Gain; Compensation; Scale - YUSHI YUT-2800 Manual

Table of Contents

Advertisement

waveform will move horizontally. The adjustment range is from 0 to 2000mm.
On the parameter menu, press
press
(left decrease) or
On the displacement submenu, press OK
100.

3.1.4 Gain

The gain submenu enables you to adjust the sensitivity of the instrument. Its unit is dB. The higher
is the gain, the more sensitive the instrument is, and vise versa. The adjustment range is from 0 to
110dB.
On the parameter menu, press
(left decrease) or
On the gain submenu, press Ok

3.1.5 Compensation

Owing to the different degrees of roughness of surface of the detected object and the test block
calibrating the instrument, in the practical detection, you need to increase (or decrease) the sensitivity of
instrument to reduce the influence caused by the surface of the detected object. The range of adjustment
is from -40 to +40 dB. The values used in the adjustment are related to the gain values.
On the parameter menu, press
then press
(left decrease) or
On the compensation submenu, press Ok
6.0.

3.1.6 Scale

This function enables you to adjust the display mode of the screen coordinate.
On the parameter menu, press
select the scale submenu, then press
modes.
YUSHI INSTRUMENTS
(up) or
(right increase) to adjust the values.
(up) or
(right increase) to adjust the values.
to adjust the steps. The steps are 0.1, 1.0, 2.0 and 6.0.
(up) or
(right increase) to adjust the values.
to activate the function. Press
(left) or
8
(down) to select the displacement submenu, then
to adjust the steps. The steps are 1, 10, 25, and
(down) to select the gain submenu, then press
(down) to select the compensation submenu,
to adjust the steps. The steps are 0.1, 1.0, 2.0 and
(right) to switch among various display
(up) or
(down) to

Advertisement

Table of Contents
loading