ELMED ENDO TEST II Instructions For Use

Advertisement

Quick Links

INSULATION TESTING ON LAPAROSCOPIC INSTRUMENTS

1. I
A S
(
NSERT
PONGE
CHOOSE SPONGE TO BE USED BY INSTRUMENT SHAFT DIAMETER
5
, (C2), 10
MM DIAMETER
,
ELECTRODE
USING THE THUMB AND THE INDEX FINGER
2. T
HOROUGHLY MOISTEN THE SPONGE WITH SALINE
I
F USED IN STERILE FIELD
3. I
NSERT THE INSULATION TEST ELECTRODE

TESTING SHAFT INSULATION ONLY

1. A
C
(1)
TTACH
ABLE
FEMALE
.
TEST UNIT
2. H
OLD INSTRUMENT AT HANDLE
PUSH INSTRUMENT SHAFT ALL THE WAY TOWARDS THE HANDLE ASSEMBLY
ON THE UNIT ALARM LIGHT AND BUZZER WILL SOUND ALERTING USER OF THE DEFECT
3. D
O A PHYSICAL INSPECTION OF THIS AREA TO CONFIRM AND REPAIR INSTRUMENT AS NECESSARY BEFORE USING IT IN A
SURGICAL PROCEDURE

TESTING SHAFT AND HANDLE INSULATION

L
I
APAROSCOPIC
NSTRUMENT
1. A
(1)
TTACH CABLE
FEMALE BANANA PLUG TO INSTRUMENT
.
TEST UNIT
2. P
LUG IN INSULATED TEST ELECTRODE HANDLE
INSULATED TEST ELECTRODE HANDLE
STICKING OUT ON TOP
3. T
HOROUGHLY MOISTEN SPONGE WITH SALINE
4. H
OLD THE INSTRUMENT AT THE HANDLE IN ONE HAND AND WITH THE OTHER HAND HOLD THE INSULATED TEST ELECTRODE
. W
HANDLE
ITH THE UNIT TURNED ON
. I
PARTS OF THE SHAFT
5. P
ROCEED TO WIPE DOWN THE HANDLE IN THE SAME FASHION
ACTIVATE WHEN THE DEFECTIVE SPOT IS TOUCHED
6. D
O A PHYSICAL INSPECTION OF THIS AREA TO CONFIRM FAILURE AND REPAIR INSTRUMENT AS NECESSARY BEFORE USING IT IN
A SURGICAL PROCEDURE

