UMC555 \ UMC444 TABLE OF CONTENTS SECTION DESCRIPTION INTRODUCTION SPECIFICATIONS INSTALLATION AND WIRING DIGITAL CONFIGURATION ANALOG CALIBRATION SERIAL OUTPUT OPTIONS RECOMMENDED SPARE PARTS REVISED 1/26/96...
SECTION 1 GENERAL INTRODUCTION ______________________________________________________________________________ This Digital Weight Indicator is a microprocessor based instrument specifically designed to meet the needs of the scale industry. FEATURES INCLUDE: * Analog and Digital Calibration via Front Panel Entry. * NON - VOLATILE memory for Configuration, Auto Zero, Auto Tare, PAZ values. * 10,000 Displayed Graduations;...
Page 6
POWER-UP INDICATION: THE UNIT PROVIDES AUTOMATIC DISPLAY CHECKING DURING "POWER-UP" BY ROTATING AN "8" FOR VISUAL VERIFICATION OF DISPLAY OPERATION. IN ADDITION, THE SOFTWARE "SET" ,"VERSION", AND OPTION STATUS IS SUPPLIED. FORMAT: "Pr. -x.y" EXAMPLE: PROGRAM NO. KBF1921-1 Ver 11 is "Pr.-1.11" in power up message. OPTION FORMAT: "OPt.---"...
Page 7
MODEL NUMBER GUIDE (___) (___) (___) (___) ENCLOSURE ___________________| A. Mild steel painted enclosure with tilt stand base. B. Mild steel painted enclosure with panel mtg. brackets. C. Uncased board set INPUT POWER _________________________| A. 117 VAC, 60 Hz B. 217 VAC, 50 Hz C.
Page 8
(___) (___) (___) (___) ENCLOSURE __________________| A. Mild steel painted enclosure | with tilt stand base. B. Mild steel painted enclosure | with panel mtg. brackets. C. Uncased board set D. NEMA 4X SS Enclosure INPUT POWER _______________________| A. 117 VAC, 60 Hz B.
Page 12
D I S P L A Y M E S S A G E S ______________________________________________________________________________ DISPLAY EXPLANATION REMEDY ______________________________________________________________________________ Err1 Loss of Configuration Enter Configuration Mode and Reconfigure Err1A Loss of option Enter Option Mode Reconfig ______________________________________________________________________________ Err2 Loss of Zero Calibration Enter Digital Calibration...
Page 14
SECTION 3 INSTALLATION ______________________________________________________________________________ This digital weight indicator is a general purpose unit designed for a wide variety of weighing and scale related applications. In addition to standard tabletop applications, the unit can be equipped with a tilt-stand kit for angled tabletop uses, for wall mounting, or for ceiling (or cab) mounting applications.
Page 18
LOAD CELL WIRING: FIGURE 3.4 NOTE: Traces shown are cut ONLY for six (6) wire (remote sensing) applications. NOTE: The load cell shield wire should be connected to one of the load cell cable clamp screws located on the load cell mating connector. CAUTION! Shielding is connected at ONLY one end (typically at the indicator end).
Page 19
KHA8924 POWER SUPPLY ASSEMBLY WIRING: (AC INPUT) FIGURE 3.5 ______________________________________________________________________________ KGH8924 POWER SUPPLY ASSEMBLY WIRING: (AC\DC BATTERY CHARGER)
Page 21
KHF8924 POWER SUPPLY ASSEMBLY WIRING: (EXTERNAL DC SUPPLY)
Page 23
KGY8924 CPU ASSEMBLY INPUT POWER WIRING : FIGURE 3.8...
Page 24
CONVERTING FROM 117VAC TO 217VAC: CAUTION! THIS SHOULD ONLY BE PREFORMED BY AUTHORIZED SERVICE PERSONNEL - INSTRUMENT WARRANTEE MAY BE VOIDED REMOVE POWER FROM UNIT! Remove rear outer case of unit. Remove rear connector bracket and four standoffs. Remove Power Supply assembly. Remove protective insulator from solder side of assembly.
Page 26
SECTION 4 CONFIGURATION ______________________________________________________________________________ Prior to calibration, the Digital Weight Indicator must be configured, or given its set of operating parameters. To configure the unit : Unscrew the two (2) knurled captive fasteners on the bar of the Front Panel and drop the bar down to expose the four (4) program switches in the center and the Dead Load adjustment potentiometer.
Page 27
The PARAMETER IDENT is a two digit number (1 - 15). The SELECTED DATA is a number or display symbol which represents the parameter data being entered into the unit's configuration data. See configuration table for settings. If the SELECTED DATA is correct, push the "ZERO" ("PARAMETER SELECT") key to rotate to the next parameter.
