Zevex Infinity Orange Operator's Manual

Small volume enteral feeding pump
Table of Contents

Advertisement

Quick Links

Small Volume
Enteral Feeding Pump
OPERATOR'S MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for Zevex Infinity Orange

  • Page 1 Small Volume Enteral Feeding Pump OPERATOR’S MANUAL...
  • Page 2: Table Of Contents

    INDICATIONS FOR USE .................. LIST OF WARNINGS AND CAUTIONS ............. v Warnings ......................iv 1. PUMP COMPONENTS ..................1 Infinity Orange Small Volume Enteral Feeding Pump ........1 Keypad ......................3 Display ......................4 Disposable Set ....................5 Symbols ......................6 2.
  • Page 3 Infinity ® ™ Orange Operator’s Manual...
  • Page 4: Product Overview

    Orange system includes a complete line of Di(2-ethylhexyl)phthalate (DEHP) - free pump delivery sets with automatic free-flow protection. The Infinity Orange is designed for both hospital and alternate care use. It is durable, water resistant, and easily carried in a ZEVEX carry pack. The Infinity Orange’s size, weight, accuracy, and portability promote and support health im-...
  • Page 5: List Of Warnings And Cautions

    Make sure the Infinity AC Adapter/Charger is completely dry before plugging into an electrical outlet. Do not use Infinity Orange enteral feeding pump for delivery of non-enteral solutions. Serious injury may result. If an error occurs (ER01 - ER99 will appear in display), all settings and vol- ume counters should be checked and verified before starting a feeding.
  • Page 6 Federal law (U.S.A.) restricts this device to sale by or on the order of a physi- cian, registered dietitian, registered nurse, or other licensed practitioner. Dispose of Infinity Orange Disposable Sets properly, as required by local law. If any leaks are detected in the disposable set, stop pump operation and disconnect set from patient and replace with a new disposable set.
  • Page 7 Infinity ® ™ Orange Operator’s Manual...
  • Page 8: Pump Components

    1. PUMP COMPONENTS 1. PUMP COMPONENTS Infinity Orange Small Volume Enteral Feeding Pump Figure 1-1 Pump: Front 1. Door 2. Pump Wheel 3. Display 4. Upstream Pressure Sensor 5. Air Sensor 6. Downstream Pressure Sensor 7. Receptacle for Cassette 8. Keypad...
  • Page 9: Figure 1-2 Pump: Rear

    1. PUMP COMPONENTS Infinity Orange Small Volume Enteral Feeding Pump Figure 1-2 Pump: Rear 1. Pump Instruction Label 2. Speaker 3. Receptacle for Pole Clamp 4. Serial Number Label 5. Charger/Power Connector Infinity ® ™ Orange Operator’s Manual...
  • Page 10: Keypad

    1. PUMP COMPONENTS Keypad Press key for 1.5 seconds to ON/OFF either turn Infinity Orange on or off. Press and hold key to rapidly PRIME fill disposable set with fluid. Release key to stop. Press key to start or RUN/PAUSE pause pump, or to clear an alarm.
  • Page 11: Display

    1. PUMP COMPONENTS Display The display includes large alphanumeric char- acters, as well as smaller symbols and words. All segments and symbols on the display are shown in Figure 1-4. Settings and messages are displayed in large alphanumeric characters (Figure 1-5). Figure 1-4 Display Segments When the large alphanumeric characters dis- play numbers, one of the smaller words below...
  • Page 12: Disposable Set

    1. PUMP COMPONENTS Disposable Set The only disposable sets approved for use with Infinity Orange pump are: 4 Infinity Orange 100 ml Bag Set - Order Number INF0100 4 EnteraLite Infinity 500 ml Bag Set - Order Number INF0500 4 EnteraLite Infinity 1200 ml Bag Set - Order Number INF1200...
  • Page 13: Symbols

    Figure 1-11 Consult They are simply a quick reference guide. Accompanying Documents Please read the Operator’s Manual be- fore operating the Infinity Orange. The bag symbol is printed on pump door over Figure 1-12 port where upstream tubing enters pump.
  • Page 14: Directions For Use

    3 for instructions on removing air. If set Figure 2-3 Remove Cover from is to be hung above pump, i.e. on an IV pole, you Barbed Adapter / Place Cover may skip to step 4. on Pole Clamp © Copyright 2008 ZEVEX Incorporated...
  • Page 15: Figure 2-4 Squeeze Bag

