INDICATIONS FOR USE .................. LIST OF WARNINGS AND CAUTIONS ............. v Warnings ......................iv 1. PUMP COMPONENTS ..................1 Infinity Orange Small Volume Enteral Feeding Pump ........1 Keypad ......................3 Display ......................4 Disposable Set ....................5 Symbols ......................6 2.
Orange system includes a complete line of Di(2-ethylhexyl)phthalate (DEHP) - free pump delivery sets with automatic free-flow protection. The Infinity Orange is designed for both hospital and alternate care use. It is durable, water resistant, and easily carried in a ZEVEX carry pack. The Infinity Orange’s size, weight, accuracy, and portability promote and support health im-...
Make sure the Infinity AC Adapter/Charger is completely dry before plugging into an electrical outlet. Do not use Infinity Orange enteral feeding pump for delivery of non-enteral solutions. Serious injury may result. If an error occurs (ER01 - ER99 will appear in display), all settings and vol- ume counters should be checked and verified before starting a feeding.
Page 6
Federal law (U.S.A.) restricts this device to sale by or on the order of a physi- cian, registered dietitian, registered nurse, or other licensed practitioner. Dispose of Infinity Orange Disposable Sets properly, as required by local law. If any leaks are detected in the disposable set, stop pump operation and disconnect set from patient and replace with a new disposable set.
1. PUMP COMPONENTS Keypad Press key for 1.5 seconds to ON/OFF either turn Infinity Orange on or off. Press and hold key to rapidly PRIME fill disposable set with fluid. Release key to stop. Press key to start or RUN/PAUSE pause pump, or to clear an alarm.
1. PUMP COMPONENTS Display The display includes large alphanumeric char- acters, as well as smaller symbols and words. All segments and symbols on the display are shown in Figure 1-4. Settings and messages are displayed in large alphanumeric characters (Figure 1-5). Figure 1-4 Display Segments When the large alphanumeric characters dis- play numbers, one of the smaller words below...
1. PUMP COMPONENTS Disposable Set The only disposable sets approved for use with Infinity Orange pump are: 4 Infinity Orange 100 ml Bag Set - Order Number INF0100 4 EnteraLite Infinity 500 ml Bag Set - Order Number INF0500 4 EnteraLite Infinity 1200 ml Bag Set - Order Number INF1200...
Figure 1-11 Consult They are simply a quick reference guide. Accompanying Documents Please read the Operator’s Manual be- fore operating the Infinity Orange. The bag symbol is printed on pump door over Figure 1-12 port where upstream tubing enters pump.
2. DIRECTIONS FOR USE Step 3: Turn bag upside down and gently squeeze. Tilt bag as needed to keep air at tubing port (Figure 2-4). Step 4: Gently pinch orange colored tubing below “ ” symbol. Hold this position until air is removed from tubing.
2. DIRECTIONS FOR USE Operating the Pump A Single Feeding Example If you want to deliver 50 ml of enteral feeding solution at a rate of 100 ml/hr and then stop the pump: Key Sequence for “A Single Feeding Example” FORM ULA 100 ml/hr 50 ml...
2. DIRECTIONS FOR USE To restart pump when in paused mode, press the key. The pump will RUN/PAUSE restart the feeding at the point where it was paused. NOTE: If pump is turned off and turned back on again, the volume delivered coun- ter resets to 0 and food type resets to formula.
2. DIRECTIONS FOR USE The Battery Running on Battery Power Infinity Orange will run for 24 hours at a rate of 125 ml/hr. A fully depleted battery takes ap- proximately 6 hours to fully charge. When pump is running on battery power the...
2. DIRECTIONS FOR USE User Preference Settings There are six user preference settings. These set- tings can be changed to fit the user’s needs. Alarm Volume: The alarm has two volumes, high or low. Pump default is low (Figure 2-19). Settings Lock: Figure 2-19 Alarm Volume Settings BEEP LOW...
Page 25
2. DIRECTIONS FOR USE Step 3: Press the key again. Display will read MUTE WHEN DONE or BEEP PRIME WHEN DONE. Press the + key to change to BEEP WHEN DONE or the key to change to MUTE WHEN DONE. If setting is as desired, do not press either key. Step 4: Press the key again.
3. ALARMS, MESSAGES, AND INDICATIONS ALARMS, MESSAGES, AND INDICATIONS Following is a list of all alarms, display messages, and indications that are used by the Infinity Orange enteral feeding pump. Each alarm, message, or indication is described in detail on the following pages: Figure...
3. ALARMS, MESSAGES, AND INDICATIONS Alarms ER01 - ER99 (Figure 3-1) Pump will sound a continuous beep and display will read ER followed by a two digit number to indicate a self-test has failed. To clear alarm: Turn the pump off and then back on. WARNING: If an error occurs (ER01 - ER99 will ap- pear in display), all settings and volume counters should be checked and verified before starting a feed-...