Page 29
PROMPT PROMPT DISPLAY INTERPRETATION AZM CAPTURE BAND o F F AZM "Off" " 0.5 Displayed Graduations " 1.0 Displayed Graduations " 3.0 Displayed Graduations AZM/PAZ APERATURE " 1.9 % Full Scale Capacity 100 % Full Scale Capacity MOTION BAND o F F NO MOTION INHIBIT "...
Page 31
SECTION 5 CALIBRATION ______________________________________________________________________________ The indicator must now be calibrated. All zero and span calibration is done from the front panel (Exception - When mounted in a NEMA X4 Enclosure, DEAD LOAD pot (RZ2) and S1 are located on the CPU board (KGY8924- ). The calibration procedure is as follows: Ensure that the unit is correctly wired to a load cell or scale base.
Page 33
CALIBRATION - CONTINUED Push the "ZERO" key. The display will read "CAL" for a second, and then go to a "0" ("zero") display. Now place a known calibration weight on the scale. The "GROSS/NET" key becomes the "DIGIT SELECT" key; each push of this key will "select"...
Page 34
COMPUTATION EXAMPLES: Desired capacity = 200,000 lbs x 20 lb Number of grads = 200,000 ) 20 = 10,000 Therefore in the Configuration Mode PARAMETERS 1 & 2 would be set as follows: PARAMETER "1" = 100 (10,000 grads) PARAMETER "2" = 20 (resolution) To determine the live load signal range: Load cells rated at 2mV/V ........
SECTION 6 SERIAL OUTPUT ______________________________________________________________________________ GENERAL: (BIDIRECTIONAL SERIAL PORT #1) The digital weight indicator is equipped with a Duplex (Bidirectional) Serial Communications Port for information transfer to a computer or other "host" device. The Port is normally configured for RS232C operation (Figure 6.1); 20 mA Loop can be selected (Active - Figure 6.2, or Passive - Figure 6.3).
Page 37
SERIAL PORT #1: (ACTIVE 20mA I LOOP) The Bidirectional Serial Port #1 may also be operated as an "Active" 20mA Current (I) Loop (Indicator "Active" - Remote Computer or Device "Passive"). Wiring is done per Figure 6.2 below. TO CONFIGURE PORT #1 FOR ACTIVE 20mA LOOP: JUMPER S3 - 2 to S3 - 3 JUMPER S2 - 1 to S2 - 2 FIGURE 6.2...
Page 38
SERIAL PORT #1: (PASSIVE 20mA I LOOP) If isolation from ground loop problems is desirable, or if reduced power consumption is a consideration, Serial Port #1 may also be wired as a "Passive" 20mA I Loop (Indicator "Passive" - Remote Computer or Device "Active"). Wiring is done as shown in Figure 6.3 below. In addition to Jumper changes, Integrated Circuit U8 MUST BE REMOVED from the CPU.
Page 39
SERIAL PORT #1: (PASSIVE 20mA I LOOP - continued) FIGURE 6.3 GENERAL: (SERIAL PORT #2) The digital weight indicator is also equipped with a second Simplex (Output Only) Serial Port #2. Port #2 may be configured as an RS232C output and as an Active 20mA I Loop. Wiring is done per Figure 6.4 below.
Page 41
SERIAL PORTS #1 AND #2: (DATA FORMATS) The prompts shown in Table 6.1 below control the following serial output parameters: TABLE 6.1 PARAMETER PORT #1 PORT #2 Serial Enable Prompt #11 Prompt #14 Demand/Continuous Simplex/Duplex Prompt #12 Does Not Control Apply Prompt #13 Baud Rate...
Page 43
C. CONTINUOUS OUTPUT - SINGLE LINE (NO "G-T-N") PRINTOUT * (Computer demand format outputed as shown below when OPTION 2 enabled and selected for "cptr". Will output both single & (G-T-N) multiple line.) <stx><pol><DATA><L/K><G/N><status><cr/lf> WHERE: Brackets ("<",">") are shown herein for character definition, and are not sent as a part of a serial transmission.
Page 45
status: One (1) character field data identification used in the continuous (computer) output mode to identify the status of the indicator display. Characters are listed below in order of priority. DESCRIPTION CHARACTER Calibration/Configuration <C> (ASCII 43H) Invalid Mode (Dashes on display) <I>...
Page 46
REMOTE COMMANDS: (SERIAL PORT #1) The bidirectional features of Serial Port #1 can also be used to send remote functional commands to the indicator. REMOTE COMMAND OPERATION GUIDELINES <Z><cr> Zero Scale "Gross" Mode, No Motion, Inside PAZ Aperture <N><cr> Go to "Net" in "Gross"...
