Lufft UMB ANACON 8160.UANA Operating Manual

Analogue/digital converter
Hide thumbs Also See for UMB ANACON 8160.UANA:

Advertisement

Quick Links

Operating Manual
OTT HydroMet Fellbach GmbH, Fellbach, Germany
Operating Manual
UMB Analogue/Digital Converter ANACON
Order Number: 8160.UANA
Status: V8; 27.07.2021
UMB Analogue/Digital Converter ANACON
1
WEEE DE 98036552

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the UMB ANACON 8160.UANA and is the answer not in the manual?

Questions and answers

Summary of Contents for Lufft UMB ANACON 8160.UANA

  • Page 1 Operating Manual UMB Analogue/Digital Converter ANACON Operating Manual UMB Analogue/Digital Converter ANACON Order Number: 8160.UANA Status: V8; 27.07.2021 OTT HydroMet Fellbach GmbH, Fellbach, Germany WEEE DE 98036552...
  • Page 2: Table Of Contents

    UMB Analogue/Digital Converter ANACON Operating Manual TABLE OF CONTENTS ....................4 LEASE READ BEFORE USE ........................6 ESCRIPTION ANACON UMB A ............7 NALOGUE IGITAL ONVERTER ......................11 ONFIGURATION ................12 NSTALLATION AND OMMISSIONING ....................14 ENSOR HANNEL ......................17 ECHNICAL ....................
  • Page 3 Operating Manual UMB Analogue/Digital Converter ANACON OTT HydroMet Fellbach GmbH, Fellbach, Germany...
  • Page 4: Please Read Before Use

    UMB Analogue/Digital Converter ANACON Operating Manual Please read before use Before using the equipment, please read the operating manual carefully and follow the instructions in every detail. Safety instructions  Installation and commissioning must only be carried out by suitably qualified specialist personnel. ...
  • Page 5 Operating Manual UMB Analogue/Digital Converter ANACON Symbols used Important indication concerning possible hazards to the user Important indication for the correct functioning of the device OTT HydroMet Fellbach GmbH, Fellbach, Germany...
  • Page 6: Description

    2 analogue inputs with 24 bit resolution and signal conditioning for resistance, PT100/PT1000, voltage, current, frequency and pulses as well as pre-defined settings for a large number of Lufft environmental sensors. A half-duplex RS485 interface for networking the converters together.
  • Page 7: Digital Converter

    Operating Manual UMB Analogue/Digital Converter ANACON ANACON UMB Analogue/Digital Converter Summary Power supply & RS485 bus interface Input Channel A Input Channel B Status LED’s OTT HydroMet Fellbach GmbH, Fellbach, Germany...
  • Page 8: Anacon Umb Analogue

    UMB Analogue/Digital Converter ANACON Operating Manual Analogue Inputs Channel A and Channel B Name  Channel B Power, Sensor Power Supply ANA_+, Sense High Input ANA_I, Force Low Input  Channel A ANA_-, Force High Input The terminals are suitable for stranded or rigid wires ANACON-UMB without end sleeves with cross section from 0,14 mm Connection...
  • Page 9 Operating Manual UMB Analogue/Digital Converter ANACON Mounting Rail Bus Connector Earth 24V Power Supply, Feed Earth 24V Power Supply, Feed +24V Power Supply, Feed B1, Internal Bus A1, Internal Bus NOTE: The assembly must take place on a grounded DIN rail ! OTT HydroMet Fellbach GmbH, Fellbach, Germany...
  • Page 10 UMB Analogue/Digital Converter ANACON Operating Manual Status LED’s There are 2 LED’s on the front side of the device in order to display the correct configuration and functioning of the ANACON. Their meanings are as follows: Red LED: - If this LED lights up permanently after switch-on there is a configuration fault. - If this LED lights up during operation an unforeseen fault has occurred;...
  • Page 11: Configuration

    Operating Manual UMB Analogue/Digital Converter ANACON Configuration The ANACON modules are all supplied from the factory with ID 1. If several modules are installed in a UMB network, it must be ensured that all modules on the network have a different ID. The following steps should be taken in order to configure the device: 1.
  • Page 12: Installation Andc

