Page 2
There is also 3 Digital Input or Output to synchronize the BFP system with other equipment. Since each camera, the BFP driver and the LED driver are connected to a computer via USB cables, a 4 ports USB 3.0 hub is included to connect only 1 USB cable to the computer..
2.1. BEFORE connecting anything to the computer, you need to install the camera driver that is on the USB key supplied with the system. Launch Cam33U_setup_4.7.0.1557.exe. 2.2. Install Doric Neuroscience Studio, using the version on the USB key supplied with the system; Launch setup_DoricStudio5.4.0.15.exe Or looks for the most recent release in our website : http://doriclenses.com/life-sciences/software/955-doric-neuroscience-studio.html...
Page 4
3.2. Bundle imaging Fluorescence Mini Cube 'BFMC' This is the central part of the BFP system. It contain set of dicrhroic mirrors and bandpass spectral filters with optics to have uniform illumination on optical fibre port that is re-imaged to an image snesor.
Page 5
BFP SYSTEM : Quick Starting Guide 3.3. BFP Driver 'BFPD' The BFP Driver 'BFPD' synchronize the LED driver with camera acquisition in an interleave mode. There is also 3 Digital Input or Output to synchronize the BFP system with other equipment.
Page 6
4.1. Load configurations files If a configuration already exist, you may load the configuration by clicking Load Configuration buttons in the LED driver and the BFP driver tabs (2 different files). 4.2. Set LED driver Turn ON the LED Driver, and set each channel to TTL by pushing the knob 2 times, then adjust the current to the desired value by rotating the knob.