    2. DIRECTIONS FOR USE Step 3: Turn bag upside down and gently squeeze. Tilt bag as needed to keep air at tubing port (Figure 2-4). Step 4: Gently pinch orange colored tubing below “ ” symbol. Hold this position until air is removed from tubing.
  • Page 16: Figure 2-8 Seat Cassette

    332 ml/hr. Display will read “TO STOP LET GO” (Figure 2-10). Once all air is removed from tubing, release key. Pump will stop, display will revert to last programmed rate, and pump will be in pause mode. © Copyright 2008 ZEVEX Incorporated...
  • Page 17: Operating The Pump A Single Feeding Example

    2. DIRECTIONS FOR USE Operating the Pump A Single Feeding Example If you want to deliver 50 ml of enteral feeding solution at a rate of 100 ml/hr and then stop the pump: Key Sequence for “A Single Feeding Example” FORM ULA 100 ml/hr 50 ml...
  • Page 18: Figure 2-13 Pump Is Running

    Display will show RUN/PAUSE rate, and pump will be in pause mode. Pump will save the memory of where it was in the feeding before being paused. Or, turn pump off by pressing the key. ON/OFF © Copyright 2008 ZEVEX Incorporated...
  • Page 19: Figure 2-16 Dose Done

    2. DIRECTIONS FOR USE To restart pump when in paused mode, press the key. The pump will RUN/PAUSE restart the feeding at the point where it was paused. NOTE: If pump is turned off and turned back on again, the volume delivered coun- ter resets to 0 and food type resets to formula.
  • Page 20 CON FIRM DOSE AND FOOD TYPE. NOTE: Rate, dose, and food type settings can be locked so that changes cannot be made. See the User Preference Settings section (Page 16) for instructions to change these settings. © Copyright 2008 ZEVEX Incorporated...
  • Page 21: The Battery

    2. DIRECTIONS FOR USE The Battery Running on Battery Power Infinity Orange will run for 24 hours at a rate of 125 ml/hr. A fully depleted battery takes ap- proximately 6 hours to fully charge. When pump is running on battery power the...
  • Page 22: Charging The Battery

    NOTE: If pump shut down due to low battery, charging less than 10 minutes may result in an ER99 alarm. Fully charge battery if depleted (see page 20 to clear alarm). Figure 2-18 Fuel Gauge © Copyright 2008 ZEVEX Incorporated...
  • Page 23: User Preference Settings

    2. DIRECTIONS FOR USE User Preference Settings There are six user preference settings. These set- tings can be changed to fit the user’s needs. Alarm Volume: The alarm has two volumes, high or low. Pump default is low (Figure 2-19). Settings Lock: Figure 2-19 Alarm Volume Settings BEEP LOW...
  • Page 24: Figure 2-23 Max Rate Settings

    Display will read UNLK (unlocked) or LOCK (locked). PRIME Press the key to change to LOCK (locked) or the key to change to UNLK (unlocked). If setting is as desired, do not press either key. © Copyright 2008 ZEVEX Incorporated...
  • Page 25 2. DIRECTIONS FOR USE Step 3: Press the key again. Display will read MUTE WHEN DONE or BEEP PRIME WHEN DONE. Press the + key to change to BEEP WHEN DONE or the key to change to MUTE WHEN DONE. If setting is as desired, do not press either key. Step 4: Press the key again.
  • Page 26: Alarms, Messages, And Indications

    3. ALARMS, MESSAGES, AND INDICATIONS ALARMS, MESSAGES, AND INDICATIONS Following is a list of all alarms, display messages, and indications that are used by the Infinity Orange enteral feeding pump. Each alarm, message, or indication is described in detail on the following pages: Figure...
  • Page 27: Alarms

    3. ALARMS, MESSAGES, AND INDICATIONS Alarms ER01 - ER99 (Figure 3-1) Pump will sound a continuous beep and display will read ER followed by a two digit number to indicate a self-test has failed. To clear alarm: Turn the pump off and then back on. WARNING: If an error occurs (ER01 - ER99 will ap- pear in display), all settings and volume counters should be checked and verified before starting a feed-...
  • Page 28: No Flow In

    Figure 3-5 NO FLOW OUT Alarm pecially around pressure sensors. If cleaning is necessary, refer to CHAPTER 5 - CLEANING (Page 29). 3. Back pressure from patient may also cause this alarm. Discuss with physician. © Copyright 2008 ZEVEX Incorporated...
  • Page 29: No Food