Page 47
SECTION 7 OPTIONS ______________________________________________________________________________ Currently the instrument offers two (2) digitally programmable Options; Option 1 configures the digital weight indicator for interfacing with the Analog Output Module and Option 2 when enabled configures the serial output format for Tag, or for selections between "printer or computer" output formats.
OPTION 1 ANALOG OUTPUT * Converts serial data into analog: 0-10 vdc or 4-20 ma. * Directly mounted on the rear of the indicator or operated remotely. * Self powered, 117/217 VAC provides isolated output * Mild steel, painted enclosure 7.26W x 2.06H x 3.31D, wt.2Lbs * Automatic data checking/verification * Serial data "pass-thru"...
Page 50
GENERAL: Serial data from the indicator is modified to include scaled information specifically for the analog output module. The standard communications are also included to allow the module to "pass-thru" printer or scoreboard information. If external switches (ZERO, TARE, etc) are desired, utilize the Duplex port of the instrument.
Page 51
ANALOG MODULE SETUP/WIRING: DESCRIPTION DESCRIPTION 0 - 10 VDC OUTPUT ZERO GROSS/NET 4 - 20 MA OUTPUT TARE ALARM START 7 - 5...
Page 52
+ 20mal (RX+) serial COMMON - 20mal (RX-) serial TX+ (RS232) + 20mal (TX+) serial +5VDC - 20mal (TX-) serial -20MA (SERIAL) DEMAND (PRINT) SWITCH 1 RECEIVE (baud rate from weight indicator) 1200 (switch position) 0 = open 2400 1 = closed 4800 9600 SWITCH 1...
Page 54
ANALOG MODULE SERIAL "PASS-THRU": The serial data from the host is provided as a simplex output with the same format as the host. 7 - 8...
Page 55
Refer to NOTES 1 - 3 on following page before proceeding! OPTION 1 CONFIGURATION OPTION PROMPT INTERPRETATION DISPLAY OP.1 ANALOG OPTION OP.1 OP.1 Analog Option enabled Analog tracks Display Analog tracks Gross Weight Analog tracks Net Weight NOTE 1 Indicator data output on Port1(to Analog) Indicator data output on Port2(to Analog) NOTE 1 Analog module does not pass serial data...
Page 56
NOTE 1 : Use standard setup for serial configuration under menu "11 - 15" (configuration switch S-2 closed) to establish communications before using the Option 1 menu. The selected transmit Port from the indicator to the Analog module must be set the desired baud rate and be in continuous output mode.
Page 57
cptr Computer format (condensed data output) 7 - 11...
Page 58
PARALLEL BCD OPTION (SERIALLY UPDATED) * Converts serial data into full parallel BCD * Tri-state buffer drivers (5 volt logic) * Source or sink up to six (6) milliamps * Direct mount on indicator or remote * Self powered, 117/217 VAC * Mild steel, painted enclosure 7.26W x 2.06H x 3.31D, wt.2Lbs * Automatic data checking/verification * Serial data "pass-thru"...
Page 60
RS232 or 20 ma loop continuous serial output data from a WEIGHT INDICATOR (UMC600, IQ700, UMC555/444 or UMC2000) and update the remote display with the display data from the WEIGHT INDICATOR (sending unit). In addition, the UMC555/444 REMOTE DISPLAY offers...
Page 61
REQUIREMENTS: EPROM - KBL1921-5 The standard UMC444 or UMC555 is easily converted to a REMOTE DISPLAY by installing the KBL1921-5 EPROM. CONFIGURATION: (Configuration Table provided on following page) Close S1 - 2 (CONF). This places the instrument in the configuration mode.
Page 62
CONFIGURATION TABLE PARAMETER PARAMETER DATA FUNCTION SELECT SELECTION PORT 1 BAUD RATE P1.12 1 2 0 0 Baud rate for See Note P1.24 2 4 0 0 receiving wt. below P1.48 4 8 0 0 data P1.96 9 6 0 0 PORT 2 BAUD RATE P1.12 1 2 0 0 Transmit baud...
Page 63
SERIAL PORT 1 (RS232C simplex) TO CONFIGURE PORT #1 FOR RS232C: JUMPER S3 - 1 to S3 - 2 REMOTE DISPLAY WEIGHT INDICATOR 7 - 17...
Page 64
(RS232C duplex) REMOTE DISPLAY WEIGHT INDICATOR (20mA I LOOP simplex) Remote Display configured as a passive receiver REMOTE DISPLAY WEIGHT INDICATOR 7 - 18...
