Controls, Connectors And Visual Indicators - Emerson 3200 User Manual

Emerson cough assist model 3000, 3200 user's guide

Advertisement

C
O N T R O L S
o
7
SECTION 4
,
C O N N E C T O R S
C A - 3 0 0 0 , C A - 3 2 0 0
F
R O N T
1. Manual/Auto: Changes the cycling mechanism to either manual or auto-
2. Inhale: Sets time interval (sec) for Inhale phase of automatic cycling.
3. Exhale: Sets time interval (sec) for Exhale phase of automatic cycling.
4. Pause: Sets time interval (sec) for Pause phase of automatic cycling.
5.
Power: The power switch turns on or off the CoughAssist MI-E. The "I" symbol on the
switch designates the on position. When activated a green light within the switch will
illuminate. The "O" symbol designates the off position.
6.
Manual Control Lever: Use to manually cycle the unit to inhale or exhale. Not operative
in the automatic mode.
7.
Pressure: Varies the inhalation and exhalation pressures together (also see Inhale Pressure).
8.
Handle: Recessed carrying handle.
9.
Patient Port: Connection for patient circuit.
10.
: Symbol for Type B Equipment with F-Type applied part.
11. Inhale Flow: There are two inhalation flow settings: Full (
NOTE: When using reduced inhalation flow there will be a small reduction in inhale pressure.
12. Inhale Pressure: Varies the inhalation pressure between 50% and 100% of the
exhale pressure.
13. Pressure Gauge Zero Adjust: Access to "zero" adjustment on the pressure gauge.
Use only if the gauge does not return to "O" when the unit is turned off (see Section 8).
14. Pressure Gauge: Indicates pressure in the patient circuit calibrated in cm H
&
V I S U A L I N D I C AT O R S
P
C
A N E L
matic mode.
Not operative in the manual mode.
Not operative in the manual mode.
Not operative in the manual mode.
O N T R O L S
) and Reduced (
).
O.
2