INSULATION TESTING ON MONOPOLAR + BIPOLAR FORCEPS

M
F
ONOPOLAR
ORCEPS
1. C
ONNECT THE ALLIGATOR CLIP
(A). C
SOCKET
LIP THE ALLIGATOR CLIP TO EXPOSED METAL ON THE FORCEPS TO BE TESTED
(3)
HANDLE
TO SOCKET
THE APPROPRIATE SIZE WITH
).
OUT EXCESS WATER
2. H
OLD THE FORCEPS IN ONE HAND AND THE INSULATED TEST IN THE OTHER
,
PART OF THE SPONGE
3. I
F AN INSULATION FAILURE IS DETECTED THE FAILURE LIGHT AND AUDIBLE ALARM WILL ACTIVATE
PHYSICAL INSPECTION OF THE AREA AND PERFORM NECESSARY REPAIRS BEFORE USING THE INSTRUMENT IN A SURGICAL
.
PROCEDURE
, (C3),
MM DIAMETER
INTO THE TEST ELECTRODE
,
,
FLUIDS SPONGE
ELECTRODE MUST BE STERILIZED AND FLUIDS USED MUST BE STERILE
(2)
INTO SOCKET
B
ANANA PLUG OF THE INSTRUMENT
,
ACTIVATE FOOTSWITCH ON UNIT
.
(3)
(3). I
NSERT A SPONGE INTO THE TEST ELECTRODE WITH ABOUT
.
(
SUGGESTED
,
ACTIVATE THE FOOTSWITCH AND SWIPE THE UPPER PART OF THE SPONGE ALONG ALL
F THERE IS AN INSULATION FAILURE THE FAILURE LIGHT AND ALARM WILL ACTIVATE
.
(4)
TO THE FEMALE END OF THE TEST CABLE
(B)
OF THE UNIT AND INSERT INSULATION TEST ELECTRODE
¼"
STICKING OUT ON THE TOP AND MOISTENED IT WITH SALINE OR DISTILLED WATER
SWIPE ALL INSULATED PARTS OF THE FORCEPS
ENDO TEST II MANUAL
I
NSTRUCTIONS
U
S
U
:
NIT
ET
P
1. P
U
LACE
NIT
U
(C
SED
ENTRAL
2. A
F
TTACH
OOTSWITCH
U
.
NIT
3. T
O
T
URN
N
N
:
OTE
L
IGHT
I
NDICATING
P
T
LUG
A S
TANDARD
M
INIMUM OF
I
S
MPORTANT
AFETY
P
W
ROCEDURES
(2)
,
CENTER THE SPONGE AND RELEASE PRESSURE ON ELECTRODE SHAFT
(
)
SUGGESTED
OR DISTILLED WATER
(B)
LOCATED ON THE RIGHT SIDE OF THE SOCKET BAR
RF-P
.
OST
PLUG MALE BANANA PLUG INTO SOCKET
INSERT TIP OF INSTRUMENT INTO SPONGE HOLDER
. I
RF–
. P
POST
LUG MALE BANANA PLUG INTO SOCKET
(B)
INTO SOCKET
ON UNIT
)
OR DISTILLED WATER
. I
F THERE IS AN INSULATION FAILURE THE ALARM WILL
.
. A
.
F
U
OR
SE
O
A L
S
I
N
EVEL
URFACE
S
, OR, E
TERILIZATION
T
T
S
O
HE
OCKET
T
C
T
B
O
EST
HARGE
HE
ATTERY
I
T
B
I
L
T
F
HE
ATTERY
S
OW
A
T
P
S
BOVE
HE
OWER
WITCH
T
T
U
N
HAT
HE
NIT
EEDS
C
I
T
B
HE
HARGER
NTO
HE
AC O
A
L
UTLET
ND
4 H
.
OURS
N
U
N
:
EVER
SE
OTICE
C
C
ITH
HARGER
ONNECTED
,
SIZE TO BE TESTED
BY PRESSING THE MIDDLE PART OF THE
(
SQUEEZE OUT EXCESS WATER
F A DEFECT IS FOUND ON THE SHAFT INSULATION
.
. I
NSERT INSULATION TEST ELECTRODE
¼"
(
SQUEEZE OUT EXCESS WATER
(1). P
LUG THE MALE END OF THE CABLE INTO THE
. C
ONNECT INSULATED TEST
(2)
(3). I
INTO HANDLE
CTIVATE THE FOOTSWITCH AND WITH THE UPPER
. C
T
A
T
B
N
HE
REA
O
E
.)
TC
I
T
B
O
T
N
HE
ACK
F
HE
.
Y
W
HE
ELLOW
ARNING
W
L
U
ILL
IGHT
P
T
B
C
.
O
E
HARGED
O
T
U
A
ACK
F
HE
NIT
ND
I
C
F
A
ET
T
HARGE
OR
U
F
T
NIT
OR
EST
T
U
.
O
NIT
3
, (C1),
MM DIAMETER
.
). N
:
OTE
.
.
(A)
ON THE
,
SLOWLY
(A)
ON THE
(2)
INTO
OF THE SPONGE
).
.
NSERT A SPONGE OF
(
SQUEEZE
ONFIRM FAILURE WITH

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ENDO TEST II and is the answer not in the manual?

Questions and answers

Summary of Contents for ELMED ENDO TEST II

  • Page 1 ENDO TEST II MANUAL NSTRUCTIONS 1. P LACE EVEL URFACE , OR, E ENTRAL TERILIZATION 2. A TTACH OOTSWITCH OCKET 3. T HARGE ATTERY ATTERY ELLOW ARNING IGHT BOVE OWER WITCH IGHT NDICATING EEDS HARGED HARGER AC O TANDARD UTLET...
  • Page 2 BIPOLAR FORCEPS 1. T HE SET UP OF THE UNIT AND ACCESSORIES IS THE SAME AS TESTING ONOPOLAR ORCEPS ONNECT THE ALLIGATOR CLIP TO ONE TIP OF THE FORCEPS 2. W ITH THE FORCEPS IN ONE HAND AND THE INSULATED TEST HANDLE IN THE OTHER ACTIVATE THE FOOTSWITCH AND WITH THE UPPER PART OF THE SPONGE SWIPE THE INSULATION ON THE ONE SHANK WHERE THE CLIP IS CONNECTED...