Page 65
(20mA I LOOP duplex) Remote Display configured as a passive receiver and active transmitter. REMOTE DISPLAY WEIGHT INDICATOR SERIAL PORT 2 (Simplex Only) The UMC444/555 REMOTE DISPLAY is also equipped with a second simplex output on Serial Port 2 for Pass-Thru serial I/O. Pass-Thru serial output data is the serial data received on Port 1 of the REMOTE DISPLAY from the WEIGHT INDICATOR and then sent back out on Port 2 of the REMOTE DISPLAY.
Page 66
WIRING TABLE UMC555/444 REMOTE DISPLAY UMC555/444/600/2000 PORT 1 RS232 TB2 - 5 RX (RS232) " " " " TB2 - 3 TX (RS232) " " " " TB3 - 1 sig. common sig. common PORT 1 20MAL TB2 - 1...
SECTION 8 RECOMMENDED SPARE PARTS ______________________________________________________________________________ PART NUMBER DESCRIPTION KGY8924 - 1 STD.CPU/Analog to Digital Assembly KGY8924 - 2 STD.CPU with RFI shielded A/D Assembly KFL8924 - 1 LED Display Assembly KHA8924 - 1 Power Supply KHY7237 TD. Front Panel Overlay KHY7237G UMC444 Front Panel Overlay KDK7238...
Page 68
UMC600 Digital Weight Indicator Installation Manual 30261...
Page 70
Optional and Advanced Features ....................62 8.1 Expanded Serial Communications ....................62 8.1.1 Demand Output Serial Data Format......................62 8.1.2 Continuous Condec Output Serial Data Format..................63 8.1.3 RS485 Data Formats..........................63 8.1.4 Port 1 Remote Serial Commands ......................64 8.1.5 Full Duplex Parameter Entry/Recall......................
UMC600 manual from the Condec distributor digital weight indicator. This manual provides site at www.4condec.com . information on installation, calibration, configuration, and operation of the UMC600.
Front Panel Keypad and Annunciators Figure 1-1 shows the UMC600 front panel. The UMC600 display consist of six seven-segment display digits. Table 1-1 lists the front panel keys and their functions. Low Battery Annunicator ZERO GROSS MOTION CAPACITY 5000 lb. X 0.5 lb/kg ZERO TARE...
Page 73
Table 1-2 summarizes the front panel annunciator functions. Annunciator Function LOW BATTERY On when 12 VDC input voltage drops to approximately 11.4 VDC. ZERO On when scale weight is within ±0.25 displayed graduations of zero. Used in gross or net weighing mode. On when the indicator is in net weighing mode.
The shipping carton should contain the indicator, tilt stand, the manual, and a parts kit. If any Figure 2-1. UMC600 Backplate parts were damaged in shipment, notify Condec and the shipper immediately. Power Connections Standard units are powered by one of the following Enclosure Disassembly options.
VAC Conversion Board Removal The UMC600 can be converted from 115VAC to If you must remove the UMC600 CPU board, use the 230VAC. The following steps are necessary to following procedure: complete this conversion. 1. Disconnect power to the indicator. Loosen cord grips and remove enclosure as described in Before beginning, be sure to Section 2.2...
Instrumentation Setup Notes: All indicators are configured and tested prior to • Test weights should be used to verify scale shipment to ensure that they are fully functional. The performance. unit can be turned on immediately after connecting the • The tilt stand should be secured with 1/4-20 inch input power and the load cells.
The standard connection is designed for 4-wire (non remote sensing) use. To convert to 6-wire (remote sensing) applications, cut the two PC traces on each end of TB1 as shown in Figure 2-7. Traces +EXC +SEN +SIG –SIG –SEN Check load cell –EXC color code for proper wiring.
2.10 Hardware Configuration Jumper pins are provided for: • EPROM chip selection • Front panel switch configuration (DC on/off or batch start) • Power on/off relay configuration • Dual channel (optional) • Time and date option (SRAM selection) These features are dependent on installed options and software. The location and jumper positions are noted in Figure 2-10.
2.11 Serial Port Wiring The UMC600 has two serial communications ports. Port 1 supports full-duplex RS-232, or 20mA current loop (active or passive), or half-duplex RS-485. Switch S1 (shown in Figure 2-12), selects the interface protocol of Port 1. Port 2 supports simplex (output only) RS-232, or 20mA current loop (active or passive). Access to the serial communication ports is through the cord grips (TB2 and TB4) located on the back of the enclosure (see Figure 2-11).
2.11.1 Serial Port #1 Wiring: CPU KGR8924–1 Serial Port 1 supports full-duplex RS-232, or 20mA current loop (active or passive), or half-duplex RS-485. Switch S1 (shown below), selects the interface protocol of Port 1.. DIGITAL OUTPUT RS232 20MA RS485 ACTIVE PASS PORT 1 RECEIVE...
