From Eye to Insight Leica FL400 for M530 User Manual 10 744 764 – Version 03...
Page 2
The information provided by this manual is directly related to the operation of the equipment. Medical decision remains the responsibility of the clinician. Leica Microsystems has made every effort to provide a complete and clear user manual highlighting the key areas of product use. Should additional information regarding the use of the product be required, please contact your local Leica representative.
Activating the FL400 through the GUI Operation Operating Range Limits for FL400 Pre-operation checklist Using the Leica FL400 for M530 Care and maintenance What to do, if… General Status messages Electrical Common troubleshooting steps Leica FL400 for M530 / Ref. 10 744 764 / Version 03...
Product identification In this user manual the functions of the Leica FL400 for M530 are described. For information and description of the Leica surgical The model and serial numbers of your product is located on the side microscope, please refer to the user manual of the particular and top of the Leica FL400 for M530 filter module.
NOTE Appearance of GUI screenshots The type label is located on the side of the Leica FL400 for M530 The graphical user interface (GUI) screenshots are for reference only filter module. and can vary in function of the configured options and different surgical microscopes.
Leica FL400 user. Design The Leica FL400 for M530 is an accessory for the Leica M530 surgical microscope. 4.2.1 Optics carrier with Leica FL400 for M530...
Page 7
For the joystick and the other keys presets are available according to your task. Magnifi- Working Magnifi- ARveo 8: only the switches 2 and 5 may be individually cation distance cation assigned. Selected brakes Selected brakes Leica FL400 for M530 / Ref. 10 744 764 / Version 03...
Select the "Fluorescence FL400" user preset button (3). NOTE The FL400 user presets are only available if the FL400 accessory is enabled in the accessory list of your Leica Surgical Microscope. If not, please get in contact with your local Leica Service.
"ON/OFF button (1) on the left handle. NOTE no function Do not change the settings or edit the user list when the Left Handle microscope is in direct use on a patient. The FL400 mode is automatically stopped when modifying the settings. FL400 function no function Press the joystick to the left: The main page for "Fluorescence FL400"...
Get current settings Creating your own FL400 user On the page "Main > Fluorescence FL400" for both white light and You can create and save your own FL400 settings according to your FL400 modes, a "Show Settings" button can be found.
"Fluorescence" by using the arrowheads (3). Tap the "Menu" button in the header. Select "User Settings". Select the "FL" tab to set the brightness in FL400 mode to an adequate level. Recommended and default brightness level is 100%. Select the desired "Fluorescence FL400" function you wish to assign.
Please note that a factory limitation is applied to the working After adapting the user settings, you can save your changes under a distance (see 7.1) as soon as the FL400 mode is activated. Hence, if new username. working distance (WD) limitation in White light mode is switched off and the working distance is above this limit (see 7.1) when...
Page 13
Please refer to the user settings chapter in the manual of your respective surgical microscope for detailed instructions on how to set or change passwords. Leica FL400 for M530 / Ref. 10 744 764 / Version 03...
Please note that the "passcode" has to be typed twice to confirm it. Select a surgeon profile for which the FL400 profile has been already been defined at the "Select surgeon" page: If not, you have to create a new surgeon profile.
Page 15
The dialog box is closed and the selected surgeon's profile and FL400 mode by pushing the joystick of the left handle towards function mappings are now copied into the new surgeon the left.
Page 16
6.2.3 Creating a brand new profile In both cases (copying FL400 preset or copying an existing surgeon profile), after pressing "Create New", the live screen is displayed. To create a new profile, you must first follow the steps described in You can see the surgeon profile's name on the top right corner of 6.2.1.
Page 17
Tap on the "User Inputs" button (2) to open the "User Inputs" menu. The "User Inputs" menu is displayed: Tap "Handles" button (3) to show the "Left Handle" settings. Leica FL400 for M530 / Ref. 10 744 764 / Version 03...
Page 18
When you are on the handle configuration page, click the button you want to assign for the activation of the FL400 mode. This will display the "Select function" popup window for the selected button: Tap the "Back"...
Page 19
Preparation before surgery (ARveo 8) The live screen is displayed: Leica FL400 for M530 / Ref. 10 744 764 / Version 03...
Page 20
Selecting the FL400 Preset user about the profile limitations. NOTE The preset can be used to quickly enable the FL400 ON/OFF functionality on the microscope. This preset profile can be used and modified but any changes to the profile settings will not be stored, therefore with each new start up, the default profile settings will be restored.
The Live Screen is displayed. NOTE Do not change the settings or edit the user list when the microscope is in direct use on a patient. The FL400 mode is automatically stopped when modifying the settings. Tap the "Settings" icon (2) The "Settings"...
NOTE There is a default limitation to the WD and Mag when using the FL400 mode. If the WD and Mag are above this limit before You can activate the FL400 Mode directly through the GUI. Tapping switching to FL400 mode, the WD and Mag will be reduced upon the mode icon (1) a menu of available fluorescence modes is switching to FL400 mode and the picture will get out of focus.
The status LED on the swing arm of the particular surgical • Working Distance: 225 to 350 mm microscope is illuminated in blue if the FL400 mode is active (on • Magnification: 2.0× to 7.0× all surgical microscopes), and white if switched back to white- •...
Please do not dispose waste electrical and electronic equipment in household trash. For care and maintenance of the Leica FL400 for M530, please refer to the user manual of the particular surgical microscope. Leica FL400 for M530 / Ref. 10 744 764 / Version 03...
If the active bulb is used over a long time and its reaching a critical age you receive the recommendation to exchange the bulb(s): Check the life time of the bulb(s) and replace the bulb(s) if the efficiency is too low. Leica FL400 for M530 / Ref. 10 744 764 / Version 03...
Page 26
ARveo 8 - defective lamp 2 alert Replace the lamp. Exchange the defective bulb with a new one before your next case. Leica FL400 for M530 / Ref. 10 744 764 / Version 03...
If electrically operated functions do not work properly, always check these points first: • Is the power switch switched on? • Are the power cables attached correctly? • Are all connecting cables attached correctly? Leica FL400 for M530 / Ref. 10 744 764 / Version 03...
Dark in-vivo FL in the eyepiece and on Insufficient illumination intensity Check the illumination intensity: • verify that the "Brightness FL400" is set to 100% - see the monitor chapter 5.6 and chapter 6.3 • bulb life indicator time...
Page 29
10.1 Leica FL400 characteristics compatibility (EMC) Fluorescence excitation 405 nm (blue light) For the EMC declaration of the Leica FL400 for M530, please Fluorescence signal 635 nm (red) refer to the user manual of the particular surgical microscope. 10.2 Compatibility...