    3. ALARMS, MESSAGES, AND INDICATIONS NO FOOD (Figure 3-6) Pump will sound a dual tone beep repeatedly and display will read NO FOOD. Why: This alarm has occurred because pump has detected air in the tubing. It takes approximately 1 ml of air, which is approximately 5 inches (12.7 cm) in length entering the orange colored tubing to cause an alarm.
  • Page 30: Shut Door

    This message indicates a feeding dose has been delivered. To clear pump: Either press the key to ON/OFF turn pump off, or press the key to RUN/PAUSE put pump in pause mode before starting another feeding. Figure 3-9 DOSE DONE Message © Copyright 2008 ZEVEX Incorporated...
  • Page 31: To Stop Let Go

    3. ALARMS, MESSAGES, AND INDICATIONS Indications TO STOP LET GO (Figure 3-10) When pressing and holding the PRIME key, alarm will sound once, pump will begin pumping at maximum rate of 480 ml/hr and display will read “TO STOP LET GO”. To stop priming: Release the PRIME key.
  • Page 32: Battery Doesn't Hold Its Charge

    4 Charger is connected properly. If this does not correct the problem, pump or charger may be damaged. Contact your healthcare provider for service. NOTE: There are no user serviceable or replaceable components inside the Infin- ity Orange. © Copyright 2008 ZEVEX Incorporated...
  • Page 33: Troubleshooting Guide

    5. Has the appropriate milk. food type setting on the pump been selected for this If alarm continues, contact your feeding? healthcare provider or ZEVEX Customer Service at (800) 970-2337 for assistance. Infinity ® ™ Orange Operator’s Manual...
  • Page 34 Turn pump off. Unload and reload delivery set cassette. securely? UNSURE Make sure door is closed securely. Turn pump back on. Contact your healthcare provider or ZEVEX Customer Service at (800) 970-2337 to receive a new door. 2. Is door cracked or tab on inside of door...
  • Page 35 Customer Service at (800) 970-2337 to return pump for service. 5. Is Battery Symbol and E and F of fuel gauge flashing? If condition continues, contact your healthcare provider or ZEVEX Customer Service at (800) 970-2337 to return pump for service. Infinity ® ™ Orange...
  • Page 36: Cleaning

    5. CLEANING CLEANING WARNING: To avoid electrical shock, never clean pump or Infinity Orange AC Adapter/Charger with charger plugged into an outlet or pump turned on. WARNING: Make sure the Infinity Orange AC Adapter/Charger is completely dry before plugging into an electrical outlet.
  • Page 37: Accessories

    6. ACCESSORIES ACCESSORIES Infinity Brand AC Adapter/Charger Order Number 23401-001 Plug AC Adapter/Charger into a wall outlet and plug connector into pump to operate on AC power and/or recharge battery (Figure 6-1). Figure 6-1 AC Adapter/Charger Infinity Brand Pole Clamp Order Number Z-11981 Thread bolt into receptacle on the back of pump and tighten gray wheel to mount pump...
  • Page 38: Figure 6-4 Mini Backpack

    Feeding Bag Strap 500 ml Bag Compartment Tubing Strap Pump Location Pump Strap Figure 6-8 Super-Mini Backpack Figure 6-9 Super-Mini Backpack Figure 6-7 Super-Mini Backpack (Cutaway View of Internal (Cutaway View of Internal Straps) Compartments) © Copyright 2008 ZEVEX Incorporated...
  • Page 39: Figure 6-10 Waist Pack

    6. ACCESSORIES Infinity Brand Waist Pack Order Numbers PCK3001 (black and gray) Pack holds Infinity pump with a 100 ml or 500 ml bag. Pump is secured with an elastic strap. A Velcro strap secures the neck of the bag. The tubing is loaded into pump and any excess length of tubing can be secured with a Velcro tab.
  • Page 40: Specifications

    1 ml from 10 - 3000 ml Current Dose Status VOLUME DISPLAY Accumulated When Pump is on or off PROGRAM MEMORY RETAINED PEDIATRIC USE PRIME Automatic Free-flow Protection DISPOSABLE SET No Drip Chamber FEATURES DEHP-free © Copyright 2008 ZEVEX Incorporated...
  • Page 41 Temperature: -4 °F to 149 °F (-20 °C to 65 °C) STORAGE Humidity: 10% to 95% noncondensing ENVIRONMENT Infinity Orange Enteral Feeding Pump with internal battery – Type CLASSIFICATION BF Applied Part INFORMATION Infinity Orange AC Adapter / Charger – Class II Infinity ®...
  • Page 42: Important Pediatric Considerations