2.11.2 Serial Port #2 Wiring: CPU KGR8924–1 Serial port #2 supports simplex (output only) RS-232, or 20mA current loop (active or passive). The following chart illustrates Serial Port 2 pin assignments. Port Connector/Pin # RS-232 20mA Current Loop TB2 - 1 TB2 - 5 TB2 - 6 TB4 - 1...
Page 82
Typically, digital outputs control relays which operate other equipment. Outputs 1 through 4 allow for setpoint and zero band control. Operational Mode TTL Inputs J11 Normal Batch Relay Board TB3 KGR8924 CPU Tare Start Batch TB3-6 Zero Zero TB3-5 Net/Gross Abort TB3-4 Print...
Configuration Prior to calibration, the UMC600 must be digitally When configuration is complete, set SW1-2 down to configured, or assigned a set of operating parameters. return the unit to normal operating mode. The three parameters listed in Section 3.1.1 are directly related to calibration and must be set before proceeding Parameter Selected Data...
Page 84
3.1.3 Digital Configuration Parameters Table 3-1 (shown below) shows an overview of the 14 configuration parameters and their description. Table 3-2 on page 16 lists the configuration display prompts (Prompt 1), and their value selections for displayed graduations. Prior to calibration, the UMC600 must be digitally configured, or given its set of operating parameters. The first three parameter selections are directly related to calibration and must be set up before proceeding to the calibration mode.
Page 85
Figure 3-4 provides a graphic representation of the 14 configuration parameters associated with the UMC600. Options 1-11 Shown only if Shown only if shown only if SW1-2 is on SW1-1 is on SW1-2 & 3 are on OPTION 6 OPTION 7 OPTION 8 OPTION 9 OPTION 10...
Page 86
Prompt Display Interpretation Notes Displayed Graduations Number of Displayed Graduations = Scale Capacity Resolution 1000 Legal for trade values: 500–10000 graduations 1500 2000 2500 3000 4000 5000 6000 8000 10000 12000 Not valid in NTEP legal-for-trade applications 14000 16000 18000 20000 30000 Not valid in legal-for-trade applications...
Page 87
Prompt Display Interpretation Notes Update Display rate. Sets the update rate for displayed values. Digital Averaging Averages Rate 10/sec 5/sec 2.5/sec 1 sec 2 sec 4 sec 8-4-2 Variable 16-8-4 Variable 8-4-2 Variable Selection available when option 1 is enabled. see Section 4.1. 16-8-2 Variable Tare Enable...
Page 88
Prompt Display Interpretation Notes Display Base (lb/kg) lb/kg CONV key functions only if parameter 10 is set to 10 Con lb display only kg display only lb (base) conversion Table 3-5. Configuration Display Prompts 8 –10(Continued) Prompt Display Interpretation Notes Setpoint mode Four independent modes, three dependent modes.
3.1.4 Normal Configuration Setup Parameters You must be in the configuration mode (SW1-2 closed to set the indicator from the PC) to be able to write configuration parameters. To write commands the following sequence must be used. The following Jxxyyzz<CR> parameters and their setup numbers are as follows: Parameter 1 = Grads J0600 03 = Both...
Serial Configuration The UMC600 has two serial ports. Both serial ports 1 and 2 are ASCII-compatible, 20 mA current loop, or RS-232 outputs. The serial format is compatible with most printers, scoreboards, and other remote devices. Each output can be disabled, set for print on demand mode, or set to output data continuously. Table 3-8 shows the configuration selections for parameter 14, which control the configuration of ports 1 and 2.
Page 91
Subparameter Subparmeter Data Parameter Selection Selection Explanation Net/Gross Tare Lb/Kg Conversion Key which affects associated parameters 14. dU Port 1 duplex 14.1 Demand print configuration in net mode 14.1Gtn Three line output G-N-T 14.1nE Single line, NET print 14.2oFF No delay after CR 14.2 1 second delay after CR 14.2...
Options Configuration The UMC600 offers a selection of optional features that In the option mode, primary function keys ( ZERO , etc.) operate as secondary function keys are available in the configuration of the indicator. The GROSS/NET (Figure 4-2). options setup mode allows the operator to expand the capabilities of the indicator.
Option 1 – Expanded Resolution Enabling option 1 increases the indicator resolution by allowing display graduation selections beyond the normal 20,000 in parameter 1 of the configuration mode. Up to 80,000 displayed graduations are available when this option is on; however, applying such high gains to the data may cause undesirable display instability in some applications.
Option 2 – Analog Output Option Indicator option 2 provides settings for zero, span, and trim adjustments for the analog output. The settings are all digital therefore no potentiometers are required. During the trim adjustments for zero/span, the analog output is forced to the zero/span previously selected in option mode parameters 2.5 and 2.6.
