Page 1
Start up Guide Series Balances Transcat.com | 800.828.1470...
Page 2
1. CONTENT Models: TCS-203 Balance Draft shield Weighing pan Power adapter ×1 ×1 ×1 ×1 Models: TCS-602 | TCS-2002 | TCS-3101 Balance Draft shield Weighing pan Power adapter ×1 ×1 ×1 ×1 *The plug type may vary by country. 2. UNPACKING Keep the packaging should a warranty claim or service be required.
3. RECOMMENDATIONS AND WARNINGS Operate the device in a room where the temperature ranges +40 °C between 10–40 °C (50–104 °F) and where the relative humidity is +104°F below 80%. +10 °C Place the balance on a solid surface to ensure stability. To obtain +50 °F stable and repeatable weighing results, an anti-vibration table is recommended.
5. COMPONENTS ASSEMBLY Models: TCS-203 Models: TCS-602 | TCS-2002 | TCS-3101 6. PREPARING FOR WORK 6.1. Temperature stabilization time Prior to switching the device on, let it Connect the power Connect the reach room temperature (estimated adapter to DC power adapter to stabilization time: 1-8 hours).
6.3. Switch the balance on Your balance is ready to use! To switch the device on, press ON / OFF key. 7. PANEL AND SCREEN 7.1. Panel keys Switching Navigation ON/OFF and taring Selecting working Navigation mode / ESC and zeroing Sending weighing result to a printer or a computer / ENTER...
Page 6
8. BASIC OPERATIONS 8.1. Adjustment External adjustment is carried out using an external mass Balance Mass standard of the right accuracy and weight value, which model standard [g] value depends on balance type and capacity. TCS-203 TCS-602 Prior to adjustment prepare the right mass 2 000 TCS-2002 standard.
Page 7
Unload the weighing pan. Adjustment completed. 8.2. Zeroing key. The balance has been zeroed. With unloaded weighing pan press 8.3. Taring With loaded weighing pan, The balance has been tared. upon result stabilization, press key.
8.4. Working modes accessibility Simultaneously press keys to enter Using key select “P3.Func” parameter home screen. and press key. Using key select "3.1.1Acc" parameter While “3.1.UUGG” parameter is on, press key. and press key. In order to enable other working modes, repeat the steps Save working mode settings - section...
8.5. Selecting working mode Using key enter working modes menu. Using key select working mode and press The menu features the following working modes: Detailed overview of the settings of the working modes is to be • • Weighing Percent found in balance’s main user •...
10. DEVICE CLEANING 10.1. General Recommendations Apply the cleaner onto the cloth. Prior to cleaning it is necessary Use soft cloth made of either Mind not to apply the cleaner to disconnect the device from natural or synthetic fiber directly onto the device. the mains.