Maintenance; Warranty - JP Selecta PERCOM N-M II Manual

Peristaltic pump
Table of Contents

Advertisement

Available languages

Available languages

MANUAL DE INSTRUCCIONES
J
.P. SELECTA s.a.u.
CODIGO 80357 REV A
8.3 Tubing Technical Parameters

9. Maintenance

Unplug the equipment from the mains before removing the oven's cover to
handle inside.
The manipulation of the equipment internal electronic circuits by un-
authorized personnel can cause irreparable damage.
Take it to one of J.P.SELECTA's authorized technical services.
The pump is life time lubricated.
Check condition of power cord (and foot switch tube) annually for cracks and wear.
When pump is not working, please loose the cartridges of pressing the tubing for
avoiding changing the shape of tubing because of longtime extrusion.
Pump head cannot resist super corrosive liquid. Please pay attention to it when
it is using.
Keep the rollers of pump head clean and dry. Do not use aggressive solvents.
If the surface of rollers is not clean, they can quicken the tubing wearing, and
reduce the useful life of tubing.
If there is liquid on the rollers, please dry them. Longtime moisture can damage
the rollers.

10. Warranty

This product is guaranteed for one year. It does not cover damages caused by an
incorrect use or causes beyond the control of J. P. SELECTA, s.a.u.
Any manipulation of the equipment by personnel not authorized by J.P. SELECTA,
s.a.u. automatically cancels the guarantee.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
OCT/2018
(Sujetas a modificaciones sin previo aviso) Pag.: 15
Tel 34 937 700 877 Fax 34 937 702 362

Advertisement