Page 95
Table 4-2 shows the option 2 configuration parameters Prompt Option Display Interpretation Notes All off -------- Mode not enabled (options 1 through 10 — turned off) OP.2 OP.2 Analog option disabled — OP.2 Analog option enabled Analog tracks display — Analog tracks gross weight Analog tracks net weight Indicator data out on port 1 (to analog)
Page 96
Figure 4-5. Analog Module Setup and Wiring SW1 Setting Baud rate 1200 2400 4800 9600 Mode Normal operation 0 VDC/4 mA (test only) 10 VDC/20 mA (test only) 0–10 VDC/4–20 mA ramp (test only) Table 4-3. Baud Rate and Switch SW1 Settings Description Description 4–20 mA...
4.2.5 Analog Wiring to Host Indicator The following diagram illustrates the wiring layout to the host indicator. Active 20 mA 20 mA (passive) active RX+ (passive) Rx - (analog) (analog) PORT 2 PORT 1 Figure 4-6. Analog Wiring to Indicator 4.2.6 Analog Module Serial Pass-Through The serial data from the host digital weight indicator is provided as a simplex output with the same format as the...
Option 3 – Time and Date The main purpose for time and date is to allow the operator to quickly change the time when the time changes between standard and daylight savings time. The time and date clock is attached to the battery-backed memory (U4) and will continue running when the indicator is off.) Prompt Display Interpretation...
Option 4 – Accumulator The accumulate function is used to add weight data to a register for later access by the user. The accumulator can keep a running total of weights entered, either automatically using setpoints, or manually pressing the PRINT when the accumulate function is active.
Option 5 – Expanded Serial Communications The expanded serial communications options provide RS-485 communications for the UMC600. Option 5 features include: • Duplex communication for up to two weight indicators on a single serial port. • RS-485 half-duplex with U21 installed. •...
ID HELD This function is enabled when 5.3 is selected On. When ID held is enabled, the current keyed in ID number is stored in memory and is sent out with each valid demand serial print. There are two methods of ID entry. •...
Option 7 – Serial I/O The serial I/O option offers flexibility for an operator to customize the serial output format for individual system requirements. The selections under option 7 can be divided into three groups. • Customizing of serial output data. •...
Page 104
Option 7 Menu CALIBR OPTION 6 OPTION 7 OPTION 8 OPTION 10 OPTION 11 CONFIG XXXXXXX OPTION 1 OPTION 2 XXXXXXX OPTION 3 OPTION 4 OPTION 5 XXXXXXX XXXXXXX OPTION 9 port 1 gross port 1 net port 1 total port 2 net port 2 gross port 2 total...
4.7.2 Macro File Setup There may be times when you need to print more than the 30 character codes that are allowed in the custom transmit files (7.1 - 7.8). To do custom transmits larger than 30 character codes, the UMC600 provides eight “macro” files (7.9 to 7.16) that can be called in from the custom transmit files.
4.7.3 Standard Serial Configuration The Serial I/O option allows standard serial output ports, configured in parameter 14 of the configuration mode, and options OP3 (time and date), OP5 (tag, ID, counter), and OP4 (accumulator), to be modified and imported into the serial output data stream.
Page 107
Option Prompt Display Interpretation 7.19 Transmit parity selection Mark parity Odd parity Even parity Space 7.20 Receive data parity Disabled – parity observed when receiving data. Enabled – parity is ignored when receiving data. Table 4-15. Option 7 Configuration Parameters (continued) NOTE: Selections 7.19 and 7.20 are available in EPROM (KDA1921-3).
Option 8 – Analog Output/Relay Option Option 8 only addresses the setup of the analog output section on the dual function circuit board. NOTE: Option 8 utilizes the KJN8924, a dual function circuit board mounted internally to the back plate of the indicator. It provides a four relay output for setpoint process control and a 4 - 20mA or 0 - 10 VDC analog output which is configured and trimmed for zero and span in similar fashion as the older option 2 analog output (KHD8924).
When connecting the ribbon cable from J11 of the CPU board to J1 of the analog relay option Caution board, be sure to match pin one of J11 to pin one of J1. Use the color stripe on the cable to identify pin one at each connector end.
4.10 Option 10 – Setpoints This option allows an active setpoint to be operated using the display, gross, or net data. When option 10 is turned on the selection of data that setpoint 1 operates off of is configured in 10.1 and 10.2 for setpoint 2. Table 4-19 shows the option 10 parameters.
4.12 Additional Options Additional options are available with the UMC600. While they are not digitally accessed and set up under the option mode (front panel switches 2 and 3 closed), they are available. Additional options include the following: • Relay board (KHL8924) installed or when ordered as a kit (KBT7116-11/12). •...
