This operating instructions contains safety information that if ignored can endanger life or result in serious injury. The original instruction is in English. Read these instructions carefully before use and keep them for future reference. Information and specifications on this manual could be uncorrect or could have printing errors. Specifications are subject to change without notice.
PURPOSE OF USE EQUIPMENT INTENDED FOR THE MEASUREMENT OF CHLORINE IN WATER. AND SAFETY Do not use in explosive area (EX). Do not use with flammable chemicals. Do not use with radioactive chemicals. Use the equipment in accordance with the data and specifications printed on the label. Do not modify or use in a manner inconsistent with the provisions of the operating manual.
Environmental Work area safety Always keep the area clean to avoid and/or discover emissions. Recycling guidelines EWC code: 16 02 16 Always recycle according to these guidelines: 1. If the unit or parts are accepted by an authorized recycling company, then follow local recycling laws and regulations.
Page 6
Transportation and An unsuitable transportation or storage can cause damages. storage Use original box to pack the amperometric cell. Observe storage conditions also for transportation. Although packed, always protect the unit against humidity and the action of chemicals. ⎘ Before returning the probe to the manufacturer repair service, refer to Decommissioning.
INTRODUCTION Open ECL Series is designed for measuring free chlorine (both organic and inorganic). amperometric Open amperometric cells comprise an Off-line probe holders, a sensing electrode and a flow electrode. cells Probe holders can contain up to three probes (temperature, pH and ORP). The flow of water within this cell must remain constant and within 40 l/h.
MODELS • ECL6 for free chlorine (organic and inorganic) - pH and ORP probe-holder • ECL6/E for free chlorine (organic and inorganic) • ECL12 for free chlorine (organic and inorganic) for salt water - pH and ORP probe-holder • ECL12/E for free chlorine (organic and inorganic) for salt water...
ECL12 ECL12 ECL12/E ECL12/E Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05 Connection 2 wires (+red; -black) Measuring system amperometric - 2 electrodes (platinum/silver) Ph working range 6-8 pH First polarization: 2 h about Run-in-time Next polarizations: 50 min.
Operating ECL amperometric cells have an amperometric measuring system. principle Measurement method is an electrochemical technique: the two electrodes (platinum and copper, or platinum and silver) produces a current directly proportional to the concentration of chlorine. Precautions Before any operation (preparation, cleaning and replacements) and before handly the amperometric cell you MUST FOLLOW these PRECAUTIONS.
Installation Secure amperometric cell on a stable wall by screwing the two fixing screws on the body of the amperometric cell (⎘ Dimensional). Connect water inlet and outlet to the pipe. We recommend the installation of a filter before the amperometric cell. Install pH, ORP and temperature probes (if any) in their seats.
07/02/2017 ±0.05 Materiale: Commissioning Open water flow into the cell turning water inlet valve. ECL6 CELLA AMPEROMETRICA Edizione Foglio Esploso_ECL6 1 / 1 Turn until the float reaches the black mark on the cell. SEPR LED lights to indicate the proper flow into the cell.
Electrode replacing DANGER AUTHORIZED AND QUALIFIED PERSONNEL Installation and maintenance tasks should be carried out by AUTHORIZED AND QUALIFIED PERSONNEL only in accordance with local regulations. DANGER POWER SUPPLY DISCONNECTION Always disconnect power to the equipment before you perform any installation or maintenance tasks.
Calibration Perform calibration of amperometric cell on a monthly basis or more, if necessary. Amperometric cell must be calibrated with the instrument to which it is connected. Calibration is carried out on two points: 0 and a second point near the work value. For calibration procedure refer to the manual of the instrument.
MAINTENANCE Maintenance DANGER schedule MAINTENANCE SCHEDULE In order to ensure the requirements of potable drinking water treated and the maintenance of the improvements as declared by the manufacturer, this equipment must be checked at least once a week. DANGER OPERATOR PROTECTION Use safety equipment according to the company regulations.
TROUBLESHOOTING PROBLEM CAUSES SOLUTIONS Sensor cannot be calibrated Run-in period too short Observe the run-in period and measured value Troublesome substances in Check water for troublesome greater than DPD measurement the water substances and contact supplier Short circuit in the measuring Identify short circuit and line eliminate the cause.
ECL12/E Progettato da: Controllato da: Data: Peso Lordo: Peso Netto: Toll. Gen. Massimo_F Massimo_F 22/07/2015 ±0.05 Materiale: ECL12E CELLA AMPEROMETRICA Edizione Foglio ECL12E 1 / 1 9.00 81.00 171.70 Progettato da: Controllato da: Data: Massimo_F Massimo_F 22/07/20 ECL12E CELLA AMPEROMETRIC...
Page 22
MOD 7.5 B1 Q Ed. 1 - rev. 0 21/02/2012 PRODUCT SERVICE REPAIR FORM ENCLOSE THE PRESENT FORM TO THE DELIVERY NOTE DATE ..........SENDER Company name ............................Address ..............................Phone no..............................Contact person............................. PRODUCT TYPE (see product label) DEVICE CODE ..............................
Page 24
Disposal of end-of-life equipment by users This symbol warns you not to dispose of the product with normal waste. Respect human health and the environment by giving the discarded equipment to a designated collection center for the recycling of electronic and electrical equipment. For more information visit the online site.