FONIX 6500-CX TARGET 2CC SSPL 90 FONIX ® 6500 TARGET 2CC F.O.G. NOTE: Use the remote module for this procedure. NOTE: Use the remote module for these procedures. GENERAL PROCEDURE — using all AVERAGE EAR data: 1. In the probe menu highlight CREATE TARGET a.
FONIX 6500-CX QUIK-PROBE FONIX 6500-CX QUIK-PROBE (Automatic) INSERTION GAIN (IG) TEST (Automatic) SPL TARGET TEST 1. Enter probe menu by pressing [PROBE] (or [START/STOP] on 1. Enter probe menu by pressing [PROBE] or [START/STOP] on remote) and then pressing [MENU].
6500-CX AGC/ADAPTIVE AGC FONIX 6500-CX LINEAR ANSI S3.22-1987 TEST ANSI S3.22–1987 TEST 1. LEVEL (only if UNLEVELED). 1. LEVEL (only if UNLEVELED) 2. Attach hearing aid & coupler to microphone, set full on 2. Attach hearing aid & coupler to microphone, set full on gain, place in test box.
FONIX 6500-CX FONIX 6500-CX Multi-Curve—Pure Tone Multi-Curve—Composite (Software Version 2.50 and up) (Software Version 2.50 and up) Procedure: Using Multi-Curve to test four different hearing aids. All Procedure: Using Multi-Curve to test four different hearing aids. these test examples can be run as either SPL or Gain.
FONIX 6500-CX IEC TEST 1. LEVEL (only if UNLEVELED) 2. Attach the hearing aid and appropriate coupler to microphone. 3. Linear aids—set to full-on gain Compression aids—set to maximum output (usually minimum compression). 4. Place in test box. 5. Press [MENU], then [IEC].
FONIX 6500-CX AGC/ADAPTIVE AGC FONIX 6500-CX LINEAR ANSI S3.22-1996 TEST ANSI S3.22-1996 TEST 1. LEVEL (only if UNLEVELED). 1. LEVEL (only if UNLEVELED) 2. Attach hearing aid and coupler to microphone, set full on gain, 2. Attach hearing aid and coupler to microphone, set full on gain, place in place in sound chamber.
ANSI S3.22-1996 Specifications and Tolerances OSPL90 Curve: Max SPL shall not exceed the mfgr. spec. by +3 dB HFA-OSPL90 or SPA-OSPL90: Shall be within ± 4 dB of mfgr. spec. Full-on Gain (FOG): 50 or 60 dB inputs HFA-FOG or SPA-FOG: within ± 5 dB of mfgr. spec. Reference Test Gain: information only Freq.
ANSI S3.22–1987 Recommended Measurements, Specifications and Tolerances SSPL9O CURVE: Max. shall not exceed that spec. mfgr + 3dB HFA-SSPL9O or SPA-SSPL9O: Shall be within ± 4 dB of mfgr. spec. FULL-ON GAIN: 50 or 60 dB inputs HFA/FOG or SPA/FOG: Within ± 5 dB of mfgr. spec. REFERENCE TEST GAIN: Shall be stated for info only FREQUENCY RESPONSE CURVE: 60 db input FREQUENCY RANGE: (f1 - f2): f1 ±...
Family of Frequency Response Curves: These curves are developed by adjusting the noise input level in 10-dB steps over a selected range. The preferred levels are 50, 60, 70, 80, and 90dB SPL. Frye has added 40dB SPL to the preferred levels.
6500-CX Digital Speech in Noise— 6500-CX Digital Speech in Noise— Sound Chamber Probe 1. Level (only if unleveled) 1. Follow the instructions to do SPL testing. 2. Once in the main SPL testing screen press menu 2. Set up the hearing aid and coupler with the microphone.