Page 112
UMC600 SUPRESSOR TB1-1 OUT1 TB1-2 KGR8924 CPU TB1-3 OUT2 TB1-4 TB1-5 OUT3 TB1-6 TB1-7 OUT4 TB1-8 TB1-9 SPARE SPARE RIBBON CABLE BRINGS CONTROL LOGIC FROM TB1-10 CPU TO THE OUTPUT RELAYS RELAY OUTPUT BOARD KHL8924– POWER TO OUTPUTS Figure 4-13. KHL8924 Relay Output Wiring NOTES: •...
4.12.2 Relay Output/Input Board (KJN8924–) The relay output board KJN8924– is a dual function circuit board that contains both an analog output section (0 – 10VDC or 4 – 20mA) and a relay output/input section with four solid state relays and four solid state input relays. Additional features of KJN8924–...
Page 114
SUPRESSOR OUTPUT RELAYS OUT1 TB2-1 OUT2 TB2-2 K5 THROUGH K8 OUT3 LOGIC CONTROL TB2-3 FROM CPU BOARD OUT4 TB2-4 NOT USED TB2-6 NOT USED TB5-1 KGR8924 CPU CONTROLS ONLY OUTPUTS 1– 4 NOT USED TB5-2 NOT USED TB5-3 KGR8924 RELAY TB2-5 TB3-1 AC COMMON...
Page 115
Operational Mode TTL Inputs J11 Normal Batch Relay Board TB3 KGR8924 CPU Print Print TB3-10 Tare Start Batch TB3-9 Abort Zero TB3-8 Net/Gr Abort TB-3-7 Tare Start Batch TB3-6 Zero Zero TB3-5 Net/Gross Abort TB3-4 Print Print TB3-3 CPU Board TB-4 Print Print TB4-5...
4.12.3 Serial Command Option Parameters The following table lists the option parameter serial commands for the UMC600. Option 1 K0306 00 = TI (time entry) K0701 00 = Off K0717 00 = Off K0906 00 = CHG K0100 00 = Off K0306 01 = DA (date entry) K0701 01 = On K0717 01 = On...
Calibration The UMC600 indicator can be calibrated using single slope span calibration or five-point linearization. Zero must be calibrated (see Section 5.1) before either span or linearization calibration can be performed. Zero Calibration Zero calibration is accomplished by the following steps: 1.
Five-Point Linear Calibration There are five calibration points (La, Lb, Lc, Ld, and Le) that can be entered sequentially. Calibrations using fewer than five points will linearize the curve up to the last data point. The curve is then extrapolated from the last entry point.
Normal Weighing Mode Operations This section provides the operator with a description of front panel key functions and associated annunciators (LEDs) used to operate the UMC600 in the normal weighing mode. After the unit is configured and calibrated, the unit is placed in the weighing or normal weighing mode (SW1-1 through SW1-4 open).
6.4.3 TARE Key Function 6.5.1 Lb/Kg Conversion If the tare acquisition is greater than zero (+1/2 When parameter 10 in the configuration mode is set to graduation), set the display to the net mode and apply Con (conversion), pressing the key toggles the UNIT the new tare.
Serial Output Pressing the local PRINT key or issuing a serial print command (unit must be set up for bidirectional serial communication). 6.6.1 Serial Data Formats Serial data is transmitted in ASCII-compatible format and consists of the following: • One start bit •...
Status Character Definitions The status character (<STAT>) provides information to the receiving device about the current indicator operation. Table 6-4 shows the status characters returned by the UMC600 indicator. Status Characters Description <D> (44H) Digital calibration mode <A> (41H) Analog calibration mode <I>...
Truck Weighing Mode The truck weighing mode provides a weigh-in/weigh-out system for single scale applications. The UMC600 has a system capacity of up to three hundred transactions, each with up to six digit identification numbers (ID’s). The truck weighing system is configured using parameter 13 and transaction printouts are performed only on port 2 and the operation is only in gross mode.
Page 124
Mode Functional Description Full Truck In mode Transaction begins with a full truck (gross weight) entering landfill, and leaves empty (tare weight). The transaction uses the following steps. Truck enters the scale full. Scale settles out of motion. PRINT Driver inserts ticket and presses Display responds with ID No prompt.
Setpoints The UMC600 is equipped with four independent control outputs, which are 5VDC logic level signals capable of sinking 12mA. Setpoints can be configured in the following modes: • Mode A – Setpoint and zero band • Mode B – Setpoint, with preact, and zero band •...
