Software Manual
Control Panel for Pegasus Units
Models: MD 421
MD 422
Software Manual ............................................
1-18
EN
V. 4.14

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MD 421 and is the answer not in the manual?

Questions and answers

Summary of Contents for Ecolab MD 421

  • Page 1 Software Manual Control Panel for Pegasus Units Models: MD 421 MD 422 Software Manual …………………………………….. 1-18 V. 4.14...
  • Page 2: Table Of Contents

    1. DESCRIPTION 2. DISPLAY LAYOUT 3. MODES 3.1 Power On 3.2 Standby 3.3 Power off 3.4 Error / Warning 4. CONTENT OF MAIN- AND SUBMENUS 5. OVERVIEW OF MAIN- AND SUB-MENUS 5.1 Content of the Info menu 5.2 Content of the Errors menu – CAUSE & REMEDY 5.2.1 Error List 5.3 Content of the Setup menu 5.3.1 Submenu setting...
  • Page 3: Description

    1. Description This software manual describes the operation of the control panel of the Single Booster unit. The unit is operated via the control panel (see below illustration). 1. OFF Button (1) switches OFF the unit. 2. ON/STANDBY Button (2) switches ON the unit. 3.
  • Page 4: Display Layout

    2. Display layout Example: Standard Standard pressure: 25 bar pressure: 25 bar When switched on the display will show an arrow symbol in the top right corner equivalent to the active navigation buttons on the control board. The arrows illustrate that it is possible to scroll up and down between the menus. Scrolling to the right makes it possible to enter a value.
  • Page 5: Modes

    Date & time 2006-01-01 12:00 Set date and time using navigation buttons . Set date & time 2006-01-01 12:00 Confirm by pressing OK button. 3.2 Standby The start up display will look as the one below. ECOLAB Pegasus 12:00:00 110003604B-EN Software Manual_Pegasus.
  • Page 6: Power Off

    3.3 Power Off When the unit is off the display will only show a small dot in the top right corner. In Off mode it is possible to access all menus without the pump running by pressing the OK button on the panel. Dot is equivalent to the OK confirm button on the panel.
  • Page 7: Content Of Main- And Submenus

    4. Content of Main- and Submenus The control panel contains the below main menu screen displays. Scroll between the menus using the navigation buttons. Each main menu contains submenus which will be described subsequently. Info Errors Setup Readings 110003604B-EN Software Manual_Pegasus.
  • Page 8: Overview Of Main- And Sub-Menus

    5. Overview of Main- and Sub-menus 1. Info (A) 1 Info (AEA) 2. Errors (B) 3. Setup (C) (default password 6802) 1 Password (CA) 2 Settings (E) Standard pressure (ER) Auto off delay (EZ) Date & Time (EP) 3 Languages & units (F) Language (FA) Units (FB) 4 Communication (I)
  • Page 9: Content Of The Info Menu

    C2 Low Supply Pressure sure. 2. Check inlet filter for impurities/rinse filter 3. Contact your Ecolab technician C3 High Water 1. Water pump top temperature is above 80° 1. Lower the inlet water temp. (max Temp. 2. Water consumption is too low (pump column has 70°C)
  • Page 10 1. Insufficient water supply C8 Low Tank 1. Check correct water supply Level 2. Water supply valve not open 2. Secure suffcient air supply to valve 3. Contact your Ecolab technician 1. Secure sufficient air supply to the C9 High Tank 1. Water supply valve not closed Level valve.
  • Page 11: Content Of The Setup Menu

    Password Note: Default user password is 6802. Enter the settings menu to change the default password. To avoid unintended access to the system the Languages & Ecolab technician can insert a blocking preventing access to the password units controlled menus.
  • Page 12: Submenu Setting

    2006-01-01 12:00 2006-01-01 12:00 Set date and time 5.3.2 Submenu Language & Units Language: Languages & Language: units English English Units: Units: Metric Imperial 5.3.3 Submenu Installation Installation * For the use of authorised Ecolab technician only 110003604B-EN Software Manual_Pegasus.
  • Page 13: Content Of Readings

    5.4 Content of Readings Output frequency of frequency inverter. Readings Pump state Pump-top pres. 24 bar Pump-top temp . 29 °C Supply pressure 2.5 bar Supply temp. 29 °C Flow status No Flow 110003604B-EN Software Manual_Pegasus.
  • Page 14: Upgrade Firmware

    6. Upgrade Firmware Upgrading from software version 4.10 or higher 6.1 Upgrade using USB Cable Connect USB cable (Item No. 110001526) to PC USB port (Remove other USB cables connected to PC) Connect Display cable to USB cable In the menu go to, Setup->”password”->Communication-> Firmware upgrade by cable- >Confirm Firmware upgrade and press “OK” Display should now be looking like this: Upgrade Ready Power Cycle and...
  • Page 15 Press UPGRADE button in FWU 200 tool 14. After a short while the FWU tool should start upgrading, the upgrade button should turn Yellow and change to “RUNNING” 15. The progress bar “Program” must go to the end before the upgrade is complete (less than 1 minute) 16. When the upgrade is complete the FWU tool should be looking like this, Button turns Green and text changes to “DONE/NEW” Now the display is upgraded 18. To verify software version in display, in the menu go to, Info->Machine info 19. Here you will find display software version and control software version. SW ID “Display software”, HW ID “type of device”, CT ID Controller board software(if display is not connected to a control board, control software will be 0/300) 110003604B-EN Software Manual_Pegasus.
  • Page 16: Upgrade Using Irda Cable

    6.2 Upgrade using IrDA Cable Connect power to equipment In the menu go to, Setup->”password”->Communication-> Firmware upgrade by Infrared- >Confirm Firmware upgrade and press “OK” Display should now be looking like this: Upgrade Ready Power Cycle and Start PC App. Turn Off equipment and turn it on again (Power cycle) Make sure the All three led’s are on (Green “ON”, red “ERROR” and red “OFF”) and the display is blank Connect Irda module (Item No.
  • Page 17 12. When the file is located press “Open” in browser window 13. The file name will now be written in file text line Press UPGRADE button in FWU 200 tool 15. After a short while the FWU tool should start upgrading, the upgrade button should turn Yellow and change to “RUNNING” 16. The progress bar “Program” must go to the end before the upgrade is complete (less than 1 minute) 17. When the upgrade is complete the FWU tool should be looking like this, Button turns Green and text changes to “DONE/NEW” Now the display is upgraded 19. To verify software version in display, in the menu go to, Info->Machine info 20. Here you will find display software version and control software version. SW ID “Display software”, HW ID “type of device”, CT ID Controller board software(if display is not connected to a control board, control software will be 0/300) 110003604B-EN Software Manual_Pegasus.
  • Page 18: Upgrade Controller Board

    6.3 Upgrade Controller Board Make sure display is connected to controller Power up equipment In the menu go to, Setup->”password”->Communication-> Upgrade Controller->Confirm upgrade and press “OK” Display will now look like this: Upgrade Ready Power Cycle and wait for upgrade Turn off power to equipment and turn it back on (Power cycle) After power on display will look like this: Upgrading Erase 0%...
  • Page 19 110003604B-EN Software Manual_Pegasus.
  • Page 20 © 2006 All rights reserved ECOLAB Engineering GmbH F&B Engineering Competence Centre EMEA Raiffeisenstrasse 7 D-83313 Siegsdorf, Germany Telephone: (+49) 8662/610 Fax: (+45) 8662/61235 Email: engineering-mailbox@ecolab.com...