PSETUP
Command
PDATAOUT
Command
Module Specific Syntax 3-2
Figure 3-1, Relay Diagram
The PSETUP command causes the specified module setup to be
displayed as a response. The syntax used is:
PSETUP <module address>[;<module address>]
[;<module address>]
where <module address> is the switch card address.
The responses to the PSETUP command for the 1260-12 Relay
Actuator Module are as follows:
<module address>.1260-12 RELAY ACTUATOR MODULE
<module address>.BBM
<module address>.END
The response to the PSETUP command consists of a header on
the first line. The header describes the model number. The next
line designates the setup mode for the scanning which, by
default, is Break-Before-Make (BBM). The last line containing
the "END" characters denotes no more information to report.
The PDATAOUT command causes the specified module to
transmit the state of the CLOSED relays (within the switching
module) to the 1260 Controller. The syntax used is:
PDATAOUT <module address>[;<module address>]
[;<module address>]
The responses to the PDATAOUT command are as follows:
<module address>.1260-12 RELAY ACTUATOR MODULE
1260-12 User Manual
EADS North America Defense
Test and Services, Inc. © 1993