    The specifications of primary impor- tance are: 4 The flow rate range of Infinity Orange is 0.1 ml/hr to 10 ml/hr in 0.1 ml/hr incre- ments and 10 ml/hr to 300 ml/hr in 1 ml/hr increments.
  • Page 43: Additional Technical Information

    Guidance and manufacturer’s declaration - electromagnetic emissions The Infinity Orange is intended for use in the electromagnetic environment specified below. The customer or the user of the Infinity Orange should assure that it is used in such an environment. Emissions Test...
  • Page 44 Therefore, approximately 2.5 ml of fluid remains in the disposable set when pump stops feeding. Routine Maintenance There are no user serviceable parts or routine calibration or adjustment procedures required for Infinity Orange. All service requests should be referred to ZEVEX tech- nical service. © Copyright 2008 ZEVEX Incorporated...
  • Page 45: Warranty

    WARRANTY Solely for the benefit of the original buyer, ZEVEX INCORPORATED (“ZEVEX”), warrants all new Infinity Orange products of its manufacture to be free from defects in material and workmanship, and will replace or repair, F.O.B., at its factory in...
  • Page 46: List Of Figures

    Tubing Segment Being Pinched ..............8 Figure 2-7 Tubing Segment Being Punctured by Fingernail ......... 8 Figure 2-8 Seat Cassette ....................9 Figure 2-9 Close Pump Door ..................9 Figure 2-10 Priming Indication ..................9 © Copyright 2008 ZEVEX Incorporated...
  • Page 47 LIST OF FIGURES Figure 2-11 Program Rate to 100 ml/hr ................. 10 Figure 2-12 Program Dose to 50 ml ................10 Figure 2-13 Pump is Running ..................11 Figure 2-14 Amount Currently Delivered ................. 11 Figure 2-15 Total Amount Delivered ................11 Figure 2-16 DOSE DONE ....................
  • Page 48 Figure 6-11 Waist Pack (Cutaway View of Internal Compartment and Straps) ..32 Figure 6-12 Backpack ...................... 32 Figure 6-13 Backpack (Cutaway View of Internal Compartments) ........ 32 Figure 6-14 Backpack (Cutaway View of Internal Straps) ..........32 Figure 8-1 Head Height Effect on Accuracy ..............35 © Copyright 2008 ZEVEX Incorporated...
  • Page 49: Who To Call

    YOUR HEALTHCARE PROVIDER: _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ YOUR PHYSICIAN: _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ ZEVEX Incorporated 4314 ZEVEX Park Lane Salt Lake City, Utah 84123 USA Customer Service and Technical Support: (800) 970-2337 www.zevex.com Infinity ® ™ Orange Operator’s Manual...
  • Page 50: Index

    AC Adapter/Charger 15, 30 To clean the AC Adapter/Charger 29 Accessories To clean the Carrying Packs 29 AC Adapter/Charger 30 To clean the Infinity Orange Enteral Adjustable Angle Pole Clamp 30 Feeding Pump 29 Backpack 32 CLEAR Key 3 Mini Backpack 31...
  • Page 51 INDEX Electromagnetic Emissions/Interference Maintenance 37 Messages 23 DOSE DONE 12, 23 Error, ER01 - ER99 20 Mini Backpack 31 Feeding Completion 12 Feeding Example (Single Feeding) 10 Occlusion (Blockage) 21 Flow Rate 10, 33, 35 ON/OFF Key 3 Foam 7, 22 Operating Orientation 33 Food Type 3, 11, 24, 26 Fuel Gauge 4, 15...
  • Page 52 User Preference Settings 16 Adjusting User Preference Settings 17 Alarm Volume 16 Backlight 16 Dose Done 16 Max Dose 17 Max Rate 16 Settings Lock 16 VOL/TOTAL Key 3, 11 Waist Pack 32 Water 29 © Copyright 2008 ZEVEX Incorporated...
  • Page 53 NOTES Infinity ® ™ Orange Operator’s Manual...
  • Page 54 D503,799, D503,978, D504,506, D505,199 and D507,647 and Foreign Patent No’s 40108978.9, 40209539.1, 40302879.5, ZL 01338370.1, ZL 02369422.4, 03306228.5, 015998, 701976-701985, 16,112, 2104612, 3012123, 3010564, 129,165, 129,630 and 241366. Other US and Foreign patents pending. Infinity, Orange, and ZEVEX are registered trademarks of ZEVEX Incorporated. 28945-001 Rev. C...