3. ALARMS, MESSAGES, AND INDICATIONS NO FOOD (Figure 3-6) Pump will sound a dual tone beep repeatedly and display will read NO FOOD. Why: This alarm has occurred because pump has detected air in the tubing. It takes approximately 1 ml of air, which is approximately 5 inches (12.7 cm) in length entering the orange colored tubing to cause an alarm.
3. ALARMS, MESSAGES, AND INDICATIONS Indications TO STOP LET GO (Figure 3-10) When pressing and holding the PRIME key, alarm will sound once, pump will begin pumping at maximum rate of 480 ml/hr and display will read “TO STOP LET GO”. To stop priming: Release the PRIME key.
5. Has the appropriate milk. food type setting on the pump been selected for this If alarm continues, contact your feeding? healthcare provider or ZEVEX Customer Service at (800) 970-2337 for assistance. Infinity ® ™ Orange Operator’s Manual...
Page 34
Turn pump off. Unload and reload delivery set cassette. securely? UNSURE Make sure door is closed securely. Turn pump back on. Contact your healthcare provider or ZEVEX Customer Service at (800) 970-2337 to receive a new door. 2. Is door cracked or tab on inside of door...
Page 35
Customer Service at (800) 970-2337 to return pump for service. 5. Is Battery Symbol and E and F of fuel gauge flashing? If condition continues, contact your healthcare provider or ZEVEX Customer Service at (800) 970-2337 to return pump for service. Infinity ® ™ Orange...
5. CLEANING CLEANING WARNING: To avoid electrical shock, never clean pump or Infinity Orange AC Adapter/Charger with charger plugged into an outlet or pump turned on. WARNING: Make sure the Infinity Orange AC Adapter/Charger is completely dry before plugging into an electrical outlet.
6. ACCESSORIES ACCESSORIES Infinity Brand AC Adapter/Charger Order Number 23401-001 Plug AC Adapter/Charger into a wall outlet and plug connector into pump to operate on AC power and/or recharge battery (Figure 6-1). Figure 6-1 AC Adapter/Charger Infinity Brand Pole Clamp Order Number Z-11981 Thread bolt into receptacle on the back of pump and tighten gray wheel to mount pump...
6. ACCESSORIES Infinity Brand Waist Pack Order Numbers PCK3001 (black and gray) Pack holds Infinity pump with a 100 ml or 500 ml bag. Pump is secured with an elastic strap. A Velcro strap secures the neck of the bag. The tubing is loaded into pump and any excess length of tubing can be secured with a Velcro tab.
Page 41
Temperature: -4 °F to 149 °F (-20 °C to 65 °C) STORAGE Humidity: 10% to 95% noncondensing ENVIRONMENT Infinity Orange Enteral Feeding Pump with internal battery – Type CLASSIFICATION BF Applied Part INFORMATION Infinity Orange AC Adapter / Charger – Class II Infinity ®...
The specifications of primary impor- tance are: 4 The flow rate range of Infinity Orange is 0.1 ml/hr to 10 ml/hr in 0.1 ml/hr incre- ments and 10 ml/hr to 300 ml/hr in 1 ml/hr increments.
Guidance and manufacturer’s declaration - electromagnetic emissions The Infinity Orange is intended for use in the electromagnetic environment specified below. The customer or the user of the Infinity Orange should assure that it is used in such an environment. Emissions Test...
WARRANTY Solely for the benefit of the original buyer, ZEVEX INCORPORATED (“ZEVEX”), warrants all new Infinity Orange products of its manufacture to be free from defects in material and workmanship, and will replace or repair, F.O.B., at its factory in...
Page 47
LIST OF FIGURES Figure 2-11 Program Rate to 100 ml/hr ................. 10 Figure 2-12 Program Dose to 50 ml ................10 Figure 2-13 Pump is Running ..................11 Figure 2-14 Amount Currently Delivered ................. 11 Figure 2-15 Total Amount Delivered ................11 Figure 2-16 DOSE DONE ....................
YOUR HEALTHCARE PROVIDER: _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ YOUR PHYSICIAN: _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ ZEVEX Incorporated 4314 ZEVEX Park Lane Salt Lake City, Utah 84123 USA Customer Service and Technical Support: (800) 970-2337 www.zevex.com Infinity ® ™ Orange Operator’s Manual...
AC Adapter/Charger 15, 30 To clean the AC Adapter/Charger 29 Accessories To clean the Carrying Packs 29 AC Adapter/Charger 30 To clean the Infinity Orange Enteral Adjustable Angle Pole Clamp 30 Feeding Pump 29 Backpack 32 CLEAR Key 3 Mini Backpack 31...