Page 126
Mode Description Setpoint Mode C Setpoint with dribble control Dribble control provides two-speed cutoff (fast or slow feed) to enhance accuracy. Cutoff occurs SP1-Dr when WT≥ SP – dr – Pr. Use the following steps to set Mode C setpoint values. SETPOINT 1 SETPOINT 2 Press...
Page 127
Mode Description Setpoint Mode E One setpoint with preact, dribble control, and tolerance band output SP1-Pr Converts setpoint two (SP2) to be a tolerance output when WT > SP1 – Lo, but < SP1 + HI. SP1-Pr-dr Use the following steps to set Mode E setpoint values SETPOINT 1 Enter value for SETPOINT 2...
Page 128
Mode Description Setpoint Mode G Over/Under checkweigh control Provides High, Low, and Accept outputs except a direct entry is made for a target weight and high/ ACCEPT low settings become dependent. HIGH Use the following steps to set Mode G setpoint values. SETPOINT 1 Press Enter the desired target value;...
Page 129
Modes A, B, C, and D can be used in a independent straight setpoint system or as part of either a manual batching or automatic batching system. When configured for batching, the setpoints are de-energized except when the batch is in process.
Page 130
Parameter Data Subparameter Subparameter Selection Selection Data Selection Explanation Net/Gross Tare Lb/Kg Key which affects associated parameters 11.Ou.Un Over/Under scale configuration 11.1 11.1.hL Mode F entry 11.1.tGt Mode G entry 11.2 11.2POS Low is off when weight is below range 11.2.zEr Low is on when weight is below range 11.3...
Page 131
Parameter Data Subparameter Subparameter Selection Selection Data Selection Explanation Net/Gross Tare Lb/Kg Key which affects associated parameters 11.5 11.5.POS Setpoint 2 is off when weight < SP 11.5.zEr Setpoint 2 is on when weight < SP 11.6 11.6 oFF Timer #1 oFF (pause) 11.6 delay (seconds) 11.6...
Optional and Advanced Features Expanded Serial Communications The UMC600 has two serial ports. Both serial ports 1 and 2 are ASCII-compatible, RS-232, or 20 mA current loop outputs. The serial format is compatible with most printers, scoreboards, and other remote devices. Each output can be disabled, set for “print on demand”...
8.1.2 Continuous Condec Output Serial Data Format The continuous mode output serial data format is used to interface computers, scoreboards, and other remote devices requiring constant data updating. Continuous mode transmission occurs at the end of each display update. <STX> <POL> <DATA> <L/K> <G/N> <STAT> <CR> <LF>...
If the initiating device address matches the port address of an UMC600 listening on the RS485 network, the indicator responds with the following format shown in Figure 8-4. <:> <00> <DATA DEMAND> <P> <POL> <DATA> <L/K> <G/N> <STAT> <CR> L = pounds Print ASCII 13 Message...
8.1.5 Full Duplex Parameter Entry/Recall One of the capabilities of the UMC600, with the full duplex serial port 1 is the ability to enter or recall setpoints from a computer or terminal. The following table illustrates full duplex setpoint parameter entry/recall. Setpoint Parameter SE<n>...
Appendix ASCII Character Chart Use the decimal values for ASCII characters listed in Table 9.1 when specifying print format string. The actual character depends on the character mapping used by the output device. The UMC600 can send or receive any ASCII character value (decimal 0-255), but the characters that can be shown on the indicator are limited by the 14-segment display.
Parameter Control Code Chart Use the following code values for parameter descriptions listed in Table 9-2 when specifying the format string. The actual character depends on the character mapping used by the output device. Code Description Code Description Code Description Gross weight &...
Specifications Power Environmental Power Input 115/230 VAC, 50-60 Hz AC Operating Temperature –10 to +40 °C (14 °F to 104 °F) Analog Specifications Enclosure Full Scale Input Signal Up to 33 mV Enclosure Dimensions 9.0 in x 6.44 in x 4.0 in Load Cell Excitation 10 VDC, fixed 23 cm x 16 cm x 10 cm...
Page 139
OPTIONAL PANEL MOUNTING KIT 4.14 9.12 PANEL THICKNESS .03 THRU .75 ZERO GROSS MOTION CAPACITY 6.35 6.62 3.94 FRONT VIEW SIDE VIEW MODEL NO. SER. NO. DATE INPUT VOLTAGE VAC 50/60 Hz ADD'L INFO SCALE ACCESSORY LR47913 ® REAR VIEW Appendix...
Condec. Condec warrants the equipment against faulty workmanship and defective materials. If any equipment fails to conform to these warranties, Condec will, at its option, repair or replace such goods returned within the warranty period subject to the following conditions: •...
Need help?
Do you have a question about the UMC555 and is the answer not in the manual?
Questions and answers