Operator’s Manual...
Page 43
Locking lever in locked position Connect additional vaporizers on the anesthetic delivery system as needed: For a 2 vaporizer anesthesia delivery system, ensure the interlock pins of both vaporizers are in direct contact with each other. Direct contact V60 Anesthetic Vaporizer Operator’s Manual...
If the concentration cannot be set, do not force the control dial. Check that all other vaporizers connected are in “0” position and that the interlock device is operational. Stop use of the vaporizer if the control dial loosens or is damaged. V60 Anesthetic Vaporizer Operator’s Manual...
(2) Keep within the permissible temperature and humidity range. (3) Observe use-by date of the anesthetic agent. If the vaporizer has to be removed from the anesthetic delivery system, see 6.4 Disconnecting the Vaporizer and 6.5 Moving when Filled. V60 Anesthetic Vaporizer Operator’s Manual...
When the vaporizer is detached from the anesthetic delivery system and transported separately, the control dial must remain at the “0” position. V60 Anesthetic Vaporizer Operator’s Manual...
Do not allow the cleaning solutions to get into the gas inlet, gas outlet, or filling system. CAUTION Liquids other than the anesthetic agents specified for the vaporizer get into the vaporizer, may cause injury to the patient. V60 Anesthetic Vaporizer Operator’s Manual...
Sodium hypochlorite solution (10% available chlorine) Super Sani-Cloth (0.5% Quaternary ammonium chloride and 55% Isopropyl alcohol) WARNING Do not sterilize the vaporizer and accessories. This may damage the vaporizer and may cause incorrect output concentration. V60 Anesthetic Vaporizer Operator’s Manual...
NOTE All repairs must be carried out by qualified service personnel. Replace damaged parts with components manufactured or sold by Mindray. After repair, test the unit to make sure that it complies with the manufacturer’s published specifications. Contact Mindray for service assistance.
±0.25 vol.% or ±20% rel., +0.35/-0.25 vol.% or +30/-20% rel., whichever is greater whichever is greater Test result If the corrected measured value is within the permissible range of output concentration, the vaporizer can be put into operation. V60 Anesthetic Vaporizer Operator’s Manual...
The locking lever is undamaged and not buckled. The interlock device is undamaged, guides easily and cannot be removed. Two interlock pins are present. Seals are undamaged. Manufacturer’s plate on the back of the vaporizer is present and secure. V60 Anesthetic Vaporizer Operator’s Manual...
Page 52
FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
Repair by trained service personnel if the vaporizer is defective. The vaporizer is incorrectly If necessary, re-install the vaporizer or installed or the plug-in adapter is have it repaired by trained service damaged. personnel. V60 Anesthetic Vaporizer Operator’s Manual...
Page 54
The vaporizer is overfilled. Drain the vaporizer to the maximum incorrect filling level is mark and check the concentration. shown in the sight glass. Sight glass display is faulty. Repair by trained service personnel. V60 Anesthetic Vaporizer Operator’s Manual...
Contact us immediately. There is foreign body between Remove foreign body. the plug-in connector and plug-in adapter. V60 Anesthetic Vaporizer Operator’s Manual...
Page 56
FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual...
Do not transport the vaporizer with anesthetic agent filled, or it may cause incorrect output concentration. NOTE Liquid anesthetic agents and filled vaporizers are subject to Hazardous Goods Regulations. These regulations do not apply to the residues of anesthetic agents left in the wick after draining. V60 Anesthetic Vaporizer Operator’s Manual 10-1...
Page 58
FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual 10-2...
Fresh gas enters by the inlet D. Some of the fresh gas is routed through the vaporizing chamber G, and charged with anesthetic agent in soaked wick F. The rest of the fresh gas is routed past the airway I and through the temperature compensator E. V60 Anesthetic Vaporizer Operator’s Manual 11-1...
Page 60
B, finally flows out from the outlet C. The vaporizing chamber G is completely shut off from the gas flow by valves J and K. No anesthetic-agent can escape from the vaporizing chamber G. V60 Anesthetic Vaporizer Operator’s Manual 11-2...
If the temperature of the vaporizer before use is outside 10 to 40 , a time of 15 min/ to be allowed for temperature adjustment so that the concentration remains within the accuracy specified. V60 Anesthetic Vaporizer Operator’s Manual 11-3...
Page 62
Influence of Running Time). The diagrams show typical temperature dependence when operating with a 2 L/min flow of Air. If temperature is not within this range, the deviations are shown as following figures, despite continuing compensation: V60 Anesthetic Vaporizer Operator’s Manual 11-4...
The diagrams show the influence of flow on the concentration delivered after 1 minute at 22 , 101.3 kPa when operating with Air. Isoflurane Set concentration Output concentration vol.% vol.% 0.6% Fresh-gas flow L/min V60 Anesthetic Vaporizer Operator’s Manual 11-5...
The higher the concentration set; The lower the gas flow; The more extreme the change of gas type. The extent of this dynamic deviation increases as gas flow increases, though the duration of the deviation will decrease. V60 Anesthetic Vaporizer Operator’s Manual 11-6...
Under no circumstances should the vaporizer ever be used at an atmospheric pressure and / or temperature at which the anesthetic agent could start to boil, as the concentration delivered will rise and be uncontrolled. V60 Anesthetic Vaporizer Operator’s Manual 11-7...
E. Fresh gas flow of 1 L/min, concentration set of 6%, running time of 6 hours. F. Fresh gas flow of 4 L/min, concentration set of 5%, running time of 4 hours. G. Fresh gas flow of 1 L/min, concentration set of 5%, running time of 6 hours. V60 Anesthetic Vaporizer Operator’s Manual 11-8...
(L/min) output concentration (vol.%) Note: The rate of consumption of anesthetic agent depends primarily on flowrate and vapor output concentration. The figures are approximate and are intended for general guidance only. V60 Anesthetic Vaporizer Operator’s Manual 11-9...
Page 68
FOR YOUR NOTES V60 Anesthetic Vaporizer Operator’s Manual 11-10...
Weight 6 ± 0.5kg (empty) Dimensions Height:239 ±3 mm Width: 120 ±3 mm Depth: 210 ±10 mm Filling volume 360 ml (dry wick) 300 ml (moist wick) 260 ml (between the minimum and maximum marks) V60 Anesthetic Vaporizer Operator’s Manual...
Alone, freestanding During operation (fixed position) Pressure difference Difference between pressure range and ambient pressure on the -10 to 20 kPa vaporizer outlet Flow resistance(without connectors) at 10L/min Air at 22 switch on/off: <70cm H V60 Anesthetic Vaporizer Operator’s Manual...
Percentage %rel Relative deviation from value in % Degree Celsius, unit of temperature Degree, unit of plane angle Kilogram, unit of mass Kilopascal, unit of pressure Hundred Pascal, unit of pressure Pascal, unit of pressure V60 Anesthetic Vaporizer Operator’s Manual...
Page 76
While filling, adopt filling adapter to connect the anesthetic agent bottle and filler together. While draining, turn on draining valve by specific key, and adopt draining Quik-Fil adapter and drainage funnel to connect the anesthetic agent bottle and the anesthetic vaporizer to drain anesthetic agent. V60 Anesthetic Vaporizer Operator’s Manual...