It is easy to use with a simple user interface and keypad. ZEVEX technology allows an accuracy rate of +/- 5%, and the ability to safely operate in any ori- entation.
LIST OF WARNINGS AND CAUTIONS WARNINGS Use ONLY feeding solutions prescribed by the responsible physician, regis- tered dietitian, registered nurse, or other licensed practitioner. Use ONLY EnteraLite ® Infinity ™ Disposable Sets to ensure proper fluid deliv- ery. Others will not deliver the correct dose, may allow dangerous free-flow conditions, and may generate hazardous fluid pressures which may activate occlusion alarms at unpredictable pressures.
1. Pump Components Display The display includes large alphanumeric characters, as well as smaller symbols and words. All segments and symbols on the dis- play are shown in Figure 1-4. Figure 1-4 Display Segments Settings and messages are displayed in large alphanumeric characters.
1. Pump Components Symbols Brief instructions for operation of pump as Consult Accompanying well as a brief explanation of each alarm Documents! message which pump may communicate are printed on the label attached to pump. Figure 1-11 These are not intended to be used in place Consult Accompanying Documents of the Operator's Manual.
Page 15
2. Directions for Use NOTE: If set is to be used with a carry pack, all Tubing air must be removed from bag and tubing. Port Continue to step 3 for instructions on removing air. If set is to be hung above pump, i.e.
2. Directions for Use Operating the Pump A Single Feeding Example If you want to deliver 500 ml of enteral feeding solution at a rate of 120 ml/hr and then stop the pump: Key Sequence for “A Single Feeding Example” 120 ml/h 500 ml Step 1:...
Page 19
2. Directions for Use When you would like to restart pump, press the ON/OFF and/or the RUN/PAUSE key. Pump will save the memory of where it was in the feeding before stopping. Press the RUN/PAUSE key and pump will restart at the point where it was stopped.
Page 21
2. Directions for Use Step 4: Press the key to change the rate to 50 ml/hr. Hold down either key to change rapidly (Figure 2-17). Step 5: Figure 2-17 Program Press the RATE/DOSE key to display dose. Press the Rate to 50 ml/hr key until a dose of 100 ml is displayed (Figure 2-18).
Page 23
2. Directions for Use Feeding Completion: When the dose has been completely delivered, pump will stop running and display will read NEXT DOSE then give the number of hours and minutes until feeding will be repeated (Figure 2-23). Display will show the run symbol flashing without the arcs rotating around it.
2. Directions for Use The Battery Charging the Battery To charge the battery, insert plug from AC Adapter/Charger into port on the left side of pump. Plug charger into a wall outlet. Pump will charge whether it is turned off or running.
Page 27
2. Directions for Use To adjust these settings: Key Sequence for “Adjusting User Preference Settings” HIGH LOCK MUTE LITE ON UNLK BEEP LITE OFF Step 1: With pump turned off, press and hold for 1.5 seconds the and the ON/OFF keys at the same time.
3. Alarms, Messages, and Indications Alarms ERRA - ERRZ (Figure 3-1) Pump will sound a continuous beep and display will read ERR followed by a letter to indicate a self-test has failed. To clear alarm: Turn the pump off and then back on. LOAD SET (Figure 3-2) Pump will sound a dual tone beep repeatedly and dis- Figure 3-1 ERRA...
3. Alarms, Messages, and Indications NO FOOD (Figure 3-6) Pump will sound a dual tone beep repeatedly and dis- play will read NO FOOD. Why: This alarm has occurred because pump has detected air in the tubing. It takes approximately 1 ml of air, which is approximately 5 inches (12.7 cm) in length entering the teal colored tubing to cause an alarm.
3. Alarms, Messages, and Indications Indications CHEK (Figure 3-10) Pump will sound a dual tone beep once and CHEK, RATE, DOSE, and FEED INTERVAL will all blink on the display. Figure 3-10 Why: This alarm sounds when the RATE, DOSE, and FEED CHEK Indication INTERVAL are not compatible.
4. Cleaning 4. CLEANING WARNING: To avoid electrical shock, never clean pump or EnteraLite ® Infinity ™ AC Adapter/Charger with charger plugged into an outlet or pump turned on. WARNING: Make sure the EnteraLite ® Infinity ™ AC Adapter/Charger is com- pletely dry before plugging into an electrical outlet.
Page 37
5. Accessories EnteraLite ® Infinity ™ Mini-Backpack Order Numbers PCK1001 (red), PCK1002 (teal) Pack holds EnteraLite ® Infinity ™ with a 500 ml bag in the front section or can accommodate a 1200 ml bag in the rear section. Pump is secured with a Velcro ®...
Page 39
6. Specifications VOLUME DISPLAY Current Dose Status Accumulated INTERVAL FEED Interval Feed Indication Displays Time to Next Feed PROGRAM MEMORY RETAINED When Pump is on or off PEDIATRIC USE PRIME ALARMS Chek Programming error Load Set Set not loaded properly Low Batt Battery Low No Flow In...
Accuracy % = (Volume Delivered - Volume Programmed) x 100 Volume Programmed utilizing ZEVEX EnteraLite® Infinity™ disposable sets fluid head height at +6.0 inches ± 0.3 inches (+15.24 cm ± 0.76 cm) with respect to center of rotor.
Routine Maintenance There are no user serviceable parts or routine calibration or adjustment pro- cedures required for EnteraLite ® Infinity ™ . All service requests should be referred to ZEVEX technical service. EnteraLite ® Infinity ™ Operator’s Manual...
Infinity ™ products of its manufacture to be free from defects in material and workmanship, and will replace or repair, F.O.B., at its factory in Salt Lake City, Utah, or other location designated by ZEVEX, any EnteraLite ® Infinity ™...
Page 47
Figure Description Page 2-29 Backlight Settings: LITE OFF and LITE ON ......19 ERRA and ERRZ Alarm ............22 LOAD SET Alarm ..............22 LOW BATT Alarm ............... 22 NO FLOW IN Alarm ............23 NO FLOW OUT Alarm ............23 NO FOOD Alarm ..............