    UMB Analogue/Digital Converter ANACON Operating Manual Installation and Commissioning 1. First configure all the modules using the UMB-Config-Tool configuration software in accordance with your requirements (see Configuration chapter). 2. Clip the required number of mounting rail bus connectors onto the DIN rail and then connect all the ANACON modules.
  • Page 13 Operating Manual UMB Analogue/Digital Converter ANACON 8160.UANA Channel A Channel B Input Pin1 Pin2 Pin3 Pin4 Pin1 Pin2 Pin3 Pin4 Resistance (3 wire), PT100/ PT1000 Voltage Current self-powered Current ANACON powered Frequency Impulse/ Digital Input OttHydroMet Sensors 8160.TF# (T) white white 8353.10 (Tipping Bucket) grey...
  • Page 14: Sensor Channel List

    UMB Analogue/Digital Converter ANACON Operating Manual Sensor Channel List Channel assignment equipment class 6 universal measurement transmitters UMB Channel Measurement Range Special Input Variable Unit Temperature temperature -200.0 450.0 °C temperature -328.0 842.0 °F temperature -200.0 450.0 °C temperature -328.0 842.0 °F dewpoint...
  • Page 15 Operating Manual UMB Analogue/Digital Converter ANACON Wind Vect. avg Wind Speed 100.0 Wind Speed 360.0 km/h Wind Speed 223.7 Wind Speed 194,4 Wind Direction 359.9 ° Precipitation precipitation absol. precipitation absol. precipitation diff. precipitation diff. precip. intens. mm/h precip. intens. mm/h Digital input Digital Input...
  • Page 16 UMB Analogue/Digital Converter ANACON Operating Manual Resistance 10200 10220 10240 10260 Resistance 2000 10201 10221 10241 10261 Resistance 2000 Frequency 10300 10320 10340 10360 Frequency 10000 10301 10321 10341 10361 Frequency 10000 Pulses 10500 impulse absol. 65520 pulse 10501 impulse absol. 65520 pulse 10520...
  • Page 17: Technical Data

    Operating Manual UMB Analogue/Digital Converter ANACON Technical Data Power supply Power supply 24VDC +/- 10% Power consumption < 50mA at 24V; excluding sensor Connection Phoenix bus connector, gold, 5 pole Function display Green LED, flashes every 10 seconds RS485 interface internal bus Standard RS485 2 wire, half-duplex Transmission type...
  • Page 18 UMB Analogue/Digital Converter ANACON Operating Manual Measurement Variables Variable Measurement Range Resolution Accuracy (25°C) Voltage 0 ... 100mV 1,526 µV +/- 200 µV Voltage 0 ... 1000 mV 15,26 µV +/- 1 mV Current *1) 0 ... 20 mA (max: 24mA) 305.25 nA +/- 20 µA Current *2)
  • Page 19 Operating Manual UMB Analogue/Digital Converter ANACON LUFFT Sensors Measurement Measurement Physical Physical Sensor Variable Range Variable Measurement Range Temperature Sensor 8160.TF# Temperature -50 ... 100 °C Resistance 60.34 – 138.5 or 8160.TF#s PT100 Ohms Rain Gauge with 0.2mm or Rain 0...
  • Page 20 UMB Analogue/Digital Converter ANACON Operating Manual Storage conditions Permissible ambient temp. -40 ... 70°C Permissible rel. humidity 0 ... 95% RH (non-condensing) Operating conditions Permissible operating temp. -30 ... 60°C Permissible rel. humidity 0 ... 95% RH (non-condensing) Housing Material PA-V0, green Width x Height x Depth 22.5 x 99 x 118.6 mm...
  • Page 21: Maintenance And Care

    Operating Manual UMB Analogue/Digital Converter ANACON Maintenance and Care The product is designed in such a way as to require no maintenance. It does not contain any parts (e.g. batteries) requiring maintenance or exchange. Disposal The device must be disposed of in accordance with European Directives 2002/96/EC and 2003/108/EC (waste electrical and electronic equipment).
  • Page 22: Onformity

    UMB Analogue/Digital Converter ANACON Operating Manual EC Certificate of Conformity Product: Analogue/Digital Converter Type: ANACON UMB (Part No.: 8160.UANA) We herewith certify that the above mentioned equipment complies in design and construction with the Directives of the European Union and specifically the EMC Directive in accordance with 89/336/EC and the Low Voltage Directive in accordance with 73/23/EC.

Table of Contents