Page 1
SterOx System V-Series Short user guide The SterOx System V-Series short user guide does not replace the SterOx System V-Series instructions for use. For information on intended use, indications, contraindications, warnings and precautions, please refer to the instructions for use supplied with the SteriBase V-Series.
Page 2
For the training to be fully effective, end-users of the device should be present during the installation meeting. If you have decided to use commercially available nitrogen supply cylinders that are not distributed by SteriLux, you must be extra careful when connecting the nitrogen supply to your device as it could permanently alter the efficiency of the device if done improperly.
Products overview SteriBase V-Series SteriBox V-Series Touchscreen Start button Quartz window Handle Incorporated RFID tag Manual levers Cooling fan Module Thermal printer Bottom Nitrogen supply plug Locks Power switch Power cable port Humidify Nitrogen Kit Ozone Strip Ozone Strip Blotting Paper Type 1 Chemical Indicator Type 4 Chemical Indicator 2.2L disposable nitrogen...
Installation Select a suitable location to install your SteriBase: Away from direct sunlight or a heat source Ideal ambient temperature: 20-23°C Clean, dust-free room with a good ventilation system Plug your SteriBase into a single electrical socket Turn the power switch to ON and click on the start button Take your nitrogen kit Remove the rubber protection from one nitrogen cylinder Take the pressure reducer and verify that the knob is in closed position (the closed position of the knob is all the way up)
Setting up your device From the main menu select WIFI to connect your device to a wireless network From the PARAMETERS menu: Set the date and time by clicking on Date and time (sets automatically if you’re connected to the WiFi) Set your preferred language by clicking on Language OPTIONAL: add users and/or contents by clicking on User/content settings OPTIONAL: add your email address by clicking on Email settings to automatically receive cycle reports...
Using your device Prior to launching a cycle: Check your SteriBox, it must always be clean and dry – pay special attention to the quartz window and module Instruments to be sterilized must be clean and dry Avoid preparing and closing the SteriBox in advance Launching a cycle: Open your SteriBox and remove the module Arrange your instruments –...
Page 7
Using your device Cycle report 9 At the end of the cycle: ###### SteriBase 000010 ###### SteriBox 000031 Your SteriBase automatically prints a label (i.e. CYCLE COMLPETED or CYCLE FAILED) Cycle number If you have entered an email address, you will receive the cycle report in your inbox Cycle type Pouch Performed by...
Page 8
Maintenance & software updates Always make sure that your SteriBox is spotless and pay special attention to the quartz window and module In case you have not been using your device for more than 3-4 weeks or if you are experiencing failed cycles due to insufficient ozone dose (E429) or ozone concentration measure too low (E433): Verify that nitrogen supply is well connected and not empty From the main menu select MAINTENANCE...
Customer support SteriLux remains at your disposal to assist you with the installation, maintenance and service of your device. Support is available from Monday to Friday from 8:00 to 18:00 (CET). support@sterilux.ch +41 (0) 21 546 43 34 SteriLux SA...